A. Al-Zain, I. Alshehri, H.MH. Jamalellail, R. Price
{"title":"电池放电对预算光固化装置输出的影响","authors":"A. Al-Zain, I. Alshehri, H.MH. Jamalellail, R. Price","doi":"10.1055/s-0042-1757161","DOIUrl":null,"url":null,"abstract":"\n Objectives The manufacturers of budget light-curing units (LCUs) often claim to provide high-quality units that are equivalent to LCUs from major manufacturers. This study investigated the effects of battery discharge on the light output from different budget LCUs compared to a major manufacturer.\n Materials and Methods Two brands of budget LCUs (LY-A180 and LED-CL) were compared to a control LCU from a major manufacturer (3M). The LCUs were fully charged, and their light outputs were measured over one battery discharge cycle using repeated 10-second exposures at a 0-mm distance.\n Statistical Analysis Data were analyzed using one-way analysis of variance and Bonferroni post-hoc test.\n Results The budget LCUs delivered fluctuating light output values. In their first exposure, the budget LCUs delivered between 205 and 444 mW power, an irradiance between 533 and 1154 mW/cm2, and a radiant exposure between 5.3 and 11.5 J/cm2. As the number of exposures increased, their light output decreased between 24 and 81%, while the control LCU showed only a 4.9% decrease in power and irradiance. The light outputs from the budget LCUs were significantly less than the control LCU, and they were significantly from each other.\n Conclusion The budget LCUs tested could not maintain their power, irradiance, and radiant exposure output values as the battery discharged. This supports the recommendation that clinicians should be very cautious when using budget LCUs in their clinical practice.","PeriodicalId":37771,"journal":{"name":"European Journal of General Dentistry","volume":" ","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2022-05-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Effect of Battery Discharge on the Output from Budget Light-Curing Units\",\"authors\":\"A. Al-Zain, I. Alshehri, H.MH. Jamalellail, R. Price\",\"doi\":\"10.1055/s-0042-1757161\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"\\n Objectives The manufacturers of budget light-curing units (LCUs) often claim to provide high-quality units that are equivalent to LCUs from major manufacturers. This study investigated the effects of battery discharge on the light output from different budget LCUs compared to a major manufacturer.\\n Materials and Methods Two brands of budget LCUs (LY-A180 and LED-CL) were compared to a control LCU from a major manufacturer (3M). The LCUs were fully charged, and their light outputs were measured over one battery discharge cycle using repeated 10-second exposures at a 0-mm distance.\\n Statistical Analysis Data were analyzed using one-way analysis of variance and Bonferroni post-hoc test.\\n Results The budget LCUs delivered fluctuating light output values. In their first exposure, the budget LCUs delivered between 205 and 444 mW power, an irradiance between 533 and 1154 mW/cm2, and a radiant exposure between 5.3 and 11.5 J/cm2. As the number of exposures increased, their light output decreased between 24 and 81%, while the control LCU showed only a 4.9% decrease in power and irradiance. The light outputs from the budget LCUs were significantly less than the control LCU, and they were significantly from each other.\\n Conclusion The budget LCUs tested could not maintain their power, irradiance, and radiant exposure output values as the battery discharged. This supports the recommendation that clinicians should be very cautious when using budget LCUs in their clinical practice.\",\"PeriodicalId\":37771,\"journal\":{\"name\":\"European Journal of General Dentistry\",\"volume\":\" \",\"pages\":\"\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2022-05-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"European Journal of General Dentistry\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1055/s-0042-1757161\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"Dentistry\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"European Journal of General Dentistry","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1055/s-0042-1757161","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"Dentistry","Score":null,"Total":0}
Effect of Battery Discharge on the Output from Budget Light-Curing Units
Objectives The manufacturers of budget light-curing units (LCUs) often claim to provide high-quality units that are equivalent to LCUs from major manufacturers. This study investigated the effects of battery discharge on the light output from different budget LCUs compared to a major manufacturer.
Materials and Methods Two brands of budget LCUs (LY-A180 and LED-CL) were compared to a control LCU from a major manufacturer (3M). The LCUs were fully charged, and their light outputs were measured over one battery discharge cycle using repeated 10-second exposures at a 0-mm distance.
Statistical Analysis Data were analyzed using one-way analysis of variance and Bonferroni post-hoc test.
Results The budget LCUs delivered fluctuating light output values. In their first exposure, the budget LCUs delivered between 205 and 444 mW power, an irradiance between 533 and 1154 mW/cm2, and a radiant exposure between 5.3 and 11.5 J/cm2. As the number of exposures increased, their light output decreased between 24 and 81%, while the control LCU showed only a 4.9% decrease in power and irradiance. The light outputs from the budget LCUs were significantly less than the control LCU, and they were significantly from each other.
Conclusion The budget LCUs tested could not maintain their power, irradiance, and radiant exposure output values as the battery discharged. This supports the recommendation that clinicians should be very cautious when using budget LCUs in their clinical practice.
期刊介绍:
European Journal of General Dentistry (EJGD) is one of the leading open-access international dental journal within the field of Dentistry. The aim of EJGD is publishing novel and high-quality research papers, as well as to influence the practice of dentistry at clinician, research, industry and policy-maker level on an international basis. EJGD publishes articles on all disciplines of dentistry including the cariology, orthodontics, oral surgery, preventive dentistry, periodontology, endodontology, operative dentistry, fixed and removable prosthodontics, dental biomaterials science, long-term clinical trials including epidemiology and oral health, technology transfer of new scientific instrumentation or procedures, as well as clinically relevant oral biology and translational research.Moreover, EJGD also publish the scientific researches evaluating the use of new biomaterials, new drugs and new methods for treatment of patients with different kinds of oral and maxillofacial diseases or defects, the diagnosis of oral and maxillofacial diseases with new methods, etc. Moreover, researches on the quality of life, psychological interventions, improving disease treatment outcomes, the prevention, diagnosis and management of cancer therapeutic complications, rehabilitation, palliative and end of life care, and support teamwork for cancer care and oral health care for old patients are also welcome. EJGD publishes research articles, case reports, reviews and comparison studies evaluating materials and methods in the all fields of related to dentistry.