利用碰撞池技术和微滴校准推进元素分析,通过 LA-ICP-TOFMS 实现生物成像应用

IF 5.7 2区 化学 Q1 CHEMISTRY, ANALYTICAL Analytica Chimica Acta Pub Date : 2024-10-19 DOI:10.1016/j.aca.2024.343345
Sarah Theiner , Elisabeth Foels , Gunda Koellensperger
{"title":"利用碰撞池技术和微滴校准推进元素分析,通过 LA-ICP-TOFMS 实现生物成像应用","authors":"Sarah Theiner ,&nbsp;Elisabeth Foels ,&nbsp;Gunda Koellensperger","doi":"10.1016/j.aca.2024.343345","DOIUrl":null,"url":null,"abstract":"<div><h3>Background</h3><div>Bioimaging applications by laser ablation inductively coupled plasma time-of-flight mass spectrometry (LA-ICP-TOFMS) require comprehensive elemental analysis spanning the entire mass range. Within biological systems, endogenous elements are critical for maintaining metal homeostasis in cells, a fundamental aspect in disease progression when disrupted. Additionally, elements from the higher mass range may originate from various sources such as metal-tags for immuno-mass spectrometry, nanoparticles, or metal-based anticancer drugs. This study assesses the efficacy of collision/reacton cell (CCT) mode for simultaneously analysing elements across the complete mass spectrum. Furthermore, we demonstrate the accuracy of the LA-ICP-TOFMS methodology through the analysis of serum reference material.</div></div><div><h3>Results</h3><div>Our findings demonstrate that the CCT mode outperforms the standard/no gas mode for LA-ICP-TOFMS measurements, particularly in quantifying endogenous elements susceptible to interferences. Through the analysis of picolitre-volume micro-droplet standards and serum reference material (deposited as micro-droplets), we observed accurate quantification of elements such as iron and selenium, with isotope ratios closely resembling natural compositions. As key advantage, the utilization of the CCT mode eliminated the need for labor-intensive post-data processing, streamlining analytical procedures. Additionally, the CCT mode also provided enhanced sensitivity (factor of 1.5–2) for elements in the higher mass range compared to the standard mode, without compromising the sensitivity for endogenous elements.</div></div><div><h3>Significance</h3><div>We successfully integrated Seronorm serum reference material-containing micro-droplets into LA-ICPMS bioimaging workflows for quality control, enabling validated measurements of key biological elements. The use of the CCT mode is highly recommended for bioimaging applications that address the analysis of elements across the entire mass range.</div></div>","PeriodicalId":240,"journal":{"name":"Analytica Chimica Acta","volume":"1332 ","pages":"Article 343345"},"PeriodicalIF":5.7000,"publicationDate":"2024-10-19","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Advancing elemental analysis by collision/reaction cell technology and micro-droplet calibration for bioimaging applications by LA-ICP-TOFMS\",\"authors\":\"Sarah Theiner ,&nbsp;Elisabeth Foels ,&nbsp;Gunda Koellensperger\",\"doi\":\"10.1016/j.aca.2024.343345\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Background</h3><div>Bioimaging applications by laser ablation inductively coupled plasma time-of-flight mass spectrometry (LA-ICP-TOFMS) require comprehensive elemental analysis spanning the entire mass range. Within biological systems, endogenous elements are critical for maintaining metal homeostasis in cells, a fundamental aspect in disease progression when disrupted. Additionally, elements from the higher mass range may originate from various sources such as metal-tags for immuno-mass spectrometry, nanoparticles, or metal-based anticancer drugs. This study assesses the efficacy of collision/reacton cell (CCT) mode for simultaneously analysing elements across the complete mass spectrum. Furthermore, we demonstrate the accuracy of the LA-ICP-TOFMS methodology through the analysis of serum reference material.</div></div><div><h3>Results</h3><div>Our findings demonstrate that the CCT mode outperforms the standard/no gas mode for LA-ICP-TOFMS measurements, particularly in quantifying endogenous elements susceptible to interferences. Through the analysis of picolitre-volume micro-droplet standards and serum reference material (deposited as micro-droplets), we observed accurate quantification of elements such as iron and selenium, with isotope ratios closely resembling natural compositions. As key advantage, the utilization of the CCT mode eliminated the need for labor-intensive post-data processing, streamlining analytical procedures. Additionally, the CCT mode also provided enhanced sensitivity (factor of 1.5–2) for elements in the higher mass range compared to the standard mode, without compromising the sensitivity for endogenous elements.</div></div><div><h3>Significance</h3><div>We successfully integrated Seronorm serum reference material-containing micro-droplets into LA-ICPMS bioimaging workflows for quality control, enabling validated measurements of key biological elements. The use of the CCT mode is highly recommended for bioimaging applications that address the analysis of elements across the entire mass range.</div></div>\",\"PeriodicalId\":240,\"journal\":{\"name\":\"Analytica Chimica Acta\",\"volume\":\"1332 \",\"pages\":\"Article 343345\"},\"PeriodicalIF\":5.7000,\"publicationDate\":\"2024-10-19\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Analytica Chimica Acta\",\"FirstCategoryId\":\"92\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S0003267024011462\",\"RegionNum\":2,\"RegionCategory\":\"化学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q1\",\"JCRName\":\"CHEMISTRY, ANALYTICAL\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Analytica Chimica Acta","FirstCategoryId":"92","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0003267024011462","RegionNum":2,"RegionCategory":"化学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"CHEMISTRY, ANALYTICAL","Score":null,"Total":0}
引用次数: 0

摘要

背景使用激光烧蚀电感耦合等离子体飞行时间质谱(LA-ICP-TOFMS)进行生物成像应用需要在整个质量范围内进行全面的元素分析。在生物系统中,内源元素对维持细胞中的金属平衡至关重要,一旦受到破坏,就会成为疾病发展的一个基本方面。此外,更高质 量范围的元素可能来自各种来源,如用于免疫质谱分析的金属标签、纳米粒子或基于金属的抗癌药物。本研究评估了碰撞/反应池(CCT)模式同时分析整个质谱中元素的功效。结果我们的研究结果表明,在 LA-ICP-TOFMS 测量中,CCT 模式优于标准/无气模式,尤其是在定量易受干扰的内源元素方面。通过分析皮升体积的微滴标准物质和血清参考物质(沉积为微滴),我们观察到了铁和硒等元素的精确定量,其同位素比值与天然成分非常接近。CCT 模式的主要优点是无需进行劳动密集型的后数据处理,从而简化了分析程序。此外,与标准模式相比,CCT 模式还提高了对高质范围元素的灵敏度(系数为 1.5 到 2),同时不影响对内源性元素的灵敏度。我们强烈建议在分析整个质量范围内元素的生物成像应用中使用 CCT 模式。
本文章由计算机程序翻译,如有差异,请以英文原文为准。

摘要图片

摘要图片

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Advancing elemental analysis by collision/reaction cell technology and micro-droplet calibration for bioimaging applications by LA-ICP-TOFMS

Background

Bioimaging applications by laser ablation inductively coupled plasma time-of-flight mass spectrometry (LA-ICP-TOFMS) require comprehensive elemental analysis spanning the entire mass range. Within biological systems, endogenous elements are critical for maintaining metal homeostasis in cells, a fundamental aspect in disease progression when disrupted. Additionally, elements from the higher mass range may originate from various sources such as metal-tags for immuno-mass spectrometry, nanoparticles, or metal-based anticancer drugs. This study assesses the efficacy of collision/reacton cell (CCT) mode for simultaneously analysing elements across the complete mass spectrum. Furthermore, we demonstrate the accuracy of the LA-ICP-TOFMS methodology through the analysis of serum reference material.

Results

Our findings demonstrate that the CCT mode outperforms the standard/no gas mode for LA-ICP-TOFMS measurements, particularly in quantifying endogenous elements susceptible to interferences. Through the analysis of picolitre-volume micro-droplet standards and serum reference material (deposited as micro-droplets), we observed accurate quantification of elements such as iron and selenium, with isotope ratios closely resembling natural compositions. As key advantage, the utilization of the CCT mode eliminated the need for labor-intensive post-data processing, streamlining analytical procedures. Additionally, the CCT mode also provided enhanced sensitivity (factor of 1.5–2) for elements in the higher mass range compared to the standard mode, without compromising the sensitivity for endogenous elements.

Significance

We successfully integrated Seronorm serum reference material-containing micro-droplets into LA-ICPMS bioimaging workflows for quality control, enabling validated measurements of key biological elements. The use of the CCT mode is highly recommended for bioimaging applications that address the analysis of elements across the entire mass range.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Analytica Chimica Acta
Analytica Chimica Acta 化学-分析化学
CiteScore
10.40
自引率
6.50%
发文量
1081
审稿时长
38 days
期刊介绍: Analytica Chimica Acta has an open access mirror journal Analytica Chimica Acta: X, sharing the same aims and scope, editorial team, submission system and rigorous peer review. Analytica Chimica Acta provides a forum for the rapid publication of original research, and critical, comprehensive reviews dealing with all aspects of fundamental and applied modern analytical chemistry. The journal welcomes the submission of research papers which report studies concerning the development of new and significant analytical methodologies. In determining the suitability of submitted articles for publication, particular scrutiny will be placed on the degree of novelty and impact of the research and the extent to which it adds to the existing body of knowledge in analytical chemistry.
期刊最新文献
Development of a new spectral database of photo-induced pesticide compounds for an automatic monitoring and identification system. Glutathione-Functionalized Covalent Organic Frameworks@Silica as A Hydrophilic-Hydrophobic Balanced Mixed-Mode Stationary Phase for Highly Efficient Separation of Compounds with A Wide Range of Polarity Development of a portable SERS tool to evaluate the effectiveness of washing methods to remove pesticide residue from fruit surface Therapeutic Drug Monitoring of Methotrexate by Disposable SPCE Biosensor for Personalized Medicine Nanozymes Featuring a Mesoporous Silica Shell for Rapid Enrichment and Ultrasensitive Lateral Flow Immunoassay of Influenza A
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1