尼日利亚卡杜纳州一些放射诊断中心x射线暴露参数的准确性

Abeabe, Christopher Ishiekwen, A. Aliyu
{"title":"尼日利亚卡杜纳州一些放射诊断中心x射线暴露参数的准确性","authors":"Abeabe, Christopher Ishiekwen, A. Aliyu","doi":"10.48153/jrrs.v32i1.223228","DOIUrl":null,"url":null,"abstract":"\n \n \n \nBackground: Dose levels in radio-diagnostic settings depend largely on equipment status and examination techniques. It is essential to carry out quality control (QC) on the x-ray systems to ensure proper optimization and the maintenance of quality for better diagnostic outcome, and to avoid the risk of injury. \nObjective: To carryout QC for the assessment of the accuracy of exposure parameters for the purpose of optimization of radiation protection. \nMethodology: Quality control measurements including tube output, kVp, mA and timer accuracy, linearity and reproducibility were carried out using a multi-function quality control kit, RMI, 181B and measuring tape. \nResults: The results of these measurements indicate that the tube output of the machines within the kVp range of 60 to 120 falls between 13.52 x 10-3 and 161.26 x 10-3 with a linear relationship between the kVp and tube output in all the hospitals. It also showed exposure parameters’ operation to be within tolerance limits except in a single hospital with aspects of out-of-range performance. \nConclusion: All equipment under study functioned within safe limits but further investigation and corrective action is required in one of the centres. \n  \n \n \n \n \n ","PeriodicalId":16919,"journal":{"name":"Journal of Radiography and Radiation Sciences","volume":"29 1","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2022-03-23","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Accuracy of X-Ray Exposure Parameters in some Radio-diagnostic Centres in Kaduna State, Nigeria\",\"authors\":\"Abeabe, Christopher Ishiekwen, A. Aliyu\",\"doi\":\"10.48153/jrrs.v32i1.223228\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"\\n \\n \\n \\nBackground: Dose levels in radio-diagnostic settings depend largely on equipment status and examination techniques. It is essential to carry out quality control (QC) on the x-ray systems to ensure proper optimization and the maintenance of quality for better diagnostic outcome, and to avoid the risk of injury. \\nObjective: To carryout QC for the assessment of the accuracy of exposure parameters for the purpose of optimization of radiation protection. \\nMethodology: Quality control measurements including tube output, kVp, mA and timer accuracy, linearity and reproducibility were carried out using a multi-function quality control kit, RMI, 181B and measuring tape. \\nResults: The results of these measurements indicate that the tube output of the machines within the kVp range of 60 to 120 falls between 13.52 x 10-3 and 161.26 x 10-3 with a linear relationship between the kVp and tube output in all the hospitals. It also showed exposure parameters’ operation to be within tolerance limits except in a single hospital with aspects of out-of-range performance. \\nConclusion: All equipment under study functioned within safe limits but further investigation and corrective action is required in one of the centres. \\n  \\n \\n \\n \\n \\n \",\"PeriodicalId\":16919,\"journal\":{\"name\":\"Journal of Radiography and Radiation Sciences\",\"volume\":\"29 1\",\"pages\":\"\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2022-03-23\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Radiography and Radiation Sciences\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.48153/jrrs.v32i1.223228\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Radiography and Radiation Sciences","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.48153/jrrs.v32i1.223228","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

背景:放射诊断设置中的剂量水平主要取决于设备状态和检查技术。必须对x射线系统进行质量控制(QC),以确保适当的优化和维护质量,以获得更好的诊断结果,并避免受伤的风险。目的:对辐照参数的准确性进行质量控制,以达到优化辐射防护的目的。方法:使用多功能质控试剂盒、RMI、181B和卷尺进行质量控制测量,包括管输出、kVp、mA和计时器精度、线性度和重现性。结果:测量结果表明,在kVp 60 ~ 120范围内,各医院机器的管输出在13.52 × 10-3 ~ 161.26 × 10-3之间,kVp与管输出呈线性关系。它还表明,暴露参数的操作在容忍范围内,除了在一家医院有超出范围的表现。结论:研究中的所有设备都在安全范围内运行,但需要在其中一个中心进行进一步调查和纠正措施。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Accuracy of X-Ray Exposure Parameters in some Radio-diagnostic Centres in Kaduna State, Nigeria
Background: Dose levels in radio-diagnostic settings depend largely on equipment status and examination techniques. It is essential to carry out quality control (QC) on the x-ray systems to ensure proper optimization and the maintenance of quality for better diagnostic outcome, and to avoid the risk of injury. Objective: To carryout QC for the assessment of the accuracy of exposure parameters for the purpose of optimization of radiation protection. Methodology: Quality control measurements including tube output, kVp, mA and timer accuracy, linearity and reproducibility were carried out using a multi-function quality control kit, RMI, 181B and measuring tape. Results: The results of these measurements indicate that the tube output of the machines within the kVp range of 60 to 120 falls between 13.52 x 10-3 and 161.26 x 10-3 with a linear relationship between the kVp and tube output in all the hospitals. It also showed exposure parameters’ operation to be within tolerance limits except in a single hospital with aspects of out-of-range performance. Conclusion: All equipment under study functioned within safe limits but further investigation and corrective action is required in one of the centres.    
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
CT features, Pathological findings and Clinical data of Follicular Dendritic Cell Sarcoma in the Mediastinum: Case Report and Literature Review KNOWLEDGE AND PRACTICE OF IMAGING OF CHILDREN WITH SUSPECTED CASES OF NON-ACCIDENTAL INJURIES AMONG RADIOGRAPHERS WITHIN MAIDUGURI METROPOLIS, NORTHEASTERN NIGERIA Effective doses in chest and abdominal radiography following the ICRP recommendations of 1991 and 2007 in a regional hospital. Investigation of the Effect of Solar Panel Temperature on Power Output Efficiency in Brass, Nigeria. An Assessment of the Lead rubber aprons in Radiodiagnostic centres in South-South Nigeria
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1