基于StereoPHAN幻影的SRS MapCHECK作为SBRT计划新预处理系统验证的评估

IF 0.7 Q4 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Polish Journal of Medical Physics and Engineering Pub Date : 2022-05-28 DOI:10.2478/pjmpe-2022-0010
B. Sadowski, Marta Fillmann, Dariusz Szałkowski, P. Kukołowicz
{"title":"基于StereoPHAN幻影的SRS MapCHECK作为SBRT计划新预处理系统验证的评估","authors":"B. Sadowski, Marta Fillmann, Dariusz Szałkowski, P. Kukołowicz","doi":"10.2478/pjmpe-2022-0010","DOIUrl":null,"url":null,"abstract":"Abstract Introduction: The aim of this study was to evaluate the new 2-Dimensional diode array SRS MapCHECK (SunNuclear, Melbourne, USA) with dedicated phantom StereoPHAN (SunNuclear, Melbourne, USA) for the pre-treatment verification of the stereotactic body radiotherapy (SBRT). Material and methods: For the system, the short and mid-long stability, dose linearity with MU, angular dependence, and field size dependence (ratio of relative output factor) were measured. The results of verification for 15 pre-treatment cancer patients (5 brains, 5 lungs, and 5 livers) performed with SRS MapCHECK and EBT3 Gafchromic films were compared. All the SBRT plans were optimized with the Eclipse (v. 15.6, Varian, Palo Alto, USA) treatment planning system (TPS) using the Acuros XB (Varian, Palo Alto, USA) dose calculation algorithm and were delivered to the Varian EDGE® (Varian, Palo Alto, USA) accelerator equipped with a high-definition multileaf collimator. The 6MV flattening-filter-free beam (FFF) was used. Results: Short and mid-long stability of SRS MapCHECK was very good (0.1%-0.2%), dose linearity with MU and dependence of the response of the detector on field size results were also acceptable (for dose linearity R2 = 1 and 6% difference between microDiamond and SRS MapCHECK response for the smallest field of 1 × 1 cm2). The angular dependence was very good except for the angles close to 90° and 270°. For pre-treatment plan verification, the gamma method was used with the criteria of 3% dose difference and 3 mm distance to agreement (3%/3 mm), and 2%/2 mm, 1%/1 mm, 3%/1 mm, and 2%/1 mm. The highest passing rate for all criteria was observed on the SRS MapCHECK system. Conclusions: It is concluded that SRS MapCHECK with StereoPHAN has sufficient potential for pre-treatment verification of the SBRT plans, so that verification of stereotactic plans can be significantly accelerated.","PeriodicalId":53955,"journal":{"name":"Polish Journal of Medical Physics and Engineering","volume":"64 1 1","pages":"84 - 89"},"PeriodicalIF":0.7000,"publicationDate":"2022-05-28","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Evaluation of SRS MapCHECK with StereoPHAN phantom as a new pre-treatment system verification for SBRT plans\",\"authors\":\"B. Sadowski, Marta Fillmann, Dariusz Szałkowski, P. Kukołowicz\",\"doi\":\"10.2478/pjmpe-2022-0010\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Abstract Introduction: The aim of this study was to evaluate the new 2-Dimensional diode array SRS MapCHECK (SunNuclear, Melbourne, USA) with dedicated phantom StereoPHAN (SunNuclear, Melbourne, USA) for the pre-treatment verification of the stereotactic body radiotherapy (SBRT). Material and methods: For the system, the short and mid-long stability, dose linearity with MU, angular dependence, and field size dependence (ratio of relative output factor) were measured. The results of verification for 15 pre-treatment cancer patients (5 brains, 5 lungs, and 5 livers) performed with SRS MapCHECK and EBT3 Gafchromic films were compared. All the SBRT plans were optimized with the Eclipse (v. 15.6, Varian, Palo Alto, USA) treatment planning system (TPS) using the Acuros XB (Varian, Palo Alto, USA) dose calculation algorithm and were delivered to the Varian EDGE® (Varian, Palo Alto, USA) accelerator equipped with a high-definition multileaf collimator. The 6MV flattening-filter-free beam (FFF) was used. Results: Short and mid-long stability of SRS MapCHECK was very good (0.1%-0.2%), dose linearity with MU and dependence of the response of the detector on field size results were also acceptable (for dose linearity R2 = 1 and 6% difference between microDiamond and SRS MapCHECK response for the smallest field of 1 × 1 cm2). The angular dependence was very good except for the angles close to 90° and 270°. For pre-treatment plan verification, the gamma method was used with the criteria of 3% dose difference and 3 mm distance to agreement (3%/3 mm), and 2%/2 mm, 1%/1 mm, 3%/1 mm, and 2%/1 mm. The highest passing rate for all criteria was observed on the SRS MapCHECK system. Conclusions: It is concluded that SRS MapCHECK with StereoPHAN has sufficient potential for pre-treatment verification of the SBRT plans, so that verification of stereotactic plans can be significantly accelerated.\",\"PeriodicalId\":53955,\"journal\":{\"name\":\"Polish Journal of Medical Physics and Engineering\",\"volume\":\"64 1 1\",\"pages\":\"84 - 89\"},\"PeriodicalIF\":0.7000,\"publicationDate\":\"2022-05-28\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Polish Journal of Medical Physics and Engineering\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.2478/pjmpe-2022-0010\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Polish Journal of Medical Physics and Engineering","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.2478/pjmpe-2022-0010","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

摘要

摘要简介:本研究的目的是评估新型二维二极管阵列SRS MapCHECK (SunNuclear, Melbourne, USA)与专用幻影StereoPHAN (SunNuclear, Melbourne, USA)用于立体定向体放疗(SBRT)的治疗前验证。材料与方法:测定了该系统的短、中长期稳定性、剂量与MU的线性关系、角度依赖性、场大小依赖性(相对输出因子比)。对15例治疗前肿瘤患者(5例脑、5例肺、5例肝)进行SRS MapCHECK和EBT3 Gafchromic膜的验证结果进行比较。使用acros XB (Varian, Palo Alto, USA)剂量计算算法,使用Eclipse (v. 15.6, Varian, Palo Alto, USA)治疗计划系统(TPS)对所有SBRT计划进行优化,并交付给配备高清多叶准直器的Varian EDGE®(Varian, Palo Alto, USA)加速器。采用6MV无压平滤波波束(FFF)。结果:SRS MapCHECK的短、中长稳定性良好(0.1% ~ 0.2%),与MU的剂量线性关系和探测器响应与场大小的依赖关系也可以接受(在最小场为1 × 1 cm2时,microDiamond与SRS MapCHECK的剂量线性关系R2 = 1和6%)。除了接近90°和270°的角度外,角度依赖性很好。对于治疗前计划的验证,采用伽玛方法,标准为剂量差3%,距离一致3mm (3%/ 3mm), 2%/ 2mm, 1%/ 1mm, 3%/ 1mm和2%/ 1mm。在SRS MapCHECK系统上观察到所有标准的最高通过率。结论:结合StereoPHAN的SRS MapCHECK在SBRT方案的预处理验证中具有足够的潜力,可以显著加快立体定向方案的验证。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Evaluation of SRS MapCHECK with StereoPHAN phantom as a new pre-treatment system verification for SBRT plans
Abstract Introduction: The aim of this study was to evaluate the new 2-Dimensional diode array SRS MapCHECK (SunNuclear, Melbourne, USA) with dedicated phantom StereoPHAN (SunNuclear, Melbourne, USA) for the pre-treatment verification of the stereotactic body radiotherapy (SBRT). Material and methods: For the system, the short and mid-long stability, dose linearity with MU, angular dependence, and field size dependence (ratio of relative output factor) were measured. The results of verification for 15 pre-treatment cancer patients (5 brains, 5 lungs, and 5 livers) performed with SRS MapCHECK and EBT3 Gafchromic films were compared. All the SBRT plans were optimized with the Eclipse (v. 15.6, Varian, Palo Alto, USA) treatment planning system (TPS) using the Acuros XB (Varian, Palo Alto, USA) dose calculation algorithm and were delivered to the Varian EDGE® (Varian, Palo Alto, USA) accelerator equipped with a high-definition multileaf collimator. The 6MV flattening-filter-free beam (FFF) was used. Results: Short and mid-long stability of SRS MapCHECK was very good (0.1%-0.2%), dose linearity with MU and dependence of the response of the detector on field size results were also acceptable (for dose linearity R2 = 1 and 6% difference between microDiamond and SRS MapCHECK response for the smallest field of 1 × 1 cm2). The angular dependence was very good except for the angles close to 90° and 270°. For pre-treatment plan verification, the gamma method was used with the criteria of 3% dose difference and 3 mm distance to agreement (3%/3 mm), and 2%/2 mm, 1%/1 mm, 3%/1 mm, and 2%/1 mm. The highest passing rate for all criteria was observed on the SRS MapCHECK system. Conclusions: It is concluded that SRS MapCHECK with StereoPHAN has sufficient potential for pre-treatment verification of the SBRT plans, so that verification of stereotactic plans can be significantly accelerated.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Polish Journal of Medical Physics and Engineering
Polish Journal of Medical Physics and Engineering RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING-
CiteScore
1.30
自引率
0.00%
发文量
19
期刊介绍: Polish Journal of Medical Physics and Engineering (PJMPE) (Online ISSN: 1898-0309; Print ISSN: 1425-4689) is an official publication of the Polish Society of Medical Physics. It is a peer-reviewed, open access scientific journal with no publication fees. The issues are published quarterly online. The Journal publishes original contribution in medical physics and biomedical engineering.
期刊最新文献
Parametrization of subsegmental infarcts using high spatial resolution 2DSTE and synthetic ultrasonic data Automated differential diagnostics of respiratory diseases using an electronic stethoscope Evaluating the impact of anatomical changes on dose distributions in head and neck cancer Comparing eDQE and eNEQ metrics – is there an alternative approach to assessing image quality in digital mammography? Vascular stiffness in cold pressor test hyper-reactors
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1