{"title":"用安特里插管代替气管管?和管交换器]。","authors":"Tetsuro Kimura, Hirokazu Furuuchi, Kensuke Kobayashi, Soichiro Mimuro, Akira Suzuki, Yoshiki Nakajima","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>A tracheal tube can be safely replaced by using a tube exchanger (TE). However, only a thin TE can be used to replace a double-lumen tracheal tube (DLT) with a standard single-lumen tracheal tube (SLT). We successfully replaced a DLT to a SLT by inserting an Aintree Intubation Catheter® (AIC) over a TE in two cases. The AIC (diameter : 19 Fr, overall length : 56 cm) is mainly used for the replacement of various supra- glottic apparatuses using a SLT. In our cases, an AIC with an internal diameter of 4.7 mm was placed over a thin TE with an external diameter of 3.7 mm (11 Fr) to increase the support, and the difference between the SLT with an internal diameter of 7.5 mm and an AIC with an external diameter of 6.3 mm (19 Fr) was decreased, resulting in smooth replacement of the tubes. Even for those cases in which tube replacement might be difficult, acute administration of oxygen could be provided using an AIC with a larger internal lumenthan TE. In conclusion, replacement of a DLT with a SLT i safe and useful through the concomitant use of an AI( and a TE.</p>","PeriodicalId":18254,"journal":{"name":"Masui. The Japanese journal of anesthesiology","volume":"66 3","pages":"322-326"},"PeriodicalIF":0.0000,"publicationDate":"2017-03-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"[Replacement of a Tracheal Tube by Concomitant Use of an Aintree Intubation Catheter? and a Tube Exchanger].\",\"authors\":\"Tetsuro Kimura, Hirokazu Furuuchi, Kensuke Kobayashi, Soichiro Mimuro, Akira Suzuki, Yoshiki Nakajima\",\"doi\":\"\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>A tracheal tube can be safely replaced by using a tube exchanger (TE). However, only a thin TE can be used to replace a double-lumen tracheal tube (DLT) with a standard single-lumen tracheal tube (SLT). We successfully replaced a DLT to a SLT by inserting an Aintree Intubation Catheter® (AIC) over a TE in two cases. The AIC (diameter : 19 Fr, overall length : 56 cm) is mainly used for the replacement of various supra- glottic apparatuses using a SLT. In our cases, an AIC with an internal diameter of 4.7 mm was placed over a thin TE with an external diameter of 3.7 mm (11 Fr) to increase the support, and the difference between the SLT with an internal diameter of 7.5 mm and an AIC with an external diameter of 6.3 mm (19 Fr) was decreased, resulting in smooth replacement of the tubes. Even for those cases in which tube replacement might be difficult, acute administration of oxygen could be provided using an AIC with a larger internal lumenthan TE. In conclusion, replacement of a DLT with a SLT i safe and useful through the concomitant use of an AI( and a TE.</p>\",\"PeriodicalId\":18254,\"journal\":{\"name\":\"Masui. The Japanese journal of anesthesiology\",\"volume\":\"66 3\",\"pages\":\"322-326\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2017-03-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Masui. The Japanese journal of anesthesiology\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Masui. The Japanese journal of anesthesiology","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
摘要
气管管可以通过使用管交换器(TE)安全地更换。然而,只有薄TE可以用标准的单腔气管管(SLT)代替双腔气管管(DLT)。我们在两个病例中通过在TE上插入Aintree插管®(AIC)成功地将DLT替换为SLT。AIC(直径:19 Fr,全长:56 cm)主要用于使用SLT替换各种声门上器械。在我们的案例中,将内径4.7 mm的AIC放置在外径3.7 mm (11 Fr)的薄TE上以增加支撑,并且减小了内径7.5 mm的SLT与外径6.3 mm (19 Fr)的AIC之间的差异,从而顺利更换管。即使对于那些可能难以更换导管的病例,也可以使用具有比TE更大内腔的AIC来提供急性给氧。总之,通过联合使用AI(和TE),用SLT替换DLT是安全且有用的。
[Replacement of a Tracheal Tube by Concomitant Use of an Aintree Intubation Catheter? and a Tube Exchanger].
A tracheal tube can be safely replaced by using a tube exchanger (TE). However, only a thin TE can be used to replace a double-lumen tracheal tube (DLT) with a standard single-lumen tracheal tube (SLT). We successfully replaced a DLT to a SLT by inserting an Aintree Intubation Catheter® (AIC) over a TE in two cases. The AIC (diameter : 19 Fr, overall length : 56 cm) is mainly used for the replacement of various supra- glottic apparatuses using a SLT. In our cases, an AIC with an internal diameter of 4.7 mm was placed over a thin TE with an external diameter of 3.7 mm (11 Fr) to increase the support, and the difference between the SLT with an internal diameter of 7.5 mm and an AIC with an external diameter of 6.3 mm (19 Fr) was decreased, resulting in smooth replacement of the tubes. Even for those cases in which tube replacement might be difficult, acute administration of oxygen could be provided using an AIC with a larger internal lumenthan TE. In conclusion, replacement of a DLT with a SLT i safe and useful through the concomitant use of an AI( and a TE.