Jae Seong Kim, Jingyeong Kim, Jae-Seok Kim, Wooseong Kim, Chang-Soo Lee
{"title":"利用浓度梯度生成技术在液滴中进行无标记单细胞抗菌药物敏感性测试","authors":"Jae Seong Kim, Jingyeong Kim, Jae-Seok Kim, Wooseong Kim, Chang-Soo Lee","doi":"10.1039/d4lc00629a","DOIUrl":null,"url":null,"abstract":"Bacterial communities exhibit significant heterogeneity, resulting in the emergence of specialized phenotypes that can withstand antibiotic exposure. Unfortunately, the existence of subpopulations resistant to antibiotics often goes unnoticed during treatment initiation. Thus, it is crucial to consider the concept of single-cell antibiotic susceptibility testing (AST) to tackle bacterial infections. Nevertheless, its practical application in clinical settings is hindered by its inability to conduct AST efficiently across a wide range of antibiotics and concentrations. This study introduces a droplet-based microfluidic platform designed for rapid single-cell AST by creating an antibiotic concentration gradient. The advantage of a microfluidic platform is achieved by executing bacteria and antibiotic mixing, cell encapsulation, incubation, and enumeration of bacteria in a seamless workflow, facilitating susceptibility testing of each antibiotic. Firstly, we demonstrate the rapid determination of minimum inhibitory concentration (MIC) of several antibiotics with Gram-negative E. coli and Gram-negative S. aureus, which enables us to bypass the time-consuming bacteria cultivation, speeding up the AST in 3 hrs from 1 – 2 days of conventional methods. Additionally, we assess 10 clinical isolates including methicillin-resistant Staphylococcus aureus (MRSA) and multidrug-resistant Staphylococcus aureus (MDRSA) against clinically important antibiotics for analyzing MIC, compared to the gold standard AST method from the United States Clinical and Laboratory Standards Institute (CLSI), which becomes available only after 48 h. Furthermore, by monitoring single cells within individual droplets, we have found a spectrum of resistance levels among genetically identical cells, revealing phenotypic heterogeneity within isogenic populations. This discovery not only advances clinical diagnostics and treatment strategies but also significantly contributes to the field of antibiotic stewardship, underlining the importance of our approach in addressing bacterial resistance.","PeriodicalId":85,"journal":{"name":"Lab on a Chip","volume":null,"pages":null},"PeriodicalIF":6.1000,"publicationDate":"2024-09-20","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Label-free single-cell antimicrobial susceptibility testing in droplets with concentration gradient generation\",\"authors\":\"Jae Seong Kim, Jingyeong Kim, Jae-Seok Kim, Wooseong Kim, Chang-Soo Lee\",\"doi\":\"10.1039/d4lc00629a\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Bacterial communities exhibit significant heterogeneity, resulting in the emergence of specialized phenotypes that can withstand antibiotic exposure. Unfortunately, the existence of subpopulations resistant to antibiotics often goes unnoticed during treatment initiation. Thus, it is crucial to consider the concept of single-cell antibiotic susceptibility testing (AST) to tackle bacterial infections. Nevertheless, its practical application in clinical settings is hindered by its inability to conduct AST efficiently across a wide range of antibiotics and concentrations. This study introduces a droplet-based microfluidic platform designed for rapid single-cell AST by creating an antibiotic concentration gradient. The advantage of a microfluidic platform is achieved by executing bacteria and antibiotic mixing, cell encapsulation, incubation, and enumeration of bacteria in a seamless workflow, facilitating susceptibility testing of each antibiotic. Firstly, we demonstrate the rapid determination of minimum inhibitory concentration (MIC) of several antibiotics with Gram-negative E. coli and Gram-negative S. aureus, which enables us to bypass the time-consuming bacteria cultivation, speeding up the AST in 3 hrs from 1 – 2 days of conventional methods. Additionally, we assess 10 clinical isolates including methicillin-resistant Staphylococcus aureus (MRSA) and multidrug-resistant Staphylococcus aureus (MDRSA) against clinically important antibiotics for analyzing MIC, compared to the gold standard AST method from the United States Clinical and Laboratory Standards Institute (CLSI), which becomes available only after 48 h. Furthermore, by monitoring single cells within individual droplets, we have found a spectrum of resistance levels among genetically identical cells, revealing phenotypic heterogeneity within isogenic populations. This discovery not only advances clinical diagnostics and treatment strategies but also significantly contributes to the field of antibiotic stewardship, underlining the importance of our approach in addressing bacterial resistance.\",\"PeriodicalId\":85,\"journal\":{\"name\":\"Lab on a Chip\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":6.1000,\"publicationDate\":\"2024-09-20\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Lab on a Chip\",\"FirstCategoryId\":\"5\",\"ListUrlMain\":\"https://doi.org/10.1039/d4lc00629a\",\"RegionNum\":2,\"RegionCategory\":\"工程技术\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q1\",\"JCRName\":\"BIOCHEMICAL RESEARCH METHODS\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Lab on a Chip","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1039/d4lc00629a","RegionNum":2,"RegionCategory":"工程技术","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"BIOCHEMICAL RESEARCH METHODS","Score":null,"Total":0}
Label-free single-cell antimicrobial susceptibility testing in droplets with concentration gradient generation
Bacterial communities exhibit significant heterogeneity, resulting in the emergence of specialized phenotypes that can withstand antibiotic exposure. Unfortunately, the existence of subpopulations resistant to antibiotics often goes unnoticed during treatment initiation. Thus, it is crucial to consider the concept of single-cell antibiotic susceptibility testing (AST) to tackle bacterial infections. Nevertheless, its practical application in clinical settings is hindered by its inability to conduct AST efficiently across a wide range of antibiotics and concentrations. This study introduces a droplet-based microfluidic platform designed for rapid single-cell AST by creating an antibiotic concentration gradient. The advantage of a microfluidic platform is achieved by executing bacteria and antibiotic mixing, cell encapsulation, incubation, and enumeration of bacteria in a seamless workflow, facilitating susceptibility testing of each antibiotic. Firstly, we demonstrate the rapid determination of minimum inhibitory concentration (MIC) of several antibiotics with Gram-negative E. coli and Gram-negative S. aureus, which enables us to bypass the time-consuming bacteria cultivation, speeding up the AST in 3 hrs from 1 – 2 days of conventional methods. Additionally, we assess 10 clinical isolates including methicillin-resistant Staphylococcus aureus (MRSA) and multidrug-resistant Staphylococcus aureus (MDRSA) against clinically important antibiotics for analyzing MIC, compared to the gold standard AST method from the United States Clinical and Laboratory Standards Institute (CLSI), which becomes available only after 48 h. Furthermore, by monitoring single cells within individual droplets, we have found a spectrum of resistance levels among genetically identical cells, revealing phenotypic heterogeneity within isogenic populations. This discovery not only advances clinical diagnostics and treatment strategies but also significantly contributes to the field of antibiotic stewardship, underlining the importance of our approach in addressing bacterial resistance.
期刊介绍:
Lab on a Chip is the premiere journal that publishes cutting-edge research in the field of miniaturization. By their very nature, microfluidic/nanofluidic/miniaturized systems are at the intersection of disciplines, spanning fundamental research to high-end application, which is reflected by the broad readership of the journal. Lab on a Chip publishes two types of papers on original research: full-length research papers and communications. Papers should demonstrate innovations, which can come from technical advancements or applications addressing pressing needs in globally important areas. The journal also publishes Comments, Reviews, and Perspectives.