X Chenivesse, B Anliker, A Daas, S Ferro, R Meier, M Renner, A Costanzo
{"title":"紫外分光光度法测定人基因治疗产品载体制剂中质粒DNA浓度的评价。","authors":"X Chenivesse, B Anliker, A Daas, S Ferro, R Meier, M Renner, A Costanzo","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>The European Pharmacopoeia (Ph. Eur.) general chapter 5.14. Gene transfer medicinal products for human use suggests the use of absorbance measurements at 260 nm to determine the DNA concentration of plasmid vectors used for the preparation of gene therapy products for human use. An international collaborative study was organised by the European Directorate for the Quality of Medicines & HealthCare (EDQM) to confirm the suitability of UV spectrophotometry for the quantification of plasmid vectors used in gene therapy (GT). Three Official Medicine Control Laboratories (OMCLs of the European OMCL Network) and members of the OMCL Working Group for GT products took part in the study, in which various types of spectrophotometers were assessed using common test samples. Results of the study demonstrated that UV spectrophotometry can be considered suitable for the quantification of plasmid DNA in GT products regardless of the instrument used.</p>","PeriodicalId":39192,"journal":{"name":"Pharmeuropa bio & scientific notes","volume":"2017 ","pages":"88-112"},"PeriodicalIF":0.0000,"publicationDate":"2017-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Assessment of UV spectrophotometry for determination of plasmid DNA concentration in vector preparations for human gene therapy products.\",\"authors\":\"X Chenivesse, B Anliker, A Daas, S Ferro, R Meier, M Renner, A Costanzo\",\"doi\":\"\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>The European Pharmacopoeia (Ph. Eur.) general chapter 5.14. Gene transfer medicinal products for human use suggests the use of absorbance measurements at 260 nm to determine the DNA concentration of plasmid vectors used for the preparation of gene therapy products for human use. An international collaborative study was organised by the European Directorate for the Quality of Medicines & HealthCare (EDQM) to confirm the suitability of UV spectrophotometry for the quantification of plasmid vectors used in gene therapy (GT). Three Official Medicine Control Laboratories (OMCLs of the European OMCL Network) and members of the OMCL Working Group for GT products took part in the study, in which various types of spectrophotometers were assessed using common test samples. Results of the study demonstrated that UV spectrophotometry can be considered suitable for the quantification of plasmid DNA in GT products regardless of the instrument used.</p>\",\"PeriodicalId\":39192,\"journal\":{\"name\":\"Pharmeuropa bio & scientific notes\",\"volume\":\"2017 \",\"pages\":\"88-112\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2017-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Pharmeuropa bio & scientific notes\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"Medicine\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Pharmeuropa bio & scientific notes","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"Medicine","Score":null,"Total":0}
Assessment of UV spectrophotometry for determination of plasmid DNA concentration in vector preparations for human gene therapy products.
The European Pharmacopoeia (Ph. Eur.) general chapter 5.14. Gene transfer medicinal products for human use suggests the use of absorbance measurements at 260 nm to determine the DNA concentration of plasmid vectors used for the preparation of gene therapy products for human use. An international collaborative study was organised by the European Directorate for the Quality of Medicines & HealthCare (EDQM) to confirm the suitability of UV spectrophotometry for the quantification of plasmid vectors used in gene therapy (GT). Three Official Medicine Control Laboratories (OMCLs of the European OMCL Network) and members of the OMCL Working Group for GT products took part in the study, in which various types of spectrophotometers were assessed using common test samples. Results of the study demonstrated that UV spectrophotometry can be considered suitable for the quantification of plasmid DNA in GT products regardless of the instrument used.