{"title":"比较两种不同的商用血液培养系统(BacT/Alert 3D和Autobio BC)的测定内和测定间重现性及阳性检测时间。","authors":"Nilgün Kansak , Nilay Zeynep Kalender , Neslihan Arıcı , Rıza Adaleti , Sebahat Aksaray , Handan Ankaralı , Nevriye Gönüllü","doi":"10.1016/j.ijmmb.2024.100754","DOIUrl":null,"url":null,"abstract":"<div><h3>Purpose</h3><div>In our study, we aimed to compare the performance of the BacT/Alert 3D (bioMerieux, France) system, which is currently used in our laboratory, and the Autobio BC (Autobio, China) system, which was newly introduced in our country, using standard and clinical isolates.</div></div><div><h3>Methods</h3><div>Bacterial suspension was prepared by two technicians on the same day and three consecutive days from five different standard strains with 0.5 McFarland turbidity, then serial dilution to a final concentration was adjusted and was simultaneously inoculated in aerobic blood culture bottles. The bacterial concentration was measured by making a quantitative counting plate. The same procedure was also performed for 55 clinical isolates belonging to eleven species. After simulated bacteremia with standard and clinical isolates, the growth results were confirmed by inoculation from positive blood culture bottles onto solid medium and identification was made in the next day with MALDI-TOF MS (bioMérieux). In each study, sterile saline and blood was inoculated into the bottles as a negative control to check contamination. Intra-assay and inter-assay reproducibility of recovery rates and detection times of standard strains; recovery rates and detection times of clinical isolates were compared for both systems.</div></div><div><h3>Results</h3><div>Recovery rates were 100 % in both systems, and when positive detection times were compared, it was found that there was no difference between the two devices in clinical isolates (p:0.262) but that Autobio BC gave significantly (p < 0.001) earlier results in standard strains.</div></div><div><h3>Conclusions</h3><div>In our simulated bloodstream infection study, Autobio BC was found to be comparable with BacT/Alert 3D, both recovery rates and growth detection time performance were found to be very good, and it can be used in routine microbiology laboratories.</div></div>","PeriodicalId":13284,"journal":{"name":"Indian Journal of Medical Microbiology","volume":"53 ","pages":"Article 100754"},"PeriodicalIF":1.4000,"publicationDate":"2024-11-21","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Comparison of intra-assay and inter-assay reproducibility and positive detection times of two different (BacT/Alert 3D and Autobio BC) commercial blood culture systems\",\"authors\":\"Nilgün Kansak , Nilay Zeynep Kalender , Neslihan Arıcı , Rıza Adaleti , Sebahat Aksaray , Handan Ankaralı , Nevriye Gönüllü\",\"doi\":\"10.1016/j.ijmmb.2024.100754\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Purpose</h3><div>In our study, we aimed to compare the performance of the BacT/Alert 3D (bioMerieux, France) system, which is currently used in our laboratory, and the Autobio BC (Autobio, China) system, which was newly introduced in our country, using standard and clinical isolates.</div></div><div><h3>Methods</h3><div>Bacterial suspension was prepared by two technicians on the same day and three consecutive days from five different standard strains with 0.5 McFarland turbidity, then serial dilution to a final concentration was adjusted and was simultaneously inoculated in aerobic blood culture bottles. The bacterial concentration was measured by making a quantitative counting plate. The same procedure was also performed for 55 clinical isolates belonging to eleven species. After simulated bacteremia with standard and clinical isolates, the growth results were confirmed by inoculation from positive blood culture bottles onto solid medium and identification was made in the next day with MALDI-TOF MS (bioMérieux). In each study, sterile saline and blood was inoculated into the bottles as a negative control to check contamination. Intra-assay and inter-assay reproducibility of recovery rates and detection times of standard strains; recovery rates and detection times of clinical isolates were compared for both systems.</div></div><div><h3>Results</h3><div>Recovery rates were 100 % in both systems, and when positive detection times were compared, it was found that there was no difference between the two devices in clinical isolates (p:0.262) but that Autobio BC gave significantly (p < 0.001) earlier results in standard strains.</div></div><div><h3>Conclusions</h3><div>In our simulated bloodstream infection study, Autobio BC was found to be comparable with BacT/Alert 3D, both recovery rates and growth detection time performance were found to be very good, and it can be used in routine microbiology laboratories.</div></div>\",\"PeriodicalId\":13284,\"journal\":{\"name\":\"Indian Journal of Medical Microbiology\",\"volume\":\"53 \",\"pages\":\"Article 100754\"},\"PeriodicalIF\":1.4000,\"publicationDate\":\"2024-11-21\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Indian Journal of Medical Microbiology\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S0255085724002299\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"IMMUNOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Indian Journal of Medical Microbiology","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0255085724002299","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"IMMUNOLOGY","Score":null,"Total":0}
引用次数: 0
摘要
目的:我们的研究旨在使用标准菌株和临床分离菌株,比较我们实验室目前使用的 BacT/Alert 3D(法国生物梅里埃公司)系统和我国新引进的 Autobio BC(中国 Autobio 公司)系统的性能:细菌悬液由两名技术人员在同一天和连续三天从五种不同的标准菌株中制备,浊度为 0.5 McFarland,然后连续稀释至最终浓度并同时接种到需氧血液培养瓶中。通过制作定量计数板测量细菌浓度。对属于 11 个菌种的 55 个临床分离菌也进行了同样的处理。用标准菌和临床分离菌模拟菌血症后,将阳性血液培养瓶接种到固体培养基上,确认生长结果,第二天用 MALDI-TOF MS(生物梅里埃)进行鉴定。在每项研究中,都将无菌生理盐水和血液接种到瓶中作为阴性对照,以检查污染情况。比较了两种系统对标准菌株的回收率和检测时间、临床分离菌株的回收率和检测时间的测定内和测定间重现性:结果:两种系统的回收率均为 100%,在比较阳性检测时间时发现,两种设备对临床分离菌株的检测时间没有差异(p:0.262),但 Autobio BC 的检测时间明显(pConclusions):在我们的模拟血流感染研究中发现,Autobio BC 与 BacT/Alert 3D 的性能相当,回收率和生长检测时间的表现都很好,可用于常规微生物实验室。
Comparison of intra-assay and inter-assay reproducibility and positive detection times of two different (BacT/Alert 3D and Autobio BC) commercial blood culture systems
Purpose
In our study, we aimed to compare the performance of the BacT/Alert 3D (bioMerieux, France) system, which is currently used in our laboratory, and the Autobio BC (Autobio, China) system, which was newly introduced in our country, using standard and clinical isolates.
Methods
Bacterial suspension was prepared by two technicians on the same day and three consecutive days from five different standard strains with 0.5 McFarland turbidity, then serial dilution to a final concentration was adjusted and was simultaneously inoculated in aerobic blood culture bottles. The bacterial concentration was measured by making a quantitative counting plate. The same procedure was also performed for 55 clinical isolates belonging to eleven species. After simulated bacteremia with standard and clinical isolates, the growth results were confirmed by inoculation from positive blood culture bottles onto solid medium and identification was made in the next day with MALDI-TOF MS (bioMérieux). In each study, sterile saline and blood was inoculated into the bottles as a negative control to check contamination. Intra-assay and inter-assay reproducibility of recovery rates and detection times of standard strains; recovery rates and detection times of clinical isolates were compared for both systems.
Results
Recovery rates were 100 % in both systems, and when positive detection times were compared, it was found that there was no difference between the two devices in clinical isolates (p:0.262) but that Autobio BC gave significantly (p < 0.001) earlier results in standard strains.
Conclusions
In our simulated bloodstream infection study, Autobio BC was found to be comparable with BacT/Alert 3D, both recovery rates and growth detection time performance were found to be very good, and it can be used in routine microbiology laboratories.
期刊介绍:
Manuscripts of high standard in the form of original research, multicentric studies, meta analysis, are accepted. Current reports can be submitted as brief communications. Case reports must include review of current literature, clinical details, outcome and follow up. Letters to the editor must be a comment on or pertain to a manuscript already published in the IJMM or in relation to preliminary communication of a larger study.