Yasaman Ardeshirpour PhD, Ethan D. Cohen PhD, Seth J. Seidman MSc, Biniyam Taddese PhD, Tayeb Zaidi MSc, Howard Bassen MSc
{"title":"低频安全系统直接电压感应对神经刺激器导线的影响。","authors":"Yasaman Ardeshirpour PhD, Ethan D. Cohen PhD, Seth J. Seidman MSc, Biniyam Taddese PhD, Tayeb Zaidi MSc, Howard Bassen MSc","doi":"10.1002/bem.22485","DOIUrl":null,"url":null,"abstract":"<p>Low-frequency (LF) security systems, such as antitheft electronic article surveillance (EAS) gates emit strong magnetic fields that could potentially interfere with neurostimulator operation. Some patients reported pain and shocking sensations near EAS gates, even after they turned off their pulse generator. To investigate the direct voltage induction of EAS systems on neurostimulator leads, we evaluated voltages induced by two EAS systems (14 kHz continuous wave or 58 kHz pulsed) on a 40 cm sacral neurostimulator lead formed in a circular loop attached to a pulse generator that was turned off. The lead and neurostimulator were mounted in a saline-filled rectangular phantom placed within electromagnetic fields emitted by EAS systems. The measured voltage waveforms were applied to computational models of spinal nerve axons to predict whether these voltages may evoke action potentials. Additional in vitro testing was performed on the semicircular lead geometry, to study the effect of lead geometry on EAS induced voltages. While standard neurostimulator testing per ISO 14708-3:2017 recommends electromagnetic compatibility testing with LF magnetic fields for induction of malfunctions of the active electronic circuitry while generating intended stimulating pulses, our results show that close to the EAS antenna frames, the induced voltage on the lead could be strong enough to evoke action potentials, even with the pulse generator turned off. This work suggests that patient reports of pain and shocking sensations when near EAS systems could also be correlated with the direct EAS-induced voltage on neurostimulator lead.</p>","PeriodicalId":8956,"journal":{"name":"Bioelectromagnetics","volume":null,"pages":null},"PeriodicalIF":1.8000,"publicationDate":"2023-10-23","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Effect of direct voltage induction by low-frequency security systems on neurostimulator lead\",\"authors\":\"Yasaman Ardeshirpour PhD, Ethan D. Cohen PhD, Seth J. Seidman MSc, Biniyam Taddese PhD, Tayeb Zaidi MSc, Howard Bassen MSc\",\"doi\":\"10.1002/bem.22485\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p>Low-frequency (LF) security systems, such as antitheft electronic article surveillance (EAS) gates emit strong magnetic fields that could potentially interfere with neurostimulator operation. Some patients reported pain and shocking sensations near EAS gates, even after they turned off their pulse generator. To investigate the direct voltage induction of EAS systems on neurostimulator leads, we evaluated voltages induced by two EAS systems (14 kHz continuous wave or 58 kHz pulsed) on a 40 cm sacral neurostimulator lead formed in a circular loop attached to a pulse generator that was turned off. The lead and neurostimulator were mounted in a saline-filled rectangular phantom placed within electromagnetic fields emitted by EAS systems. The measured voltage waveforms were applied to computational models of spinal nerve axons to predict whether these voltages may evoke action potentials. Additional in vitro testing was performed on the semicircular lead geometry, to study the effect of lead geometry on EAS induced voltages. While standard neurostimulator testing per ISO 14708-3:2017 recommends electromagnetic compatibility testing with LF magnetic fields for induction of malfunctions of the active electronic circuitry while generating intended stimulating pulses, our results show that close to the EAS antenna frames, the induced voltage on the lead could be strong enough to evoke action potentials, even with the pulse generator turned off. This work suggests that patient reports of pain and shocking sensations when near EAS systems could also be correlated with the direct EAS-induced voltage on neurostimulator lead.</p>\",\"PeriodicalId\":8956,\"journal\":{\"name\":\"Bioelectromagnetics\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":1.8000,\"publicationDate\":\"2023-10-23\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Bioelectromagnetics\",\"FirstCategoryId\":\"99\",\"ListUrlMain\":\"https://onlinelibrary.wiley.com/doi/10.1002/bem.22485\",\"RegionNum\":3,\"RegionCategory\":\"生物学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"BIOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Bioelectromagnetics","FirstCategoryId":"99","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1002/bem.22485","RegionNum":3,"RegionCategory":"生物学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"BIOLOGY","Score":null,"Total":0}
Effect of direct voltage induction by low-frequency security systems on neurostimulator lead
Low-frequency (LF) security systems, such as antitheft electronic article surveillance (EAS) gates emit strong magnetic fields that could potentially interfere with neurostimulator operation. Some patients reported pain and shocking sensations near EAS gates, even after they turned off their pulse generator. To investigate the direct voltage induction of EAS systems on neurostimulator leads, we evaluated voltages induced by two EAS systems (14 kHz continuous wave or 58 kHz pulsed) on a 40 cm sacral neurostimulator lead formed in a circular loop attached to a pulse generator that was turned off. The lead and neurostimulator were mounted in a saline-filled rectangular phantom placed within electromagnetic fields emitted by EAS systems. The measured voltage waveforms were applied to computational models of spinal nerve axons to predict whether these voltages may evoke action potentials. Additional in vitro testing was performed on the semicircular lead geometry, to study the effect of lead geometry on EAS induced voltages. While standard neurostimulator testing per ISO 14708-3:2017 recommends electromagnetic compatibility testing with LF magnetic fields for induction of malfunctions of the active electronic circuitry while generating intended stimulating pulses, our results show that close to the EAS antenna frames, the induced voltage on the lead could be strong enough to evoke action potentials, even with the pulse generator turned off. This work suggests that patient reports of pain and shocking sensations when near EAS systems could also be correlated with the direct EAS-induced voltage on neurostimulator lead.
期刊介绍:
Bioelectromagnetics is published by Wiley-Liss, Inc., for the Bioelectromagnetics Society and is the official journal of the Bioelectromagnetics Society and the European Bioelectromagnetics Association. It is a peer-reviewed, internationally circulated scientific journal that specializes in reporting original data on biological effects and applications of electromagnetic fields that range in frequency from zero hertz (static fields) to the terahertz undulations and visible light. Both experimental and clinical data are of interest to the journal''s readers as are theoretical papers or reviews that offer novel insights into or criticism of contemporary concepts and theories of field-body interactions. The Bioelectromagnetics Society, which sponsors the journal, also welcomes experimental or clinical papers on the domains of sonic and ultrasonic radiation.