{"title":"手持190+190行列寻址CMUT探针体积成像","authors":"Rune Sixten Grass;Mathias Engholm;Andreas Spandet Havreland;Christopher Beers;Martin Lind Ommen;Stine Løvholt Grue Pedersen;Lars Nordahl Moesner;Matthias Bo Stuart;Mudabbir Tufail Bhatti;Borislav G. Tomov;Jørgen Arendt Jensen;Erik Vilain Thomsen","doi":"10.1109/OJUFFC.2022.3213013","DOIUrl":null,"url":null,"abstract":"This paper presents the design, fabrication, and characterization of a 190+190 row-column addressed (RCA) capacitive micromachined ultrasonic transducer (CMUT) array integrated in a custom hand-held probe handle. The array has a designed 4.5 MHz center frequency in immersion and a pitch of \n<inline-formula> <tex-math>$95~\\mu $ </tex-math></inline-formula>\n m which corresponds to \n<inline-formula> <tex-math>$\\approx ~\\lambda $ </tex-math></inline-formula>\n/4. The array has a \n<inline-formula> <tex-math>$2.14\\times2.14$ </tex-math></inline-formula>\n cm\n<inline-formula> <tex-math>$^{2}$ </tex-math></inline-formula>\n footprint including an integrated apodization scheme to reduce ghost echoes when performing ultrasound imaging. The array was fabricated using a combination of fusion and anodic bonding, and a deposit, remove, etch, multistep (DREM) etch to reduce substrate coupling and improve electrode conductivity. The transducer array was wire-bonded to a rigid-flex printed circuit board (PCB), encapsulated in room temperature vulcanizing (RTV) silicone polymer, electromagnetic interference (EMI) shielded, and mounted in a 3D-milled PPSU probe handle. The probe was characterized using the SARUS experimental scanner and 3D volumetric imaging was demonstrated on scatter and wire phantoms. The imaging depth was derived from tissue mimicking phantom measurements (0.5 dB MHz\n<inline-formula> <tex-math>$^{-1} \\text{cm}^{-1}$ </tex-math></inline-formula>\n attenuation) by estimating the SNR at varying depths. For a synthetic aperture imaging sequence with 96+96 emissions the imaging depth was 3.6 cm. The center frequency measured from the impulse response spectra in transmit and pulse-echo was 6.0 ± 0.9 MHz and 5.3 ± 0.4 MHz, and the corresponding relative bandwidths were 62.8 ± 4.5 % and 86.2 ± 10.4 %. The fabrication process showed clear improvement in relative receive sensitivity and transmit pressure uniformity compared to earlier silicon-on-insulator (SOI) based designs. However, at the same time it presented yield problems resulting in only around 55 % elements with a good response.","PeriodicalId":73301,"journal":{"name":"IEEE open journal of ultrasonics, ferroelectrics, and frequency control","volume":"2 ","pages":"220-236"},"PeriodicalIF":0.0000,"publicationDate":"2022-10-10","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://ieeexplore.ieee.org/stamp/stamp.jsp?tp=&arnumber=9913981","citationCount":"1","resultStr":"{\"title\":\"A Hand-Held 190+190 Row–Column Addressed CMUT Probe for Volumetric Imaging\",\"authors\":\"Rune Sixten Grass;Mathias Engholm;Andreas Spandet Havreland;Christopher Beers;Martin Lind Ommen;Stine Løvholt Grue Pedersen;Lars Nordahl Moesner;Matthias Bo Stuart;Mudabbir Tufail Bhatti;Borislav G. Tomov;Jørgen Arendt Jensen;Erik Vilain Thomsen\",\"doi\":\"10.1109/OJUFFC.2022.3213013\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"This paper presents the design, fabrication, and characterization of a 190+190 row-column addressed (RCA) capacitive micromachined ultrasonic transducer (CMUT) array integrated in a custom hand-held probe handle. The array has a designed 4.5 MHz center frequency in immersion and a pitch of \\n<inline-formula> <tex-math>$95~\\\\mu $ </tex-math></inline-formula>\\n m which corresponds to \\n<inline-formula> <tex-math>$\\\\approx ~\\\\lambda $ </tex-math></inline-formula>\\n/4. The array has a \\n<inline-formula> <tex-math>$2.14\\\\times2.14$ </tex-math></inline-formula>\\n cm\\n<inline-formula> <tex-math>$^{2}$ </tex-math></inline-formula>\\n footprint including an integrated apodization scheme to reduce ghost echoes when performing ultrasound imaging. The array was fabricated using a combination of fusion and anodic bonding, and a deposit, remove, etch, multistep (DREM) etch to reduce substrate coupling and improve electrode conductivity. The transducer array was wire-bonded to a rigid-flex printed circuit board (PCB), encapsulated in room temperature vulcanizing (RTV) silicone polymer, electromagnetic interference (EMI) shielded, and mounted in a 3D-milled PPSU probe handle. The probe was characterized using the SARUS experimental scanner and 3D volumetric imaging was demonstrated on scatter and wire phantoms. The imaging depth was derived from tissue mimicking phantom measurements (0.5 dB MHz\\n<inline-formula> <tex-math>$^{-1} \\\\text{cm}^{-1}$ </tex-math></inline-formula>\\n attenuation) by estimating the SNR at varying depths. For a synthetic aperture imaging sequence with 96+96 emissions the imaging depth was 3.6 cm. The center frequency measured from the impulse response spectra in transmit and pulse-echo was 6.0 ± 0.9 MHz and 5.3 ± 0.4 MHz, and the corresponding relative bandwidths were 62.8 ± 4.5 % and 86.2 ± 10.4 %. The fabrication process showed clear improvement in relative receive sensitivity and transmit pressure uniformity compared to earlier silicon-on-insulator (SOI) based designs. However, at the same time it presented yield problems resulting in only around 55 % elements with a good response.\",\"PeriodicalId\":73301,\"journal\":{\"name\":\"IEEE open journal of ultrasonics, ferroelectrics, and frequency control\",\"volume\":\"2 \",\"pages\":\"220-236\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2022-10-10\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://ieeexplore.ieee.org/stamp/stamp.jsp?tp=&arnumber=9913981\",\"citationCount\":\"1\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"IEEE open journal of ultrasonics, ferroelectrics, and frequency control\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://ieeexplore.ieee.org/document/9913981/\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"IEEE open journal of ultrasonics, ferroelectrics, and frequency control","FirstCategoryId":"1085","ListUrlMain":"https://ieeexplore.ieee.org/document/9913981/","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 1
摘要
本文介绍了集成在定制手持探头手柄中的190+190行列寻址(RCA)电容式微机械超声换能器(CMUT)阵列的设计、制造和表征。该阵列的设计浸入中心频率为4.5 MHz,节距为$95~\mu $ m,对应于$\approx ~\lambda $ /4。该阵列的占地面积为$2.14\times2.14$ cm $^{2}$,包括一个集成的apodization方案,以减少执行超声成像时的鬼回波。该阵列采用融合和阳极键合的组合,以及沉积、去除、蚀刻、多步骤(DREM)蚀刻来减少衬底耦合并提高电极导电性。换能器阵列通过导线连接到刚性柔性印刷电路板(PCB)上,封装在室温硫化(RTV)有机硅聚合物中,屏蔽电磁干扰(EMI),并安装在3d研磨的PPSU探头手柄中。使用SARUS实验扫描仪对探针进行了表征,并在散射和丝影上进行了三维体积成像。成像深度是通过估计不同深度下的信噪比,从组织模拟幻象测量(0.5 dB MHz $^{-1} \text{cm}^{-1}$衰减)得出的。96+96次发射的合成孔径成像序列,成像深度为3.6 cm。发射脉冲响应谱和脉冲回波脉冲响应谱测得的中心频率分别为6.0±0.9 MHz和5.3±0.4 MHz,相应的相对带宽为62.8±4.5 % and 86.2 ± 10.4 %. The fabrication process showed clear improvement in relative receive sensitivity and transmit pressure uniformity compared to earlier silicon-on-insulator (SOI) based designs. However, at the same time it presented yield problems resulting in only around 55 % elements with a good response.
A Hand-Held 190+190 Row–Column Addressed CMUT Probe for Volumetric Imaging
This paper presents the design, fabrication, and characterization of a 190+190 row-column addressed (RCA) capacitive micromachined ultrasonic transducer (CMUT) array integrated in a custom hand-held probe handle. The array has a designed 4.5 MHz center frequency in immersion and a pitch of
$95~\mu $
m which corresponds to
$\approx ~\lambda $
/4. The array has a
$2.14\times2.14$
cm
$^{2}$
footprint including an integrated apodization scheme to reduce ghost echoes when performing ultrasound imaging. The array was fabricated using a combination of fusion and anodic bonding, and a deposit, remove, etch, multistep (DREM) etch to reduce substrate coupling and improve electrode conductivity. The transducer array was wire-bonded to a rigid-flex printed circuit board (PCB), encapsulated in room temperature vulcanizing (RTV) silicone polymer, electromagnetic interference (EMI) shielded, and mounted in a 3D-milled PPSU probe handle. The probe was characterized using the SARUS experimental scanner and 3D volumetric imaging was demonstrated on scatter and wire phantoms. The imaging depth was derived from tissue mimicking phantom measurements (0.5 dB MHz
$^{-1} \text{cm}^{-1}$
attenuation) by estimating the SNR at varying depths. For a synthetic aperture imaging sequence with 96+96 emissions the imaging depth was 3.6 cm. The center frequency measured from the impulse response spectra in transmit and pulse-echo was 6.0 ± 0.9 MHz and 5.3 ± 0.4 MHz, and the corresponding relative bandwidths were 62.8 ± 4.5 % and 86.2 ± 10.4 %. The fabrication process showed clear improvement in relative receive sensitivity and transmit pressure uniformity compared to earlier silicon-on-insulator (SOI) based designs. However, at the same time it presented yield problems resulting in only around 55 % elements with a good response.