{"title":"探索长丝密度对用于高能光子放射治疗的三维打印栓剂材料响应性的影响。","authors":"","doi":"10.1016/j.ejmp.2024.104849","DOIUrl":null,"url":null,"abstract":"<div><h3>Background</h3><div>3D-printed boluses in radiation therapy receive consideration for their ability to enhance treatment precision and patient comfort. Yet, thorough validation of 3D-printed boluses using various validation procedures and statistical analysis is missing. This study aims to determine the effectiveness of using 3D-printed boluses in radiation therapy.</div></div><div><h3>Method</h3><div>The CT Hounsfield Unit (HU) profiles of the 3D-printed materials were compared to those of the commercial bolus using the Eclipse Treatment Planning System (TPS) unit. Furthermore, absolute dose measurements were carried out to assess the efficacy of the 3D-printed samples by using concordance correlation coefficient to assess the agreement between 3D materials and the commercial bolus.</div></div><div><h3>Results</h3><div>The average HU profiles of 3D-printed materials were: −144.53 (ABS), −124.40 (ASA), 9.55 (PLA-P), −140.79 (Polycarbonate), −68.58 (PLA-S), and −113.159 (PET-G), respectively. PDD scans showed that air gaps between the bolus and surface shifted the maximum dose depth. Whereas dosimetry has shown that ASA and Polycarbonate are different in attenuation from other tested filaments. This limitation could affect their performance in specific applications within radiation therapy<em>.</em> The final analysis, using the TPS-generated datasets to assess the area under the dose curve in the build-up zone of each 3D-bolus, excluded ABS.</div></div><div><h3>Conclusions</h3><div>The results from PDD scans and dose assessments offer compelling proof that 3D-printed boluses are effective for delivering surface dosage and are like commercially available boluses. Moreover, specific materials showed a statistically significant improvement in delivering the dose. The results highlight the capability of 3D-printed boluses to enhance the effectiveness of radiation therapy.</div></div>","PeriodicalId":56092,"journal":{"name":"Physica Medica-European Journal of Medical Physics","volume":null,"pages":null},"PeriodicalIF":3.3000,"publicationDate":"2024-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Exploring the impact of filament density on the responsiveness of 3D-Printed bolus materials for high-energy photon radiotherapy\",\"authors\":\"\",\"doi\":\"10.1016/j.ejmp.2024.104849\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Background</h3><div>3D-printed boluses in radiation therapy receive consideration for their ability to enhance treatment precision and patient comfort. Yet, thorough validation of 3D-printed boluses using various validation procedures and statistical analysis is missing. This study aims to determine the effectiveness of using 3D-printed boluses in radiation therapy.</div></div><div><h3>Method</h3><div>The CT Hounsfield Unit (HU) profiles of the 3D-printed materials were compared to those of the commercial bolus using the Eclipse Treatment Planning System (TPS) unit. Furthermore, absolute dose measurements were carried out to assess the efficacy of the 3D-printed samples by using concordance correlation coefficient to assess the agreement between 3D materials and the commercial bolus.</div></div><div><h3>Results</h3><div>The average HU profiles of 3D-printed materials were: −144.53 (ABS), −124.40 (ASA), 9.55 (PLA-P), −140.79 (Polycarbonate), −68.58 (PLA-S), and −113.159 (PET-G), respectively. PDD scans showed that air gaps between the bolus and surface shifted the maximum dose depth. Whereas dosimetry has shown that ASA and Polycarbonate are different in attenuation from other tested filaments. This limitation could affect their performance in specific applications within radiation therapy<em>.</em> The final analysis, using the TPS-generated datasets to assess the area under the dose curve in the build-up zone of each 3D-bolus, excluded ABS.</div></div><div><h3>Conclusions</h3><div>The results from PDD scans and dose assessments offer compelling proof that 3D-printed boluses are effective for delivering surface dosage and are like commercially available boluses. Moreover, specific materials showed a statistically significant improvement in delivering the dose. The results highlight the capability of 3D-printed boluses to enhance the effectiveness of radiation therapy.</div></div>\",\"PeriodicalId\":56092,\"journal\":{\"name\":\"Physica Medica-European Journal of Medical Physics\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":3.3000,\"publicationDate\":\"2024-11-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Physica Medica-European Journal of Medical Physics\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S1120179724011062\",\"RegionNum\":3,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q1\",\"JCRName\":\"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Physica Medica-European Journal of Medical Physics","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S1120179724011062","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

摘要

背景:三维打印放射治疗栓因其能够提高治疗精确度和患者舒适度而受到关注。然而,目前还没有利用各种验证程序和统计分析对 3D 打印栓剂进行彻底验证。本研究旨在确定在放射治疗中使用 3D 打印栓剂的有效性:方法:使用 Eclipse 治疗计划系统(TPS)装置,将三维打印材料的 CT Hounsfield 单位(HU)曲线与商用栓剂的 CT Hounsfield 单位(HU)曲线进行比较。此外,还进行了绝对剂量测量,以评估三维打印样本的疗效:结果:3D打印材料的平均HU曲线为结果:三维打印材料的平均 HU 值分别为:-144.53(ABS)、-124.40(ASA)、9.55(PLA-P)、-140.79(聚碳酸酯)、-68.58(PLA-S)和-113.159(PET-G)。PDD 扫描显示,栓剂和表面之间的空气间隙改变了最大剂量深度。而剂量测定显示,ASA 和聚碳酸酯在衰减方面与其他测试丝不同。这一限制可能会影响它们在放射治疗特定应用中的性能。使用 TPS 生成的数据集进行最终分析,以评估每个三维脆片堆积区的剂量曲线下面积,结果排除了 ABS:PDD扫描和剂量评估的结果有力地证明了三维打印栓剂能有效地提供表面剂量,并且与市面上的栓剂相似。此外,特定材料在投放剂量方面有显著的统计学改进。这些结果凸显了三维打印栓剂提高放射治疗效果的能力。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Exploring the impact of filament density on the responsiveness of 3D-Printed bolus materials for high-energy photon radiotherapy

Background

3D-printed boluses in radiation therapy receive consideration for their ability to enhance treatment precision and patient comfort. Yet, thorough validation of 3D-printed boluses using various validation procedures and statistical analysis is missing. This study aims to determine the effectiveness of using 3D-printed boluses in radiation therapy.

Method

The CT Hounsfield Unit (HU) profiles of the 3D-printed materials were compared to those of the commercial bolus using the Eclipse Treatment Planning System (TPS) unit. Furthermore, absolute dose measurements were carried out to assess the efficacy of the 3D-printed samples by using concordance correlation coefficient to assess the agreement between 3D materials and the commercial bolus.

Results

The average HU profiles of 3D-printed materials were: −144.53 (ABS), −124.40 (ASA), 9.55 (PLA-P), −140.79 (Polycarbonate), −68.58 (PLA-S), and −113.159 (PET-G), respectively. PDD scans showed that air gaps between the bolus and surface shifted the maximum dose depth. Whereas dosimetry has shown that ASA and Polycarbonate are different in attenuation from other tested filaments. This limitation could affect their performance in specific applications within radiation therapy. The final analysis, using the TPS-generated datasets to assess the area under the dose curve in the build-up zone of each 3D-bolus, excluded ABS.

Conclusions

The results from PDD scans and dose assessments offer compelling proof that 3D-printed boluses are effective for delivering surface dosage and are like commercially available boluses. Moreover, specific materials showed a statistically significant improvement in delivering the dose. The results highlight the capability of 3D-printed boluses to enhance the effectiveness of radiation therapy.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
6.80
自引率
14.70%
发文量
493
审稿时长
78 days
期刊介绍: Physica Medica, European Journal of Medical Physics, publishing with Elsevier from 2007, provides an international forum for research and reviews on the following main topics: Medical Imaging Radiation Therapy Radiation Protection Measuring Systems and Signal Processing Education and training in Medical Physics Professional issues in Medical Physics.
期刊最新文献
Potential dose reduction and image quality improvement in chest CT with a photon-counting CT compared to a new dual-source CT Dosimetric study of bevel factors in IOERT with mobile linacs: Towards a unified code of practice Assessment of pencil beam scanning proton therapy beam delivery accuracy through machine learning and log file analysis Exploring the impact of filament density on the responsiveness of 3D-Printed bolus materials for high-energy photon radiotherapy Needle artifact redistribution technique (Needle-ART): A method for metal artifact reduction during CT interventionism based on gantry tilt
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1