IF 1.5 4区 医学 Q3 SURGERY Computer Assisted Surgery Pub Date : 2016-10-25 DOI:10.1080/24699322.2016.1240299
Ho Chang, C. Chen, Minghanbo Liu, Li-Ke Chien, Tung-Sun Huang, Bo-Hsiang Tu
{"title":"一次性无线电子内窥镜的制造与表征","authors":"Ho Chang, C. Chen, Minghanbo Liu, Li-Ke Chien, Tung-Sun Huang, Bo-Hsiang Tu","doi":"10.1080/24699322.2016.1240299","DOIUrl":null,"url":null,"abstract":"Abstract A wireless electronic endoscope that transmits signals by Wi-Fi is developed for use in single-hole endoscopic surgery and future application to natural orifice surgery. The innovative electronic endoscope developed in this study has a disposable design, completely preventing patents from taking risks of cross infection. The main components of the endoscope are a disposable component of the front tip and the hand-held part of the rear tip. The front tip consists of a lens, metallic tube, and electronic joint whereas the rear tip consists of a power switch, image converter, wireless transceiver chip, and antenna. Experimental applications of the wireless electronic endoscope show that after 3 hours and 30 minutes of normal use, the temperature of metallic tube at the front tip averages 45 °C. Regarding the measured luminance result, when the focal distance is 5 cm, luminance reaches 1410 lux. In an indoor environment, when the measured distance is 3 m and there is no disruptor, the Wi-Fi transmitted signal of the endoscope is at strength −51 dBm. If the disruptor is established between the sending and receiving ends, signal strength is −53 dBm. Increasing the measured distance to 8 m obtains a signal strength of −74 dBm. Test practical applications indicate that the wireless Wi-Fi electronic endoscope fully meets the needs of surgery room.","PeriodicalId":56051,"journal":{"name":"Computer Assisted Surgery","volume":"21 1","pages":"124 - 131"},"PeriodicalIF":1.5000,"publicationDate":"2016-10-25","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1080/24699322.2016.1240299","citationCount":"1","resultStr":"{\"title\":\"Fabrication and characterization of disposable wireless electronic endoscope\",\"authors\":\"Ho Chang, C. Chen, Minghanbo Liu, Li-Ke Chien, Tung-Sun Huang, Bo-Hsiang Tu\",\"doi\":\"10.1080/24699322.2016.1240299\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Abstract A wireless electronic endoscope that transmits signals by Wi-Fi is developed for use in single-hole endoscopic surgery and future application to natural orifice surgery. The innovative electronic endoscope developed in this study has a disposable design, completely preventing patents from taking risks of cross infection. The main components of the endoscope are a disposable component of the front tip and the hand-held part of the rear tip. The front tip consists of a lens, metallic tube, and electronic joint whereas the rear tip consists of a power switch, image converter, wireless transceiver chip, and antenna. Experimental applications of the wireless electronic endoscope show that after 3 hours and 30 minutes of normal use, the temperature of metallic tube at the front tip averages 45 °C. Regarding the measured luminance result, when the focal distance is 5 cm, luminance reaches 1410 lux. In an indoor environment, when the measured distance is 3 m and there is no disruptor, the Wi-Fi transmitted signal of the endoscope is at strength −51 dBm. If the disruptor is established between the sending and receiving ends, signal strength is −53 dBm. Increasing the measured distance to 8 m obtains a signal strength of −74 dBm. Test practical applications indicate that the wireless Wi-Fi electronic endoscope fully meets the needs of surgery room.\",\"PeriodicalId\":56051,\"journal\":{\"name\":\"Computer Assisted Surgery\",\"volume\":\"21 1\",\"pages\":\"124 - 131\"},\"PeriodicalIF\":1.5000,\"publicationDate\":\"2016-10-25\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://sci-hub-pdf.com/10.1080/24699322.2016.1240299\",\"citationCount\":\"1\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Computer Assisted Surgery\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1080/24699322.2016.1240299\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"SURGERY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Computer Assisted Surgery","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1080/24699322.2016.1240299","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"SURGERY","Score":null,"Total":0}
引用次数: 1

摘要

开发了一种通过Wi-Fi传输信号的无线电子内窥镜,用于单孔内窥镜手术,未来将应用于自然口手术。本研究开发的创新电子内窥镜采用一次性设计,完全避免了专利交叉感染的风险。内窥镜的主要部件是前尖端的一次性部件和后尖端的手持部分。前尖端由透镜、金属管和电子接头组成,后尖端由电源开关、图像转换器、无线收发芯片和天线组成。无线电子内窥镜的实验应用表明,正常使用3小时30分钟后,前尖端金属管的平均温度为45℃。对于测量亮度结果,当焦距为5 cm时,亮度达到1410勒克斯。在室内环境下,当测量距离为3m且无干扰时,内窥镜的Wi-Fi传输信号强度为−51 dBm。如果在发送端和接收端之间建立干扰,信号强度为- 53dbm。将测量距离增加到8m,信号强度为−74dbm。测试实际应用表明,无线Wi-Fi电子内窥镜完全满足手术室的需要。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Fabrication and characterization of disposable wireless electronic endoscope
Abstract A wireless electronic endoscope that transmits signals by Wi-Fi is developed for use in single-hole endoscopic surgery and future application to natural orifice surgery. The innovative electronic endoscope developed in this study has a disposable design, completely preventing patents from taking risks of cross infection. The main components of the endoscope are a disposable component of the front tip and the hand-held part of the rear tip. The front tip consists of a lens, metallic tube, and electronic joint whereas the rear tip consists of a power switch, image converter, wireless transceiver chip, and antenna. Experimental applications of the wireless electronic endoscope show that after 3 hours and 30 minutes of normal use, the temperature of metallic tube at the front tip averages 45 °C. Regarding the measured luminance result, when the focal distance is 5 cm, luminance reaches 1410 lux. In an indoor environment, when the measured distance is 3 m and there is no disruptor, the Wi-Fi transmitted signal of the endoscope is at strength −51 dBm. If the disruptor is established between the sending and receiving ends, signal strength is −53 dBm. Increasing the measured distance to 8 m obtains a signal strength of −74 dBm. Test practical applications indicate that the wireless Wi-Fi electronic endoscope fully meets the needs of surgery room.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Computer Assisted Surgery
Computer Assisted Surgery Medicine-Surgery
CiteScore
2.30
自引率
0.00%
发文量
13
审稿时长
10 weeks
期刊介绍: omputer Assisted Surgery aims to improve patient care by advancing the utilization of computers during treatment; to evaluate the benefits and risks associated with the integration of advanced digital technologies into surgical practice; to disseminate clinical and basic research relevant to stereotactic surgery, minimal access surgery, endoscopy, and surgical robotics; to encourage interdisciplinary collaboration between engineers and physicians in developing new concepts and applications; to educate clinicians about the principles and techniques of computer assisted surgery and therapeutics; and to serve the international scientific community as a medium for the transfer of new information relating to theory, research, and practice in biomedical imaging and the surgical specialties. The scope of Computer Assisted Surgery encompasses all fields within surgery, as well as biomedical imaging and instrumentation, and digital technology employed as an adjunct to imaging in diagnosis, therapeutics, and surgery. Topics featured include frameless as well as conventional stereotactic procedures, surgery guided by intraoperative ultrasound or magnetic resonance imaging, image guided focused irradiation, robotic surgery, and any therapeutic interventions performed with the use of digital imaging technology.
期刊最新文献
Ultrasound-based 3D bone modelling in computer assisted orthopedic surgery - a review and future challenges. Augmented reality technology shortens aneurysm surgery learning curve for residents. Feasibility of proton dosimetry overriding planning CT with daily CBCT elaborated through generative artificial intelligence tools. SwinD-Net: a lightweight segmentation network for laparoscopic liver segmentation. Risk prediction and analysis of gallbladder polyps with deep neural network.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1