{"title":"加速转速和转子材料对植入式微输液泵管材寿命影响的实验研究","authors":"S. Nair, A. D., Sudheesh K S","doi":"10.1080/03091902.2022.2082575","DOIUrl":null,"url":null,"abstract":"Abstract Peristaltic pumps have been put to use in various biomedical applications like devices for the transfer of body fluids as well as devices for controlled release of medication, including implantable infusion pumps. Out of the various components of a peristaltic pump, tubing is considered the most vulnerable part. This study focuses on the performance of Silicone micro-pump tubing used in such an implantable drug delivery device. Long-term implantable medical devices are expected to be operational for about 10 years. But experimental testing of the reliability of components under normal working speeds are time-consuming and thus delays the product development cycle. While simulating the conditions in the laboratory under accelerated speeds, the effect of increasing the speed must be accounted. In this study, the effect of accelerated speed and rotor material on pump tubing life is investigated. A test jig is developed which simulates the running conditions of the infusion pump for long-duration operation. Different rotor speeds and material configurations are investigated to obtain their effect on long-duration performance. Thermal effects on the roller junctions are studied and found that the Delrin silicone combination has twice the rise in junction temperature than the titanium silicone combination. The failure modes are inspected using microstructure analysis and the best configuration is identified.","PeriodicalId":39637,"journal":{"name":"Journal of Medical Engineering and Technology","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2022-06-17","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Experimental investigation on effect of accelerated speed and rotor material on life of implantable micro-infusion pump tubing\",\"authors\":\"S. Nair, A. D., Sudheesh K S\",\"doi\":\"10.1080/03091902.2022.2082575\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Abstract Peristaltic pumps have been put to use in various biomedical applications like devices for the transfer of body fluids as well as devices for controlled release of medication, including implantable infusion pumps. Out of the various components of a peristaltic pump, tubing is considered the most vulnerable part. This study focuses on the performance of Silicone micro-pump tubing used in such an implantable drug delivery device. Long-term implantable medical devices are expected to be operational for about 10 years. But experimental testing of the reliability of components under normal working speeds are time-consuming and thus delays the product development cycle. While simulating the conditions in the laboratory under accelerated speeds, the effect of increasing the speed must be accounted. In this study, the effect of accelerated speed and rotor material on pump tubing life is investigated. A test jig is developed which simulates the running conditions of the infusion pump for long-duration operation. Different rotor speeds and material configurations are investigated to obtain their effect on long-duration performance. Thermal effects on the roller junctions are studied and found that the Delrin silicone combination has twice the rise in junction temperature than the titanium silicone combination. The failure modes are inspected using microstructure analysis and the best configuration is identified.\",\"PeriodicalId\":39637,\"journal\":{\"name\":\"Journal of Medical Engineering and Technology\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2022-06-17\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Medical Engineering and Technology\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1080/03091902.2022.2082575\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"Engineering\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Medical Engineering and Technology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1080/03091902.2022.2082575","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"Engineering","Score":null,"Total":0}
Experimental investigation on effect of accelerated speed and rotor material on life of implantable micro-infusion pump tubing
Abstract Peristaltic pumps have been put to use in various biomedical applications like devices for the transfer of body fluids as well as devices for controlled release of medication, including implantable infusion pumps. Out of the various components of a peristaltic pump, tubing is considered the most vulnerable part. This study focuses on the performance of Silicone micro-pump tubing used in such an implantable drug delivery device. Long-term implantable medical devices are expected to be operational for about 10 years. But experimental testing of the reliability of components under normal working speeds are time-consuming and thus delays the product development cycle. While simulating the conditions in the laboratory under accelerated speeds, the effect of increasing the speed must be accounted. In this study, the effect of accelerated speed and rotor material on pump tubing life is investigated. A test jig is developed which simulates the running conditions of the infusion pump for long-duration operation. Different rotor speeds and material configurations are investigated to obtain their effect on long-duration performance. Thermal effects on the roller junctions are studied and found that the Delrin silicone combination has twice the rise in junction temperature than the titanium silicone combination. The failure modes are inspected using microstructure analysis and the best configuration is identified.
期刊介绍:
The Journal of Medical Engineering & Technology is an international, independent, multidisciplinary, bimonthly journal promoting an understanding of the physiological processes underlying disease processes and the appropriate application of technology. Features include authoritative review papers, the reporting of original research, and evaluation reports on new and existing techniques and devices. Each issue of the journal contains a comprehensive information service which provides news relevant to the world of medical technology, details of new products, book reviews, and selected contents of related journals.