{"title":"流体动力学和体外血液研究了解导管相关血栓形成。","authors":"Hannah Palahnuk, Boyang Su, Thaddeus Harbaugh, Cleo Gesenberg, Shouhao Zhou, Elias Rizk, Jonathan Bernstein, S Will Hazard, Keefe B Manning","doi":"10.1007/s13239-024-00761-y","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>Central venous catheters (CVCs) provide a direct route to the venous circulation but are prone to catheter-related thrombosis (CRT). A known CRT risk factor is a high catheter-to-vein ratio (CVR), or a large catheter diameter with respect to the indwelling vein size. In this study, the CVR's effect on CVC hemodynamics and its impact on CRT is investigated with in vitro and in silico experiments.</p><p><strong>Methods: </strong>An in vitro flow loop is used to characterize the hemodynamics around CVCs using particle image velocimetry. In addition, CRT is investigated using an in vitro flow loop with human blood and clinical catheters. The wall shear rate of flow around the CVC is computed numerically. CVRs of 0.20, 0.33, and 0.49 and Reynolds numbers of 200, 800, and 1300 are evaluated. No flow is used through CVC lumens to model chronic indwelling catheters.</p><p><strong>Results: </strong>Results show CVR ≥ 0.33 promotes platelet-rich clot growth at the device tip and at an increased rate compared to lower CVR cases. A high wall shear rate gradient on the CVC tip and an extended wake distal to the tip exists for higher CVR cases, promoting the aggregation of platelets and subsequent stagnation for clot formation. Further, the combination of the CVR and Reynolds number are crucial to CRT potential, not the CVR alone. Specifically, thrombosis risk is increased with low (stasis driven) and/or high (platelet activation driven) flow conditions, with the CVR and CVC's geometry playing an additional role in promoting fluid mechanic driven thrombus development. A high CVR (≥ 0.33) and high flow condition (≥ 1300) results in the highest risk for clot growth at the tip of the device; other locations of the device are at risk for thrombus development in lower flow conditions, regardless of the CVR. The importance of the device geometry and flow in promoting thrombus and fibrin sheath formation is also shown for the device investigated.</p><p><strong>Conclusions: </strong>This work demonstrates that the CVR, flow, and device geometry affect CRT. For clinical cases with CVR ≥ 0.33 and/or Re ≥ 1300, the device tip may be monitored more consistently for clot formation. Thrombosis risks remain on the entire catheter, regardless of the flow condition, for a CVR = 0.49. Device placement should be chosen carefully with respect to the combination of the Reynolds number and CVR. Further study is needed on the effect of catheterization to confirm these findings.</p>","PeriodicalId":54322,"journal":{"name":"Cardiovascular Engineering and Technology","volume":" ","pages":"116-137"},"PeriodicalIF":1.6000,"publicationDate":"2025-02-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Fluid Dynamic and in Vitro Blood Study to Understand Catheter-Related Thrombosis.\",\"authors\":\"Hannah Palahnuk, Boyang Su, Thaddeus Harbaugh, Cleo Gesenberg, Shouhao Zhou, Elias Rizk, Jonathan Bernstein, S Will Hazard, Keefe B Manning\",\"doi\":\"10.1007/s13239-024-00761-y\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Purpose: </strong>Central venous catheters (CVCs) provide a direct route to the venous circulation but are prone to catheter-related thrombosis (CRT). A known CRT risk factor is a high catheter-to-vein ratio (CVR), or a large catheter diameter with respect to the indwelling vein size. In this study, the CVR's effect on CVC hemodynamics and its impact on CRT is investigated with in vitro and in silico experiments.</p><p><strong>Methods: </strong>An in vitro flow loop is used to characterize the hemodynamics around CVCs using particle image velocimetry. In addition, CRT is investigated using an in vitro flow loop with human blood and clinical catheters. The wall shear rate of flow around the CVC is computed numerically. CVRs of 0.20, 0.33, and 0.49 and Reynolds numbers of 200, 800, and 1300 are evaluated. No flow is used through CVC lumens to model chronic indwelling catheters.</p><p><strong>Results: </strong>Results show CVR ≥ 0.33 promotes platelet-rich clot growth at the device tip and at an increased rate compared to lower CVR cases. A high wall shear rate gradient on the CVC tip and an extended wake distal to the tip exists for higher CVR cases, promoting the aggregation of platelets and subsequent stagnation for clot formation. Further, the combination of the CVR and Reynolds number are crucial to CRT potential, not the CVR alone. Specifically, thrombosis risk is increased with low (stasis driven) and/or high (platelet activation driven) flow conditions, with the CVR and CVC's geometry playing an additional role in promoting fluid mechanic driven thrombus development. A high CVR (≥ 0.33) and high flow condition (≥ 1300) results in the highest risk for clot growth at the tip of the device; other locations of the device are at risk for thrombus development in lower flow conditions, regardless of the CVR. The importance of the device geometry and flow in promoting thrombus and fibrin sheath formation is also shown for the device investigated.</p><p><strong>Conclusions: </strong>This work demonstrates that the CVR, flow, and device geometry affect CRT. For clinical cases with CVR ≥ 0.33 and/or Re ≥ 1300, the device tip may be monitored more consistently for clot formation. Thrombosis risks remain on the entire catheter, regardless of the flow condition, for a CVR = 0.49. Device placement should be chosen carefully with respect to the combination of the Reynolds number and CVR. Further study is needed on the effect of catheterization to confirm these findings.</p>\",\"PeriodicalId\":54322,\"journal\":{\"name\":\"Cardiovascular Engineering and Technology\",\"volume\":\" \",\"pages\":\"116-137\"},\"PeriodicalIF\":1.6000,\"publicationDate\":\"2025-02-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Cardiovascular Engineering and Technology\",\"FirstCategoryId\":\"5\",\"ListUrlMain\":\"https://doi.org/10.1007/s13239-024-00761-y\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2024/12/2 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q3\",\"JCRName\":\"CARDIAC & CARDIOVASCULAR SYSTEMS\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Cardiovascular Engineering and Technology","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1007/s13239-024-00761-y","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/12/2 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
Fluid Dynamic and in Vitro Blood Study to Understand Catheter-Related Thrombosis.
Purpose: Central venous catheters (CVCs) provide a direct route to the venous circulation but are prone to catheter-related thrombosis (CRT). A known CRT risk factor is a high catheter-to-vein ratio (CVR), or a large catheter diameter with respect to the indwelling vein size. In this study, the CVR's effect on CVC hemodynamics and its impact on CRT is investigated with in vitro and in silico experiments.
Methods: An in vitro flow loop is used to characterize the hemodynamics around CVCs using particle image velocimetry. In addition, CRT is investigated using an in vitro flow loop with human blood and clinical catheters. The wall shear rate of flow around the CVC is computed numerically. CVRs of 0.20, 0.33, and 0.49 and Reynolds numbers of 200, 800, and 1300 are evaluated. No flow is used through CVC lumens to model chronic indwelling catheters.
Results: Results show CVR ≥ 0.33 promotes platelet-rich clot growth at the device tip and at an increased rate compared to lower CVR cases. A high wall shear rate gradient on the CVC tip and an extended wake distal to the tip exists for higher CVR cases, promoting the aggregation of platelets and subsequent stagnation for clot formation. Further, the combination of the CVR and Reynolds number are crucial to CRT potential, not the CVR alone. Specifically, thrombosis risk is increased with low (stasis driven) and/or high (platelet activation driven) flow conditions, with the CVR and CVC's geometry playing an additional role in promoting fluid mechanic driven thrombus development. A high CVR (≥ 0.33) and high flow condition (≥ 1300) results in the highest risk for clot growth at the tip of the device; other locations of the device are at risk for thrombus development in lower flow conditions, regardless of the CVR. The importance of the device geometry and flow in promoting thrombus and fibrin sheath formation is also shown for the device investigated.
Conclusions: This work demonstrates that the CVR, flow, and device geometry affect CRT. For clinical cases with CVR ≥ 0.33 and/or Re ≥ 1300, the device tip may be monitored more consistently for clot formation. Thrombosis risks remain on the entire catheter, regardless of the flow condition, for a CVR = 0.49. Device placement should be chosen carefully with respect to the combination of the Reynolds number and CVR. Further study is needed on the effect of catheterization to confirm these findings.
期刊介绍:
Cardiovascular Engineering and Technology is a journal publishing the spectrum of basic to translational research in all aspects of cardiovascular physiology and medical treatment. It is the forum for academic and industrial investigators to disseminate research that utilizes engineering principles and methods to advance fundamental knowledge and technological solutions related to the cardiovascular system. Manuscripts spanning from subcellular to systems level topics are invited, including but not limited to implantable medical devices, hemodynamics and tissue biomechanics, functional imaging, surgical devices, electrophysiology, tissue engineering and regenerative medicine, diagnostic instruments, transport and delivery of biologics, and sensors. In addition to manuscripts describing the original publication of research, manuscripts reviewing developments in these topics or their state-of-art are also invited.