利用蒙特卡洛模拟和实验测量对核磁共振成像林纳克的电子返回效应进行量化。

IF 1.3 Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Biomedical Physics & Engineering Express Pub Date : 2024-11-13 DOI:10.1088/2057-1976/ad8ce3
Francesco Tortorelli, Cristian Borrazzo, Marica Masi, Maria Rago, Randa El Gawhary, Claudio Properzi, Domenico Marchesano, Gianmarco Grimaldi, Federico Bianciardi, Ivan Annessi, Annamaria Di Palma, Maria Valentino, Laura Verna, Giuseppina Chiarello, Plastino Wolfango, Piercarlo Gentile
{"title":"利用蒙特卡洛模拟和实验测量对核磁共振成像林纳克的电子返回效应进行量化。","authors":"Francesco Tortorelli, Cristian Borrazzo, Marica Masi, Maria Rago, Randa El Gawhary, Claudio Properzi, Domenico Marchesano, Gianmarco Grimaldi, Federico Bianciardi, Ivan Annessi, Annamaria Di Palma, Maria Valentino, Laura Verna, Giuseppina Chiarello, Plastino Wolfango, Piercarlo Gentile","doi":"10.1088/2057-1976/ad8ce3","DOIUrl":null,"url":null,"abstract":"<p><p>The integration of magnetic resonance (MR) imaging and linear accelerators into hybrid treatment systems has made MR-guided radiation therapy a clinical reality. This work aims to evaluate the influence of the Electron Return Effect (ERE) on the dose distributions. This study was conducted using MRIdian (ViewRay, Cleveland, Ohio) system. Monte-Carlo simulations (MCs) and experimental measurements with EBT3 Gafchromic films were performed to investigate the dose distribution in a slab water phantom with and without a 2-cm air gap. Plus, MCs took into account different field sizes and a lung gap. A gamma analysis compared calculated versus measured dose distributions. The MCs have shown an increase of the ERE with the radiation field size both in Percent Depth Dose (PDD) and crossline direction. As concerns to the PDD direction, the smallest field for which there was a significant dose accumulation was 4.15 × 4.15 cm<sup>2</sup>both for air-gap (13.5%) and lung-gap (3.3%). The largest field for which there was a significant dose accumulation was 24.07 × 24.07 cm<sup>2</sup>both for air-gap (39.7%) and lung-gap (4.9%). Instead for the crossline direction, the smallest field for which there was a significant dose accumulation was 2.49 × 2.49 cm<sup>2</sup>both for air-gap (8.6% ) and lung-gap (0.5%). The largest field for which there was a significant dose accumulation was 24.07 × 24.07 cm<sup>2</sup>both for air-gap (46.2%) and lung-gap (4.2%). PDD and crossline profiles showed good agreement with a gamma-passing rate higher than 91.15% for 2%/2 mm. The ERE can be adequately calculated by MC dose calculation platform available in the MRIdian Treatment Planning System. The MCs show an increase of the ERE directly proportional with the radiation field size. Good agreement was observed between the experimental measurements and calculated dose distributions.</p>","PeriodicalId":8896,"journal":{"name":"Biomedical Physics & Engineering Express","volume":" ","pages":""},"PeriodicalIF":1.3000,"publicationDate":"2024-11-13","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"A quantification of the electron return effect using Monte Carlo simulations and experimental measurements for the MRI-linac.\",\"authors\":\"Francesco Tortorelli, Cristian Borrazzo, Marica Masi, Maria Rago, Randa El Gawhary, Claudio Properzi, Domenico Marchesano, Gianmarco Grimaldi, Federico Bianciardi, Ivan Annessi, Annamaria Di Palma, Maria Valentino, Laura Verna, Giuseppina Chiarello, Plastino Wolfango, Piercarlo Gentile\",\"doi\":\"10.1088/2057-1976/ad8ce3\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>The integration of magnetic resonance (MR) imaging and linear accelerators into hybrid treatment systems has made MR-guided radiation therapy a clinical reality. This work aims to evaluate the influence of the Electron Return Effect (ERE) on the dose distributions. This study was conducted using MRIdian (ViewRay, Cleveland, Ohio) system. Monte-Carlo simulations (MCs) and experimental measurements with EBT3 Gafchromic films were performed to investigate the dose distribution in a slab water phantom with and without a 2-cm air gap. Plus, MCs took into account different field sizes and a lung gap. A gamma analysis compared calculated versus measured dose distributions. The MCs have shown an increase of the ERE with the radiation field size both in Percent Depth Dose (PDD) and crossline direction. As concerns to the PDD direction, the smallest field for which there was a significant dose accumulation was 4.15 × 4.15 cm<sup>2</sup>both for air-gap (13.5%) and lung-gap (3.3%). The largest field for which there was a significant dose accumulation was 24.07 × 24.07 cm<sup>2</sup>both for air-gap (39.7%) and lung-gap (4.9%). Instead for the crossline direction, the smallest field for which there was a significant dose accumulation was 2.49 × 2.49 cm<sup>2</sup>both for air-gap (8.6% ) and lung-gap (0.5%). The largest field for which there was a significant dose accumulation was 24.07 × 24.07 cm<sup>2</sup>both for air-gap (46.2%) and lung-gap (4.2%). PDD and crossline profiles showed good agreement with a gamma-passing rate higher than 91.15% for 2%/2 mm. The ERE can be adequately calculated by MC dose calculation platform available in the MRIdian Treatment Planning System. The MCs show an increase of the ERE directly proportional with the radiation field size. Good agreement was observed between the experimental measurements and calculated dose distributions.</p>\",\"PeriodicalId\":8896,\"journal\":{\"name\":\"Biomedical Physics & Engineering Express\",\"volume\":\" \",\"pages\":\"\"},\"PeriodicalIF\":1.3000,\"publicationDate\":\"2024-11-13\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Biomedical Physics & Engineering Express\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1088/2057-1976/ad8ce3\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Biomedical Physics & Engineering Express","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1088/2057-1976/ad8ce3","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

摘要

将磁共振(MR)成像和直线加速器整合到混合治疗系统中,使磁共振引导的放射治疗成为临床现实。这项研究使用 MRIdian(ViewRay,俄亥俄州克利夫兰市)系统进行。使用 EBT3 Gafchromic 薄膜进行了蒙特卡洛模拟(MC)和实验测量,以研究有 2 厘米气隙和无 2 厘米气隙的板状水模型中的剂量分布。此外,模拟还考虑了不同的场大小和肺间隙。伽马分析比较了计算得出的剂量分布与测量得出的剂量分布。就深度剂量百分比(PDD)方向而言,在气隙(13.5%)和肺隙(3.3%)出现显著剂量累积的最小辐射场为 4.15x4.15 平方厘米。空气间隙(39.7%)和肺间隙(4.9%)的最大剂量累积区域为 24.07x24.07 平方厘米。而在横线方向上,空气间隙(8.6%)和肺间隙(0.5%)出现明显剂量累积的最小区域为 2.49x2.49 平方厘米。气隙(46.2%)和肺隙(4.2%)的最大剂量累积区域为 24.07x24.07 平方厘米。MRIdian治疗计划系统中的MC剂量计算平台可以充分计算ERE。MC显示ERE的增加与辐射场的大小成正比。实验测量结果与计算得出的剂量分布之间具有良好的一致性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
A quantification of the electron return effect using Monte Carlo simulations and experimental measurements for the MRI-linac.

The integration of magnetic resonance (MR) imaging and linear accelerators into hybrid treatment systems has made MR-guided radiation therapy a clinical reality. This work aims to evaluate the influence of the Electron Return Effect (ERE) on the dose distributions. This study was conducted using MRIdian (ViewRay, Cleveland, Ohio) system. Monte-Carlo simulations (MCs) and experimental measurements with EBT3 Gafchromic films were performed to investigate the dose distribution in a slab water phantom with and without a 2-cm air gap. Plus, MCs took into account different field sizes and a lung gap. A gamma analysis compared calculated versus measured dose distributions. The MCs have shown an increase of the ERE with the radiation field size both in Percent Depth Dose (PDD) and crossline direction. As concerns to the PDD direction, the smallest field for which there was a significant dose accumulation was 4.15 × 4.15 cm2both for air-gap (13.5%) and lung-gap (3.3%). The largest field for which there was a significant dose accumulation was 24.07 × 24.07 cm2both for air-gap (39.7%) and lung-gap (4.9%). Instead for the crossline direction, the smallest field for which there was a significant dose accumulation was 2.49 × 2.49 cm2both for air-gap (8.6% ) and lung-gap (0.5%). The largest field for which there was a significant dose accumulation was 24.07 × 24.07 cm2both for air-gap (46.2%) and lung-gap (4.2%). PDD and crossline profiles showed good agreement with a gamma-passing rate higher than 91.15% for 2%/2 mm. The ERE can be adequately calculated by MC dose calculation platform available in the MRIdian Treatment Planning System. The MCs show an increase of the ERE directly proportional with the radiation field size. Good agreement was observed between the experimental measurements and calculated dose distributions.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Biomedical Physics & Engineering Express
Biomedical Physics & Engineering Express RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING-
CiteScore
2.80
自引率
0.00%
发文量
153
期刊介绍: BPEX is an inclusive, international, multidisciplinary journal devoted to publishing new research on any application of physics and/or engineering in medicine and/or biology. Characterized by a broad geographical coverage and a fast-track peer-review process, relevant topics include all aspects of biophysics, medical physics and biomedical engineering. Papers that are almost entirely clinical or biological in their focus are not suitable. The journal has an emphasis on publishing interdisciplinary work and bringing research fields together, encompassing experimental, theoretical and computational work.
期刊最新文献
Dose compensation for decreased biological effective dose due to intrafractional interruption during radiotherapy: integration with a commercial treatment planning system. Open-window MSR Design with Active Magnetic Compensation Coil based on COMSOL Multiphysics. A new method to assess the performance of anti-scatter grids in x-ray projection imaging. Effect of tissue viscoelasticity on delivered mechanical power in a physical respiratory system model: distinguishing between airway and tissue resistance. Local field potential-based brain-machine interface to inhibit epileptic seizures by spinal cord electrical stimulation.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1