尼日利亚西南部拉各斯市10个选定x射线中心的峰值千伏电压精度评估:确定x射线发生器能量输出精度的质量控制测试

M. Akpochafor, A. Omojola, K. Soyebi, S. Adeneye, M. Aweda, Helen Bolanle Ajayi
{"title":"尼日利亚西南部拉各斯市10个选定x射线中心的峰值千伏电压精度评估:确定x射线发生器能量输出精度的质量控制测试","authors":"M. Akpochafor, A. Omojola, K. Soyebi, S. Adeneye, M. Aweda, Helen Bolanle Ajayi","doi":"10.4103/2394-2010.184231","DOIUrl":null,"url":null,"abstract":"Background: Most X-ray centers recruited for this study admitted that there have been no maintenance checks on the energy output accuracy of their X-ray machines. Most centers complained of recurrent machine breakdown due to poor power supply, while others pointed out that X-ray machine age could be a contributory factor to poor image quality. Aim and Objectives: To determine peak kilovoltage (kVp) accuracy of individual X-ray unit and to compare their values with recommended standards and to further establish if machine age has a significant effect on mean kVp accuracy. Materials and Methods: One dedicated conventional X-ray unit in each of the ten diagnostic centers was used for this study denoted as D 1−D 10 . kVp was measured using a noninvasive Unfors ThinX RAD kVp meter. The kVp accuracy test was carried out by positioning the Unfors ThinX RAD kVp meter at the center of the beam axis of the patient′s couch at a focus to image distance of 100 cm. The X-ray machine technical factors were varied by selecting tube potential within the range of 50−120 kVp at a fixed tube current of 20 mA. Data analysis value was done using GraphPad Prism version 5.0 statistics software (GraphPad Software, Inc., San Diego, California, USA). Results: The kVp accuracy test for X-ray units D 1 , D 2 , D 5 , D 6 , D 7 , and D 8 were within an acceptable limit of ±5% at all selected tube potential from 50 to 120 kVp, X-ray units D 2 and D 6 showed accurate compliance at much lower kVp accuracy values. X-ray units D 3 and D 10 were above ±5% accuracy limit at all selected kVp tube potential. No statistically significant differences exist between machine age and mean kVp accuracy (P = 0.770). Conclusion: One-fourth of the total kVp accuracy results were above ±5% limit. Old X-ray machine had higher chances of failed kVp accuracy. Traces of failed test were also noticed in X-ray machines below 5 years.","PeriodicalId":16068,"journal":{"name":"Journal of Health Research and Reviews","volume":"68 1","pages":"60 - 65"},"PeriodicalIF":0.0000,"publicationDate":"2016-05-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"17","resultStr":"{\"title\":\"Assessment of peak kilovoltage accuracy in ten selected X-ray centers in Lagos metropolis, South-Western Nigeria: A quality control test to determine energy output accuracy of an X-ray generator\",\"authors\":\"M. Akpochafor, A. Omojola, K. Soyebi, S. Adeneye, M. Aweda, Helen Bolanle Ajayi\",\"doi\":\"10.4103/2394-2010.184231\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Background: Most X-ray centers recruited for this study admitted that there have been no maintenance checks on the energy output accuracy of their X-ray machines. Most centers complained of recurrent machine breakdown due to poor power supply, while others pointed out that X-ray machine age could be a contributory factor to poor image quality. Aim and Objectives: To determine peak kilovoltage (kVp) accuracy of individual X-ray unit and to compare their values with recommended standards and to further establish if machine age has a significant effect on mean kVp accuracy. Materials and Methods: One dedicated conventional X-ray unit in each of the ten diagnostic centers was used for this study denoted as D 1−D 10 . kVp was measured using a noninvasive Unfors ThinX RAD kVp meter. The kVp accuracy test was carried out by positioning the Unfors ThinX RAD kVp meter at the center of the beam axis of the patient′s couch at a focus to image distance of 100 cm. The X-ray machine technical factors were varied by selecting tube potential within the range of 50−120 kVp at a fixed tube current of 20 mA. Data analysis value was done using GraphPad Prism version 5.0 statistics software (GraphPad Software, Inc., San Diego, California, USA). Results: The kVp accuracy test for X-ray units D 1 , D 2 , D 5 , D 6 , D 7 , and D 8 were within an acceptable limit of ±5% at all selected tube potential from 50 to 120 kVp, X-ray units D 2 and D 6 showed accurate compliance at much lower kVp accuracy values. X-ray units D 3 and D 10 were above ±5% accuracy limit at all selected kVp tube potential. No statistically significant differences exist between machine age and mean kVp accuracy (P = 0.770). Conclusion: One-fourth of the total kVp accuracy results were above ±5% limit. Old X-ray machine had higher chances of failed kVp accuracy. Traces of failed test were also noticed in X-ray machines below 5 years.\",\"PeriodicalId\":16068,\"journal\":{\"name\":\"Journal of Health Research and Reviews\",\"volume\":\"68 1\",\"pages\":\"60 - 65\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2016-05-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"17\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Health Research and Reviews\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.4103/2394-2010.184231\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Health Research and Reviews","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.4103/2394-2010.184231","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 17

摘要

背景:本研究招募的大多数x射线中心承认,他们的x射线机没有对能量输出精度进行维护检查。大多数中心抱怨说,由于供电不足,机器经常发生故障,而另一些中心则指出,x光机的年龄可能是导致图像质量差的一个因素。目的和目的:确定单个x射线单元的峰值千伏电压(kVp)精度,并将其值与推荐标准进行比较,并进一步确定机器年龄是否对平均千伏电压精度有显着影响。材料和方法:本研究使用十个诊断中心各一台专用的常规x射线设备,编号d1 - d10。kVp采用无创Unfors ThinX RAD kVp仪测量。kVp精度测试通过将Unfors ThinX RAD kVp计定位在患者沙发光束轴的中心,聚焦成像距离为100 cm进行。在固定管电流为20 mA的条件下,选择50 ~ 120 kVp范围内的管电位来改变x射线机的技术因素。数据分析值采用GraphPad Prism 5.0版统计软件(GraphPad software, Inc., San Diego, California, USA)。结果:x射线单元d1、d2、d2、d2、d2、d2、d6、d7和d8的kVp准确度测试在所有选定的管电位从50到120 kVp范围内均在±5%的可接受范围内,x射线单元d2和d6在更低的kVp准确度值下显示准确的依从性。在所有选择的kVp管电位下,x射线单元d3和d10的准确度都在±5%以上。机器年龄与平均kVp准确率无统计学差异(P = 0.770)。结论:四分之一的kVp准确度在±5%以上。旧的x光机有更高的机会失败的kVp精度。5年以下的x光机也发现检测不合格的痕迹。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Assessment of peak kilovoltage accuracy in ten selected X-ray centers in Lagos metropolis, South-Western Nigeria: A quality control test to determine energy output accuracy of an X-ray generator
Background: Most X-ray centers recruited for this study admitted that there have been no maintenance checks on the energy output accuracy of their X-ray machines. Most centers complained of recurrent machine breakdown due to poor power supply, while others pointed out that X-ray machine age could be a contributory factor to poor image quality. Aim and Objectives: To determine peak kilovoltage (kVp) accuracy of individual X-ray unit and to compare their values with recommended standards and to further establish if machine age has a significant effect on mean kVp accuracy. Materials and Methods: One dedicated conventional X-ray unit in each of the ten diagnostic centers was used for this study denoted as D 1−D 10 . kVp was measured using a noninvasive Unfors ThinX RAD kVp meter. The kVp accuracy test was carried out by positioning the Unfors ThinX RAD kVp meter at the center of the beam axis of the patient′s couch at a focus to image distance of 100 cm. The X-ray machine technical factors were varied by selecting tube potential within the range of 50−120 kVp at a fixed tube current of 20 mA. Data analysis value was done using GraphPad Prism version 5.0 statistics software (GraphPad Software, Inc., San Diego, California, USA). Results: The kVp accuracy test for X-ray units D 1 , D 2 , D 5 , D 6 , D 7 , and D 8 were within an acceptable limit of ±5% at all selected tube potential from 50 to 120 kVp, X-ray units D 2 and D 6 showed accurate compliance at much lower kVp accuracy values. X-ray units D 3 and D 10 were above ±5% accuracy limit at all selected kVp tube potential. No statistically significant differences exist between machine age and mean kVp accuracy (P = 0.770). Conclusion: One-fourth of the total kVp accuracy results were above ±5% limit. Old X-ray machine had higher chances of failed kVp accuracy. Traces of failed test were also noticed in X-ray machines below 5 years.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
Outcomes of patients with end-stage kidney disease hospitalized with COVID-19: A single-center experience from Eastern India Biological plausibility of spirituality in public health The pH and titration values of commonly used beverages in Trinidad: A cross-sectional study A role of community nurses and challenges faced by them toward providing MCH services in selected blocks in West Bengal: A cross-sectional study Lifestyle changes during the Covid-19 lockdown among young adults in Kerala in terms of physical activity, sleeping habits, and perceived psychological problems: A cross-sectional study
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1