{"title":"商用血清肌钙蛋白快速检测试剂盒的性能分析","authors":"Raquel Ida Ferreira, Marcela R. Amaral, Maytê S. R. Brito, Matheus P. Pinto, Ana Carolina R. Maranhão, Lucas Costa, Rómulo Melo","doi":"10.5935/1676-2444.20200020","DOIUrl":null,"url":null,"abstract":"Introduction: Rapid tests represent an important diagnostic tool, providing results in a short period of time and eliminating the use of large automated equipment. Objective: To evaluate the performance of five rapid test kits for troponin I detection in serum. Materials and methods: Samples from 100 patients with suspected acute myocardial infarction (AMI) were selected from a hospital in Barbacena, MG, Brazil. They were tested in the five troponin I commercial rapid test kits and the results were compared with the study reference method (quantitative chemiluminescent immunoassay). Sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV) and correlation coefficients of the rapid tests against the reference method were defined. Results: The kits from Abon ® , Interteck ® , Medtest ® and Wama ® presented the same performance, reaching equal levels of sensitivity (98.08%), specificity (100%), PPV (100%), NPV (97.96%), and correlation coefficient (0.99). The Eco Diagnostica ® kit presented lower sensitivity (82.69%), low correlation coefficient (0.91) and NPV (84.21%), not reaching ideal levels even though lying within the confidence interval. Discussion: The results obtained with the Abon ® , Interteck ® , Medtest ® and Wama ® test kits corroborate pre-existing literature data on the diagnostic accuracy of rapid tests for troponin detection. These tests are allowed for immediate decision making by physicians and can be used to reduce unnecessary hospitalization time and costs associated with a suspected AMI. Conclusion: The kits from Abon ® , Interteck ® , Medtest ® and Wama ® showed excellent diagnostic performance, what makes them an important diagnostic tool in smaller laboratories without automated equipment.","PeriodicalId":35397,"journal":{"name":"Jornal Brasileiro de Patologia e Medicina Laboratorial","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2020-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Performance analysis of commercial rapid tests for serum troponin detection\",\"authors\":\"Raquel Ida Ferreira, Marcela R. Amaral, Maytê S. R. Brito, Matheus P. Pinto, Ana Carolina R. Maranhão, Lucas Costa, Rómulo Melo\",\"doi\":\"10.5935/1676-2444.20200020\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Introduction: Rapid tests represent an important diagnostic tool, providing results in a short period of time and eliminating the use of large automated equipment. Objective: To evaluate the performance of five rapid test kits for troponin I detection in serum. Materials and methods: Samples from 100 patients with suspected acute myocardial infarction (AMI) were selected from a hospital in Barbacena, MG, Brazil. They were tested in the five troponin I commercial rapid test kits and the results were compared with the study reference method (quantitative chemiluminescent immunoassay). Sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV) and correlation coefficients of the rapid tests against the reference method were defined. Results: The kits from Abon ® , Interteck ® , Medtest ® and Wama ® presented the same performance, reaching equal levels of sensitivity (98.08%), specificity (100%), PPV (100%), NPV (97.96%), and correlation coefficient (0.99). The Eco Diagnostica ® kit presented lower sensitivity (82.69%), low correlation coefficient (0.91) and NPV (84.21%), not reaching ideal levels even though lying within the confidence interval. Discussion: The results obtained with the Abon ® , Interteck ® , Medtest ® and Wama ® test kits corroborate pre-existing literature data on the diagnostic accuracy of rapid tests for troponin detection. These tests are allowed for immediate decision making by physicians and can be used to reduce unnecessary hospitalization time and costs associated with a suspected AMI. Conclusion: The kits from Abon ® , Interteck ® , Medtest ® and Wama ® showed excellent diagnostic performance, what makes them an important diagnostic tool in smaller laboratories without automated equipment.\",\"PeriodicalId\":35397,\"journal\":{\"name\":\"Jornal Brasileiro de Patologia e Medicina Laboratorial\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2020-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Jornal Brasileiro de Patologia e Medicina Laboratorial\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.5935/1676-2444.20200020\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"Medicine\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Jornal Brasileiro de Patologia e Medicina Laboratorial","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.5935/1676-2444.20200020","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0

摘要

简介:快速检测是一种重要的诊断工具,可在短时间内提供结果,无需使用大型自动化设备。目的:评价5种血清肌钙蛋白I快速检测试剂盒的性能。材料与方法:选取巴西MG Barbacena某医院100例疑似急性心肌梗死(AMI)患者标本。用5种肌钙蛋白I快速检测试剂盒进行检测,并与研究参考方法(定量化学发光免疫分析法)进行比较。定义快速试验与参考方法的敏感性、特异性、阳性预测值(PPV)、阴性预测值(NPV)和相关系数。结果:Abon®、Interteck®、Medtest®和Wama®的试剂盒表现相同,灵敏度(98.08%)、特异度(100%)、PPV(100%)、NPV(97.96%)、相关系数(0.99)均达到相同水平。Eco diagnostics®试剂盒的灵敏度较低(82.69%),相关系数较低(0.91),NPV(84.21%),即使处于置信区间内,也未达到理想水平。讨论:使用Abon®、Interteck®、Medtest®和Wama®检测试剂盒获得的结果证实了肌钙蛋白检测快速检测诊断准确性的已有文献数据。这些测试允许医生立即做出决定,并可用于减少与疑似AMI相关的不必要住院时间和费用。结论:Abon®、Interteck®、Medtest®和Wama®的试剂盒具有优异的诊断性能,是没有自动化设备的小型实验室的重要诊断工具。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Performance analysis of commercial rapid tests for serum troponin detection
Introduction: Rapid tests represent an important diagnostic tool, providing results in a short period of time and eliminating the use of large automated equipment. Objective: To evaluate the performance of five rapid test kits for troponin I detection in serum. Materials and methods: Samples from 100 patients with suspected acute myocardial infarction (AMI) were selected from a hospital in Barbacena, MG, Brazil. They were tested in the five troponin I commercial rapid test kits and the results were compared with the study reference method (quantitative chemiluminescent immunoassay). Sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV) and correlation coefficients of the rapid tests against the reference method were defined. Results: The kits from Abon ® , Interteck ® , Medtest ® and Wama ® presented the same performance, reaching equal levels of sensitivity (98.08%), specificity (100%), PPV (100%), NPV (97.96%), and correlation coefficient (0.99). The Eco Diagnostica ® kit presented lower sensitivity (82.69%), low correlation coefficient (0.91) and NPV (84.21%), not reaching ideal levels even though lying within the confidence interval. Discussion: The results obtained with the Abon ® , Interteck ® , Medtest ® and Wama ® test kits corroborate pre-existing literature data on the diagnostic accuracy of rapid tests for troponin detection. These tests are allowed for immediate decision making by physicians and can be used to reduce unnecessary hospitalization time and costs associated with a suspected AMI. Conclusion: The kits from Abon ® , Interteck ® , Medtest ® and Wama ® showed excellent diagnostic performance, what makes them an important diagnostic tool in smaller laboratories without automated equipment.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Jornal Brasileiro de Patologia e Medicina Laboratorial
Jornal Brasileiro de Patologia e Medicina Laboratorial Health Professions-Medical Laboratory Technology
CiteScore
1.30
自引率
0.00%
发文量
0
审稿时长
20 weeks
期刊介绍: The Jornal Brasileiro de Patologia e Medicina Laboratorial (Brazilian Journal of Pathology and Laboratory Medicine), a continuation of Jornal Brasileiro de Patologia (Brazilian Journal of Pathology), and published quarterly (March, June, September and December) is directed towards the publication of scientific articles that contribute to the development of the area of Laboratory Medicine (Clinical Pathology, Pathology, Cytopathology). It accepts the following categories of articles: original articles, review articles, case reports, short communications, updating articles, letters to editors and reviews.
期刊最新文献
The influence of tobacco and alcohol in oral cancer: literature review Umbilical cord constriction as a cause of intrauterine fetal death Consequences of late diagnosis paracoccidioidomycosis: case report Primary cutaneous diffuse large B-cell lymphoma, leg type: case report Impact of urinary selective antibiogram in primary care
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1