{"title":"[在髋关节 X 射线照相术中添加铜滤波器的效用和降低剂量的潜力]。","authors":"Tomoyoshi Kawabata, Shoko Kamiyama, Shu Onodera","doi":"10.6009/jjrt.2024-1445","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>In this study, we evaluated image quality and radiation dose reduction when a Copper (Cu) filter was added to hip joint X-ray imaging.</p><p><strong>Methods: </strong>We measured effective energy without (0 mm) and with (0.1/0.2 mm) Cu-added filter at 70 kV, and we calculated soft tissue-bone contrast and signal-difference-to-noise-ratio (SDNR) under constant entrance surface dose. After that, we estimated the dose reduction rate.</p><p><strong>Results: </strong>The effective energy was 32.07 keV for 0 mm Cu, 37.59 keV for 0.1 mm Cu, and 40.91 keV for 0.2 mm Cu. As the thickness of the Cu-added filter was increased, contrast decreased, but SDNR increased. The dose reduction rate in bone calculated measuring SDNR was 34% for 0.1 mm Cu and 47% for 0.2 mm Cu in max.</p><p><strong>Conclusion: </strong>It was suggested that adding Cu filter to hip-joint X-ray imaging could reduce entrance surface dose while maintaining the image quality based on SDNR.</p>","PeriodicalId":74309,"journal":{"name":"Nihon Hoshasen Gijutsu Gakkai zasshi","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2024-10-20","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"[Usefulness of Copper Filter Addition and Potential for Dose Reduction in Hip-joint Radiography].\",\"authors\":\"Tomoyoshi Kawabata, Shoko Kamiyama, Shu Onodera\",\"doi\":\"10.6009/jjrt.2024-1445\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Purpose: </strong>In this study, we evaluated image quality and radiation dose reduction when a Copper (Cu) filter was added to hip joint X-ray imaging.</p><p><strong>Methods: </strong>We measured effective energy without (0 mm) and with (0.1/0.2 mm) Cu-added filter at 70 kV, and we calculated soft tissue-bone contrast and signal-difference-to-noise-ratio (SDNR) under constant entrance surface dose. After that, we estimated the dose reduction rate.</p><p><strong>Results: </strong>The effective energy was 32.07 keV for 0 mm Cu, 37.59 keV for 0.1 mm Cu, and 40.91 keV for 0.2 mm Cu. As the thickness of the Cu-added filter was increased, contrast decreased, but SDNR increased. The dose reduction rate in bone calculated measuring SDNR was 34% for 0.1 mm Cu and 47% for 0.2 mm Cu in max.</p><p><strong>Conclusion: </strong>It was suggested that adding Cu filter to hip-joint X-ray imaging could reduce entrance surface dose while maintaining the image quality based on SDNR.</p>\",\"PeriodicalId\":74309,\"journal\":{\"name\":\"Nihon Hoshasen Gijutsu Gakkai zasshi\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2024-10-20\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Nihon Hoshasen Gijutsu Gakkai zasshi\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.6009/jjrt.2024-1445\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2024/9/7 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Nihon Hoshasen Gijutsu Gakkai zasshi","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.6009/jjrt.2024-1445","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/9/7 0:00:00","PubModel":"Epub","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
摘要
目的:在这项研究中,我们评估了在髋关节 X 射线成像中添加铜(Cu)滤光片后的图像质量和辐射剂量降低情况:方法:我们测量了在 70 千伏电压下未添加(0 毫米)和添加(0.1/0.2 毫米)铜滤光片时的有效能量,并计算了在入口表面剂量恒定的情况下软组织-骨骼对比度和信号差-噪声比(SDNR)。之后,我们估算了剂量减少率:结果:0 毫米铜的有效能量为 32.07 keV,0.1 毫米铜的有效能量为 37.59 keV,0.2 毫米铜的有效能量为 40.91 keV。随着加铜滤光片厚度的增加,对比度降低,但 SDNR 增加。通过测量 SDNR 计算出的骨骼剂量减少率,最大值为 0.1 mm Cu 34%,0.2 mm Cu 47%:结论:在髋关节 X 射线成像中加入铜滤光片可减少入口表面剂量,同时根据 SDNR 保持图像质量。
[Usefulness of Copper Filter Addition and Potential for Dose Reduction in Hip-joint Radiography].
Purpose: In this study, we evaluated image quality and radiation dose reduction when a Copper (Cu) filter was added to hip joint X-ray imaging.
Methods: We measured effective energy without (0 mm) and with (0.1/0.2 mm) Cu-added filter at 70 kV, and we calculated soft tissue-bone contrast and signal-difference-to-noise-ratio (SDNR) under constant entrance surface dose. After that, we estimated the dose reduction rate.
Results: The effective energy was 32.07 keV for 0 mm Cu, 37.59 keV for 0.1 mm Cu, and 40.91 keV for 0.2 mm Cu. As the thickness of the Cu-added filter was increased, contrast decreased, but SDNR increased. The dose reduction rate in bone calculated measuring SDNR was 34% for 0.1 mm Cu and 47% for 0.2 mm Cu in max.
Conclusion: It was suggested that adding Cu filter to hip-joint X-ray imaging could reduce entrance surface dose while maintaining the image quality based on SDNR.