{"title":"确认鼻饲管位置的床旁方法的准确性和成本:诊断准确性研究。","authors":"Mayara Carvalho Godinho Rigobello, Claire Nierva Herrera, Carlos Alberto Grespan Bonacim, Rosana Aparecida Pereira, Roosevelt Santos Nunes, Jorge Elias Junior, Fernanda Raphael Escobar Gimenes","doi":"10.1007/s40477-024-00960-2","DOIUrl":null,"url":null,"abstract":"<p><strong>Aim: </strong>To analyze the accuracy and costs of bedside methods for confirming the position of a nasoenteral feeding tube newly inserted blindly by nurses.</p><p><strong>Design: </strong>Diagnostic accuracy study of three clinical methods (ultrasound, epigastric auscultation, and pH measurement) compared to radiography. The direct costs of each method used to confirm the positioning of the nasoenteral tube were also measured.</p><p><strong>Methods: </strong>Seventy-six adult patients underwent a total of 87 nasoenteral tube insertion procedures in hospital units located within the Northeast region of the State of São Paulo, Brazil. The clinical methods were conducted on all study participants in the specified sequence: ultrasound (as index test), followed by epigastric auscultation and pH measurement (also index tests).</p><p><strong>Results: </strong>The outcomes regarding the confirmation of the accurate positioning of the nasoenteral tube are as follows: ultrasonography demonstrated sensitivity and specific of 79.0% and 66.7%, respectively. Epigastric auscultation exhibited a sensitivity of 81.3% and specificity of 83.3%. The pH measurement method displayed sensitivity and specificity of 89.3% and 100% respectively. Additionally, in terms of estimated direct costs, the pH measurement method incurred a higher cost (USD $8.31) compared to the other methods, with a difference of USD $6.68.</p><p><strong>Conclusions: </strong>Based on these results, X-ray examination remains the primary method for confirming the placement of nasoenteral tubes recently inserted blindly at the bedside. However, when considering the costs of the evaluated methods, it is advisable to consider the variations in expenses between non-radiological methods and X-ray examinations.</p>","PeriodicalId":51528,"journal":{"name":"Journal of Ultrasound","volume":null,"pages":null},"PeriodicalIF":1.3000,"publicationDate":"2024-10-15","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Accuracy and costs of bedside methods for confirming nasoenteral feeding tube position: a diagnostic accuracy study.\",\"authors\":\"Mayara Carvalho Godinho Rigobello, Claire Nierva Herrera, Carlos Alberto Grespan Bonacim, Rosana Aparecida Pereira, Roosevelt Santos Nunes, Jorge Elias Junior, Fernanda Raphael Escobar Gimenes\",\"doi\":\"10.1007/s40477-024-00960-2\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Aim: </strong>To analyze the accuracy and costs of bedside methods for confirming the position of a nasoenteral feeding tube newly inserted blindly by nurses.</p><p><strong>Design: </strong>Diagnostic accuracy study of three clinical methods (ultrasound, epigastric auscultation, and pH measurement) compared to radiography. The direct costs of each method used to confirm the positioning of the nasoenteral tube were also measured.</p><p><strong>Methods: </strong>Seventy-six adult patients underwent a total of 87 nasoenteral tube insertion procedures in hospital units located within the Northeast region of the State of São Paulo, Brazil. The clinical methods were conducted on all study participants in the specified sequence: ultrasound (as index test), followed by epigastric auscultation and pH measurement (also index tests).</p><p><strong>Results: </strong>The outcomes regarding the confirmation of the accurate positioning of the nasoenteral tube are as follows: ultrasonography demonstrated sensitivity and specific of 79.0% and 66.7%, respectively. Epigastric auscultation exhibited a sensitivity of 81.3% and specificity of 83.3%. The pH measurement method displayed sensitivity and specificity of 89.3% and 100% respectively. Additionally, in terms of estimated direct costs, the pH measurement method incurred a higher cost (USD $8.31) compared to the other methods, with a difference of USD $6.68.</p><p><strong>Conclusions: </strong>Based on these results, X-ray examination remains the primary method for confirming the placement of nasoenteral tubes recently inserted blindly at the bedside. However, when considering the costs of the evaluated methods, it is advisable to consider the variations in expenses between non-radiological methods and X-ray examinations.</p>\",\"PeriodicalId\":51528,\"journal\":{\"name\":\"Journal of Ultrasound\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":1.3000,\"publicationDate\":\"2024-10-15\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Ultrasound\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1007/s40477-024-00960-2\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Ultrasound","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1007/s40477-024-00960-2","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
Accuracy and costs of bedside methods for confirming nasoenteral feeding tube position: a diagnostic accuracy study.
Aim: To analyze the accuracy and costs of bedside methods for confirming the position of a nasoenteral feeding tube newly inserted blindly by nurses.
Design: Diagnostic accuracy study of three clinical methods (ultrasound, epigastric auscultation, and pH measurement) compared to radiography. The direct costs of each method used to confirm the positioning of the nasoenteral tube were also measured.
Methods: Seventy-six adult patients underwent a total of 87 nasoenteral tube insertion procedures in hospital units located within the Northeast region of the State of São Paulo, Brazil. The clinical methods were conducted on all study participants in the specified sequence: ultrasound (as index test), followed by epigastric auscultation and pH measurement (also index tests).
Results: The outcomes regarding the confirmation of the accurate positioning of the nasoenteral tube are as follows: ultrasonography demonstrated sensitivity and specific of 79.0% and 66.7%, respectively. Epigastric auscultation exhibited a sensitivity of 81.3% and specificity of 83.3%. The pH measurement method displayed sensitivity and specificity of 89.3% and 100% respectively. Additionally, in terms of estimated direct costs, the pH measurement method incurred a higher cost (USD $8.31) compared to the other methods, with a difference of USD $6.68.
Conclusions: Based on these results, X-ray examination remains the primary method for confirming the placement of nasoenteral tubes recently inserted blindly at the bedside. However, when considering the costs of the evaluated methods, it is advisable to consider the variations in expenses between non-radiological methods and X-ray examinations.
期刊介绍:
The Journal of Ultrasound is the official journal of the Italian Society for Ultrasound in Medicine and Biology (SIUMB). The journal publishes original contributions (research and review articles, case reports, technical reports and letters to the editor) on significant advances in clinical diagnostic, interventional and therapeutic applications, clinical techniques, the physics, engineering and technology of ultrasound in medicine and biology, and in cross-sectional diagnostic imaging. The official language of Journal of Ultrasound is English.