西门子E.Cam®Spect系统在加纳Korle-Bu教学医院的性能评估

E. Sosu, F. Hasford, J. Amuasi
{"title":"西门子E.Cam®Spect系统在加纳Korle-Bu教学医院的性能评估","authors":"E. Sosu, F. Hasford, J. Amuasi","doi":"10.4314/JAST.V16I1-2.64789","DOIUrl":null,"url":null,"abstract":"Performance evaluation tests on Siemens e.camR Signature Series Single Photon Emission Computed Tomogra-phy (SPECT) System at Korle-Bu Teaching Hospital (Ghana) were conducted for compliance with Manufactur-erfs specifications and usersf requirements. The tests were performed according to National Electrical Manufac-turers Association protocols, and different measuring instruments and phantoms containing 99m-Tc were used. Extrinsic uniformity, System energy resolution, System spatial resolution (without scattering), Detector shield leakage and System planar sensitivity were evaluated. The System energy resolution measured was 9.38 % as against manufacturersf specification of . 9.9 %, Planar sensitivity value measured was 232.92 c/min/mCi while manufacturersf specification was 202 c/min/mCi, and Detector shield leakage values measured were 0.001 %, 0.022 % and 0.025 % against an accepted value of < 1 %. The performance evaluation confirmed that the Syst-em met requirements for clinical nuclear medicine imaging, and the values could be used as database for future quality control measurements.","PeriodicalId":9207,"journal":{"name":"British Journal of Applied Science and Technology","volume":"39 1","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2011-03-25","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Performance evaluation of Siemens E.Cam® Spect System at Korle-Bu Teaching Hospital (Ghana\",\"authors\":\"E. Sosu, F. Hasford, J. Amuasi\",\"doi\":\"10.4314/JAST.V16I1-2.64789\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Performance evaluation tests on Siemens e.camR Signature Series Single Photon Emission Computed Tomogra-phy (SPECT) System at Korle-Bu Teaching Hospital (Ghana) were conducted for compliance with Manufactur-erfs specifications and usersf requirements. The tests were performed according to National Electrical Manufac-turers Association protocols, and different measuring instruments and phantoms containing 99m-Tc were used. Extrinsic uniformity, System energy resolution, System spatial resolution (without scattering), Detector shield leakage and System planar sensitivity were evaluated. The System energy resolution measured was 9.38 % as against manufacturersf specification of . 9.9 %, Planar sensitivity value measured was 232.92 c/min/mCi while manufacturersf specification was 202 c/min/mCi, and Detector shield leakage values measured were 0.001 %, 0.022 % and 0.025 % against an accepted value of < 1 %. The performance evaluation confirmed that the Syst-em met requirements for clinical nuclear medicine imaging, and the values could be used as database for future quality control measurements.\",\"PeriodicalId\":9207,\"journal\":{\"name\":\"British Journal of Applied Science and Technology\",\"volume\":\"39 1\",\"pages\":\"\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2011-03-25\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"British Journal of Applied Science and Technology\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.4314/JAST.V16I1-2.64789\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"British Journal of Applied Science and Technology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.4314/JAST.V16I1-2.64789","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

在加纳Korle-Bu教学医院对西门子e.camR签名系列单光子发射计算机断层扫描(SPECT)系统进行了性能评估测试,以符合制造商的规范和用户的要求。测试是根据国家电气制造商协会的协议进行的,使用了不同的测量仪器和含有99m-Tc的样品。评估了系统的外部均匀性、系统能量分辨率、系统空间分辨率(无散射)、探测器屏蔽泄漏和系统平面灵敏度。系统的能量分辨率为9.38%,与制造商的规格相符。测量的平面灵敏度值为232.92 c/min/mCi,而制造商的规格为202 c/min/mCi;测量的探测器屏蔽泄漏值分别为0.001 %,0.022%和0.025%,而可接受值< 1%。性能评价证实该系统满足临床核医学成像的要求,其值可作为将来质量控制测量的数据库。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Performance evaluation of Siemens E.Cam® Spect System at Korle-Bu Teaching Hospital (Ghana
Performance evaluation tests on Siemens e.camR Signature Series Single Photon Emission Computed Tomogra-phy (SPECT) System at Korle-Bu Teaching Hospital (Ghana) were conducted for compliance with Manufactur-erfs specifications and usersf requirements. The tests were performed according to National Electrical Manufac-turers Association protocols, and different measuring instruments and phantoms containing 99m-Tc were used. Extrinsic uniformity, System energy resolution, System spatial resolution (without scattering), Detector shield leakage and System planar sensitivity were evaluated. The System energy resolution measured was 9.38 % as against manufacturersf specification of . 9.9 %, Planar sensitivity value measured was 232.92 c/min/mCi while manufacturersf specification was 202 c/min/mCi, and Detector shield leakage values measured were 0.001 %, 0.022 % and 0.025 % against an accepted value of < 1 %. The performance evaluation confirmed that the Syst-em met requirements for clinical nuclear medicine imaging, and the values could be used as database for future quality control measurements.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
Decision Support System of Determining E-Commerce With The Topsis and Comparation With Ahp for UMKM In the Tuban Regency Area Utilization of Virtual Reality Technology in Knowing the Symptoms of Acrophobia and Nyctophobia Implementation of Multiple Intelligences Determinant Applications for Junior High School Students Using Web-Based Likert Scale Method Design and Build a Water Pump Protection Tool Using IOT (Internet Of Things) Based Water Flow Sensor The Study Of The Application Of Noise Mapping Using Golden Surfer Software To Control Noise
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1