{"title":"利用激光和雷达应用无创采集麻醉大鼠的生命数据","authors":"Toshiaki Kawabe, Shota Kita, Isao Ohmura, Ryuji Michino, Hidenori Watanabe, Guanghao Sun, Seiya Inoue","doi":"10.1177/00236772241265541","DOIUrl":null,"url":null,"abstract":"The aim of this study was to verify the possibility of obtaining vital sign information using a laser and radar sensor in a manner that is non-invasive and painless for test animals. A dataset was obtained from respiratory movement of anaesthetized male F344 rats, signals of laser and radar sensors were recorded simultaneously with vital data acquired with an integrated multiple-channel intraoperative monitor. In addition, respiratory movements were also video recorded, and used as reference data of respiration rate (RR; ref-RR). Reference data for heart rate (HR; ref-HR) were obtained from the R wave of electrocardiogram data for each epoch. Signals recorded from the radar sensor (I- and Q-signals) were input to a computer, and HR (radar-HR) and RR (radar-RR) were estimated using the frequency analysis method. Among the six positions where respiratory movements were measured by the laser sensor, the number of peak counts matched the visual counts of respiratory movements in the video records. The respiratory movements were significantly the greatest over the most caudal rib in the dorsal ( p < 0.001). The average radar-RR and ref-RR values showed correspondence (ref-RR, 69 ± 6.2 breaths/min; radar-RR, 68 ± 5.7 breaths/min ( p = 0.04–1.00); equivalence ratio, 86%). The radar-HR data showed slight variability; however, there was 80% homology compared with the ref-HR values (ref-HR, 336 ± 19.6 beats/min; radar-HR, 348 ± 34.1 ( p = 0.10–0.95)). Although comparison of the data under noradrenaline administration failed to track drug-induced changes in some cases, the HR and RR data of anesthetized rats measured from the radar sensor system showed comparable accuracy to other conventional methods.","PeriodicalId":18013,"journal":{"name":"Laboratory Animals","volume":null,"pages":null},"PeriodicalIF":1.3000,"publicationDate":"2024-09-11","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Non-invasive acquisition of vital data in anesthetized rats using laser and radar application\",\"authors\":\"Toshiaki Kawabe, Shota Kita, Isao Ohmura, Ryuji Michino, Hidenori Watanabe, Guanghao Sun, Seiya Inoue\",\"doi\":\"10.1177/00236772241265541\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"The aim of this study was to verify the possibility of obtaining vital sign information using a laser and radar sensor in a manner that is non-invasive and painless for test animals. A dataset was obtained from respiratory movement of anaesthetized male F344 rats, signals of laser and radar sensors were recorded simultaneously with vital data acquired with an integrated multiple-channel intraoperative monitor. In addition, respiratory movements were also video recorded, and used as reference data of respiration rate (RR; ref-RR). Reference data for heart rate (HR; ref-HR) were obtained from the R wave of electrocardiogram data for each epoch. Signals recorded from the radar sensor (I- and Q-signals) were input to a computer, and HR (radar-HR) and RR (radar-RR) were estimated using the frequency analysis method. Among the six positions where respiratory movements were measured by the laser sensor, the number of peak counts matched the visual counts of respiratory movements in the video records. The respiratory movements were significantly the greatest over the most caudal rib in the dorsal ( p < 0.001). The average radar-RR and ref-RR values showed correspondence (ref-RR, 69 ± 6.2 breaths/min; radar-RR, 68 ± 5.7 breaths/min ( p = 0.04–1.00); equivalence ratio, 86%). The radar-HR data showed slight variability; however, there was 80% homology compared with the ref-HR values (ref-HR, 336 ± 19.6 beats/min; radar-HR, 348 ± 34.1 ( p = 0.10–0.95)). Although comparison of the data under noradrenaline administration failed to track drug-induced changes in some cases, the HR and RR data of anesthetized rats measured from the radar sensor system showed comparable accuracy to other conventional methods.\",\"PeriodicalId\":18013,\"journal\":{\"name\":\"Laboratory Animals\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":1.3000,\"publicationDate\":\"2024-09-11\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Laboratory Animals\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1177/00236772241265541\",\"RegionNum\":4,\"RegionCategory\":\"农林科学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"VETERINARY SCIENCES\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Laboratory Animals","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1177/00236772241265541","RegionNum":4,"RegionCategory":"农林科学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"VETERINARY SCIENCES","Score":null,"Total":0}
Non-invasive acquisition of vital data in anesthetized rats using laser and radar application
The aim of this study was to verify the possibility of obtaining vital sign information using a laser and radar sensor in a manner that is non-invasive and painless for test animals. A dataset was obtained from respiratory movement of anaesthetized male F344 rats, signals of laser and radar sensors were recorded simultaneously with vital data acquired with an integrated multiple-channel intraoperative monitor. In addition, respiratory movements were also video recorded, and used as reference data of respiration rate (RR; ref-RR). Reference data for heart rate (HR; ref-HR) were obtained from the R wave of electrocardiogram data for each epoch. Signals recorded from the radar sensor (I- and Q-signals) were input to a computer, and HR (radar-HR) and RR (radar-RR) were estimated using the frequency analysis method. Among the six positions where respiratory movements were measured by the laser sensor, the number of peak counts matched the visual counts of respiratory movements in the video records. The respiratory movements were significantly the greatest over the most caudal rib in the dorsal ( p < 0.001). The average radar-RR and ref-RR values showed correspondence (ref-RR, 69 ± 6.2 breaths/min; radar-RR, 68 ± 5.7 breaths/min ( p = 0.04–1.00); equivalence ratio, 86%). The radar-HR data showed slight variability; however, there was 80% homology compared with the ref-HR values (ref-HR, 336 ± 19.6 beats/min; radar-HR, 348 ± 34.1 ( p = 0.10–0.95)). Although comparison of the data under noradrenaline administration failed to track drug-induced changes in some cases, the HR and RR data of anesthetized rats measured from the radar sensor system showed comparable accuracy to other conventional methods.
期刊介绍:
The international journal of laboratory animal science and welfare, Laboratory Animals publishes peer-reviewed original papers and reviews on all aspects of the use of animals in biomedical research. The journal promotes improvements in the welfare or well-being of the animals used, it particularly focuses on research that reduces the number of animals used or which replaces animal models with in vitro alternatives.