根据 ISO 81060-2:2018 + Amd.1:2020 协议对 TMB-2285-BT 普通人群腕式血压计进行验证。

IF 1.2 4区 医学 Q4 PERIPHERAL VASCULAR DISEASE Blood Pressure Monitoring Pub Date : 2024-09-13 DOI:10.1097/MBP.0000000000000724
Zijian Xie, Bin Peng, Jia Hu, Xinda Wang, Xiaoqin Du, Chaoya Li, Jiahui Liang
{"title":"根据 ISO 81060-2:2018 + Amd.1:2020 协议对 TMB-2285-BT 普通人群腕式血压计进行验证。","authors":"Zijian Xie, Bin Peng, Jia Hu, Xinda Wang, Xiaoqin Du, Chaoya Li, Jiahui Liang","doi":"10.1097/MBP.0000000000000724","DOIUrl":null,"url":null,"abstract":"<p><p>The objective of this study is to evaluate the accuracy of the wrist blood pressure (BP) monitor (TMB-2285-BT) in general population according to international standard of ISO 81060-2:2018+Amd.1:2020. The TMB-2285-BT BP monitor is an oscillometric device measuring BP from wrist. A single cuff for wrist sizes of 13.5-21.5 cm was used for the test device. The study was performed according to the 'International Standard (ISO 81060-2:2018 + Amd.1:2020) protocol' using the same-arm sequential BP measurement method. Subjects (n ≥ 85) fulfilling the age, gender, BP, and cuff distribution criteria of the protocol were enrolled. For criterion 1, the mean difference of SBP and DBP between the test device and the reference device was -0.27 mmHg and -0.80 mmHg, respectively. For criterion 2, the SD of the averaged paired determinations of SBP and DBP for the test device and the reference device was 2.60 and 2.06 mmHg, respectively, both of which were less than 6.95 and 6.89 mmHg. The accuracy of the test device fulfilled the requirements of the criterion 1 and 2 of ISO 81060-2:2018 + Amd.1:2020. The TMB-2285-BT wrist BP monitor can be recommended for home BP measurement for general population.</p>","PeriodicalId":8950,"journal":{"name":"Blood Pressure Monitoring","volume":" ","pages":""},"PeriodicalIF":1.2000,"publicationDate":"2024-09-13","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Validation of TMB-2285-BT wrist blood pressure monitor in general population according to the ISO 81060-2:2018 + Amd.1:2020 Protocol.\",\"authors\":\"Zijian Xie, Bin Peng, Jia Hu, Xinda Wang, Xiaoqin Du, Chaoya Li, Jiahui Liang\",\"doi\":\"10.1097/MBP.0000000000000724\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>The objective of this study is to evaluate the accuracy of the wrist blood pressure (BP) monitor (TMB-2285-BT) in general population according to international standard of ISO 81060-2:2018+Amd.1:2020. The TMB-2285-BT BP monitor is an oscillometric device measuring BP from wrist. A single cuff for wrist sizes of 13.5-21.5 cm was used for the test device. The study was performed according to the 'International Standard (ISO 81060-2:2018 + Amd.1:2020) protocol' using the same-arm sequential BP measurement method. Subjects (n ≥ 85) fulfilling the age, gender, BP, and cuff distribution criteria of the protocol were enrolled. For criterion 1, the mean difference of SBP and DBP between the test device and the reference device was -0.27 mmHg and -0.80 mmHg, respectively. For criterion 2, the SD of the averaged paired determinations of SBP and DBP for the test device and the reference device was 2.60 and 2.06 mmHg, respectively, both of which were less than 6.95 and 6.89 mmHg. The accuracy of the test device fulfilled the requirements of the criterion 1 and 2 of ISO 81060-2:2018 + Amd.1:2020. The TMB-2285-BT wrist BP monitor can be recommended for home BP measurement for general population.</p>\",\"PeriodicalId\":8950,\"journal\":{\"name\":\"Blood Pressure Monitoring\",\"volume\":\" \",\"pages\":\"\"},\"PeriodicalIF\":1.2000,\"publicationDate\":\"2024-09-13\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Blood Pressure Monitoring\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1097/MBP.0000000000000724\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"PERIPHERAL VASCULAR DISEASE\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Blood Pressure Monitoring","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/MBP.0000000000000724","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"PERIPHERAL VASCULAR DISEASE","Score":null,"Total":0}
引用次数: 0

摘要

本研究的目的是根据 ISO 81060-2:2018+Amd.1:2020 国际标准,评估腕式血压计(TMB-2285-BT)在普通人群中的准确性。TMB-2285-BT 血压计是一种从手腕测量血压的示波设备。测试设备使用的单个袖带适用于 13.5-21.5 厘米的手腕尺寸。研究根据 "国际标准(ISO 81060-2:2018 + Amd.1:2020)协议 "进行,采用同臂顺序血压测量方法。符合方案中年龄、性别、血压和袖带分布标准的受试者(n ≥ 85)被纳入研究。在标准 1 中,测试设备与参照设备之间的 SBP 和 DBP 平均差分别为 -0.27 mmHg 和 -0.80 mmHg。就标准 2 而言,测试设备和参照设备的 SBP 和 DBP 成对测定值平均值的 SD 分别为 2.60 和 2.06 mmHg,均小于 6.95 和 6.89 mmHg。测试设备的准确度符合 ISO 81060-2:2018 + Amd.1:2020 标准 1 和 2 的要求。TMB-2285-BT 腕式血压计可推荐用于普通人群的家庭血压测量。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Validation of TMB-2285-BT wrist blood pressure monitor in general population according to the ISO 81060-2:2018 + Amd.1:2020 Protocol.

The objective of this study is to evaluate the accuracy of the wrist blood pressure (BP) monitor (TMB-2285-BT) in general population according to international standard of ISO 81060-2:2018+Amd.1:2020. The TMB-2285-BT BP monitor is an oscillometric device measuring BP from wrist. A single cuff for wrist sizes of 13.5-21.5 cm was used for the test device. The study was performed according to the 'International Standard (ISO 81060-2:2018 + Amd.1:2020) protocol' using the same-arm sequential BP measurement method. Subjects (n ≥ 85) fulfilling the age, gender, BP, and cuff distribution criteria of the protocol were enrolled. For criterion 1, the mean difference of SBP and DBP between the test device and the reference device was -0.27 mmHg and -0.80 mmHg, respectively. For criterion 2, the SD of the averaged paired determinations of SBP and DBP for the test device and the reference device was 2.60 and 2.06 mmHg, respectively, both of which were less than 6.95 and 6.89 mmHg. The accuracy of the test device fulfilled the requirements of the criterion 1 and 2 of ISO 81060-2:2018 + Amd.1:2020. The TMB-2285-BT wrist BP monitor can be recommended for home BP measurement for general population.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Blood Pressure Monitoring
Blood Pressure Monitoring 医学-外周血管病
CiteScore
2.00
自引率
7.70%
发文量
110
审稿时长
>12 weeks
期刊介绍: Blood Pressure Monitoring is devoted to original research in blood pressure measurement and blood pressure variability. It includes device technology, analytical methodology of blood pressure over time and its variability, clinical trials - including, but not limited to, pharmacology - involving blood pressure monitoring, blood pressure reactivity, patient evaluation, and outcomes and effectiveness research. This innovative journal contains papers dealing with all aspects of manual, automated, and ambulatory monitoring. Basic and clinical science papers are considered although the emphasis is on clinical medicine. Submitted articles undergo a preliminary review by the editor. Some articles may be returned to authors without further consideration. Those being considered for publication will undergo further assessment and peer-review by the editors and those invited to do so from a reviewer pool.
期刊最新文献
Could soluble ST2 levels be used to detect nondipper hypertensive subgroup in newly diagnosed hypertension patients. Validation of two automatic sphygmomanometers according to the modified International Standardization Organization 81060-2:2018 protocol in adults with a mid-upper arm circumference of 22 centimeters or less. Validation of the validity of the new wrist-type fully automatic blood pressure monitor: DBP-2242 according to ISO 81060-2:2018+AMD 1:2020. Psychological distress, forced awakening, and morning blood pressure surge. Validation of the AViTA upper-arm blood pressure monitor BP636 with multiple cuff sizes in the general population according to the ANSI/AAMI/ISO 81060-2:2018/Amd1:2020.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1