将基于长短期记忆网络的运动预测应用于实时肿瘤追踪放射治疗中的成像剂量降低

IF 3.3 3区 医学 Q1 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Physica Medica-European Journal of Medical Physics Pub Date : 2024-09-01 DOI:10.1016/j.ejmp.2024.104507
{"title":"将基于长短期记忆网络的运动预测应用于实时肿瘤追踪放射治疗中的成像剂量降低","authors":"","doi":"10.1016/j.ejmp.2024.104507","DOIUrl":null,"url":null,"abstract":"<div><h3>Purpose</h3><p>To demonstrate the possibility of using a lower imaging rate while maintaining acceptable accuracy by applying motion prediction to minimize the imaging dose in real-time image-guided radiation therapy.</p></div><div><h3>Methods</h3><p>Time-series of three-dimensional internal marker positions obtained from 98 patients in liver stereotactic body radiation therapy were used to train and test the long-short-term memory (LSTM) network. For real-time imaging, the root mean squared error (RMSE) of the prediction on three-dimensional marker position made by LSTM, the residual motion of the target under respiratory-gated irradiation, and irradiation efficiency were evaluated. In the evaluation of the residual motion, the system-specific latency was assumed to be 100 ms.</p></div><div><h3>Results</h3><p>Except for outliers in the superior–inferior (SI) direction, the median/maximum values of the RMSE for imaging rates of 7.5, 5.0, and 2.5 frames per second (fps) were 0.8/1.3, 0.9/1.6, and 1.2/2.4 mm, respectively. The median/maximum residual motion in the SI direction at an imaging rate of 15.0 fps without prediction of the marker position, which is a typical clinical setting, was 2.3/3.6 mm. For rates of 7.5, 5.0, and 2.5 fps with prediction, the corresponding values were 2.0/2.6, 2.2/3.3, and 2.4/3.9 mm, respectively. There was no significant difference between the irradiation efficiency with and that without prediction of the marker position. The geometrical accuracy at lower frame rates with prediction applied was superior or comparable to that at 15 fps without prediction. In comparison with the current clinical setting for real-time image-guided radiation therapy, which uses an imaging rate of 15.0 fps without prediction, it may be possible to reduce the imaging dose by half or more.</p></div><div><h3>Conclusions</h3><p>Motion prediction can effectively lower the frame rate and minimize the imaging dose in real-time image-guided radiation therapy.</p></div>","PeriodicalId":56092,"journal":{"name":"Physica Medica-European Journal of Medical Physics","volume":null,"pages":null},"PeriodicalIF":3.3000,"publicationDate":"2024-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Application of motion prediction based on a long short-term memory network for imaging dose reduction in real-time tumor-tracking radiation therapy\",\"authors\":\"\",\"doi\":\"10.1016/j.ejmp.2024.104507\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Purpose</h3><p>To demonstrate the possibility of using a lower imaging rate while maintaining acceptable accuracy by applying motion prediction to minimize the imaging dose in real-time image-guided radiation therapy.</p></div><div><h3>Methods</h3><p>Time-series of three-dimensional internal marker positions obtained from 98 patients in liver stereotactic body radiation therapy were used to train and test the long-short-term memory (LSTM) network. For real-time imaging, the root mean squared error (RMSE) of the prediction on three-dimensional marker position made by LSTM, the residual motion of the target under respiratory-gated irradiation, and irradiation efficiency were evaluated. In the evaluation of the residual motion, the system-specific latency was assumed to be 100 ms.</p></div><div><h3>Results</h3><p>Except for outliers in the superior–inferior (SI) direction, the median/maximum values of the RMSE for imaging rates of 7.5, 5.0, and 2.5 frames per second (fps) were 0.8/1.3, 0.9/1.6, and 1.2/2.4 mm, respectively. The median/maximum residual motion in the SI direction at an imaging rate of 15.0 fps without prediction of the marker position, which is a typical clinical setting, was 2.3/3.6 mm. For rates of 7.5, 5.0, and 2.5 fps with prediction, the corresponding values were 2.0/2.6, 2.2/3.3, and 2.4/3.9 mm, respectively. There was no significant difference between the irradiation efficiency with and that without prediction of the marker position. The geometrical accuracy at lower frame rates with prediction applied was superior or comparable to that at 15 fps without prediction. In comparison with the current clinical setting for real-time image-guided radiation therapy, which uses an imaging rate of 15.0 fps without prediction, it may be possible to reduce the imaging dose by half or more.</p></div><div><h3>Conclusions</h3><p>Motion prediction can effectively lower the frame rate and minimize the imaging dose in real-time image-guided radiation therapy.</p></div>\",\"PeriodicalId\":56092,\"journal\":{\"name\":\"Physica Medica-European Journal of Medical Physics\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":3.3000,\"publicationDate\":\"2024-09-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Physica Medica-European Journal of Medical Physics\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S1120179724002527\",\"RegionNum\":3,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q1\",\"JCRName\":\"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Physica Medica-European Journal of Medical Physics","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S1120179724002527","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

摘要

方法用98名肝脏立体定向体放射治疗患者的三维内部标记位置时间序列来训练和测试长短期记忆(LSTM)网络。在实时成像中,评估了 LSTM 对三维标记位置预测的均方根误差(RMSE)、呼吸门控照射下目标的残余运动以及照射效率。结果除上-下(SI)方向的异常值外,成像速率为 7.5、5.0 和 2.5 帧/秒(fps)时,RMSE 的中值/最大值分别为 0.8/1.3、0.9/1.6 和 1.2/2.4mm。在不预测标记物位置(这是典型的临床情况)的情况下,成像速度为 15.0 帧/秒时,SI 方向的残余运动中位数/最大值为 2.3/3.6 毫米。带预测功能的 7.5、5.0 和 2.5 fps 成像速度下,相应的数值分别为 2.0/2.6、2.2/3.3 和 2.4/3.9 毫米。有标记位置预测和无标记位置预测时的照射效率没有明显差异。在较低帧频下,应用预测的几何精确度优于或相当于未应用预测的 15 帧/秒。目前临床上使用的实时图像引导放射治疗的成像速率为 15.0 fps,而没有进行预测,与之相比,成像剂量有可能减少一半或更多。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Application of motion prediction based on a long short-term memory network for imaging dose reduction in real-time tumor-tracking radiation therapy

Purpose

To demonstrate the possibility of using a lower imaging rate while maintaining acceptable accuracy by applying motion prediction to minimize the imaging dose in real-time image-guided radiation therapy.

Methods

Time-series of three-dimensional internal marker positions obtained from 98 patients in liver stereotactic body radiation therapy were used to train and test the long-short-term memory (LSTM) network. For real-time imaging, the root mean squared error (RMSE) of the prediction on three-dimensional marker position made by LSTM, the residual motion of the target under respiratory-gated irradiation, and irradiation efficiency were evaluated. In the evaluation of the residual motion, the system-specific latency was assumed to be 100 ms.

Results

Except for outliers in the superior–inferior (SI) direction, the median/maximum values of the RMSE for imaging rates of 7.5, 5.0, and 2.5 frames per second (fps) were 0.8/1.3, 0.9/1.6, and 1.2/2.4 mm, respectively. The median/maximum residual motion in the SI direction at an imaging rate of 15.0 fps without prediction of the marker position, which is a typical clinical setting, was 2.3/3.6 mm. For rates of 7.5, 5.0, and 2.5 fps with prediction, the corresponding values were 2.0/2.6, 2.2/3.3, and 2.4/3.9 mm, respectively. There was no significant difference between the irradiation efficiency with and that without prediction of the marker position. The geometrical accuracy at lower frame rates with prediction applied was superior or comparable to that at 15 fps without prediction. In comparison with the current clinical setting for real-time image-guided radiation therapy, which uses an imaging rate of 15.0 fps without prediction, it may be possible to reduce the imaging dose by half or more.

Conclusions

Motion prediction can effectively lower the frame rate and minimize the imaging dose in real-time image-guided radiation therapy.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
6.80
自引率
14.70%
发文量
493
审稿时长
78 days
期刊介绍: Physica Medica, European Journal of Medical Physics, publishing with Elsevier from 2007, provides an international forum for research and reviews on the following main topics: Medical Imaging Radiation Therapy Radiation Protection Measuring Systems and Signal Processing Education and training in Medical Physics Professional issues in Medical Physics.
期刊最新文献
Comparison of two biologically effective dose calculation models applied to single fraction stereotactic radiosurgery Advanced mathematical modeling for preciseestimation of CT energy spectrum using a calibration phantom LatticeOpt: An automatic tool for planning optimisation of spatially fractionated stereotactic body radiotherapy Development of a Monte Carlo simulation platform for the systematic evaluation of photon-counting detector-based micro-CT Investigate potential clinical benefits and linear energy transfer sparing utilizing proton arc therapy for hepatocellular carcinoma
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1