{"title":"尼日利亚卡杜纳州一些放射诊断中心x射线暴露参数的准确性","authors":"Abeabe, Christopher Ishiekwen, A. Aliyu","doi":"10.48153/jrrs.v32i1.223228","DOIUrl":null,"url":null,"abstract":"\n \n \n \nBackground: Dose levels in radio-diagnostic settings depend largely on equipment status and examination techniques. It is essential to carry out quality control (QC) on the x-ray systems to ensure proper optimization and the maintenance of quality for better diagnostic outcome, and to avoid the risk of injury. \nObjective: To carryout QC for the assessment of the accuracy of exposure parameters for the purpose of optimization of radiation protection. \nMethodology: Quality control measurements including tube output, kVp, mA and timer accuracy, linearity and reproducibility were carried out using a multi-function quality control kit, RMI, 181B and measuring tape. \nResults: The results of these measurements indicate that the tube output of the machines within the kVp range of 60 to 120 falls between 13.52 x 10-3 and 161.26 x 10-3 with a linear relationship between the kVp and tube output in all the hospitals. It also showed exposure parameters’ operation to be within tolerance limits except in a single hospital with aspects of out-of-range performance. \nConclusion: All equipment under study functioned within safe limits but further investigation and corrective action is required in one of the centres. \n \n \n \n \n \n ","PeriodicalId":16919,"journal":{"name":"Journal of Radiography and Radiation Sciences","volume":"29 1","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2022-03-23","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Accuracy of X-Ray Exposure Parameters in some Radio-diagnostic Centres in Kaduna State, Nigeria\",\"authors\":\"Abeabe, Christopher Ishiekwen, A. Aliyu\",\"doi\":\"10.48153/jrrs.v32i1.223228\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"\\n \\n \\n \\nBackground: Dose levels in radio-diagnostic settings depend largely on equipment status and examination techniques. It is essential to carry out quality control (QC) on the x-ray systems to ensure proper optimization and the maintenance of quality for better diagnostic outcome, and to avoid the risk of injury. \\nObjective: To carryout QC for the assessment of the accuracy of exposure parameters for the purpose of optimization of radiation protection. \\nMethodology: Quality control measurements including tube output, kVp, mA and timer accuracy, linearity and reproducibility were carried out using a multi-function quality control kit, RMI, 181B and measuring tape. \\nResults: The results of these measurements indicate that the tube output of the machines within the kVp range of 60 to 120 falls between 13.52 x 10-3 and 161.26 x 10-3 with a linear relationship between the kVp and tube output in all the hospitals. It also showed exposure parameters’ operation to be within tolerance limits except in a single hospital with aspects of out-of-range performance. \\nConclusion: All equipment under study functioned within safe limits but further investigation and corrective action is required in one of the centres. \\n \\n \\n \\n \\n \\n \",\"PeriodicalId\":16919,\"journal\":{\"name\":\"Journal of Radiography and Radiation Sciences\",\"volume\":\"29 1\",\"pages\":\"\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2022-03-23\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Radiography and Radiation Sciences\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.48153/jrrs.v32i1.223228\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Radiography and Radiation Sciences","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.48153/jrrs.v32i1.223228","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
Accuracy of X-Ray Exposure Parameters in some Radio-diagnostic Centres in Kaduna State, Nigeria
Background: Dose levels in radio-diagnostic settings depend largely on equipment status and examination techniques. It is essential to carry out quality control (QC) on the x-ray systems to ensure proper optimization and the maintenance of quality for better diagnostic outcome, and to avoid the risk of injury.
Objective: To carryout QC for the assessment of the accuracy of exposure parameters for the purpose of optimization of radiation protection.
Methodology: Quality control measurements including tube output, kVp, mA and timer accuracy, linearity and reproducibility were carried out using a multi-function quality control kit, RMI, 181B and measuring tape.
Results: The results of these measurements indicate that the tube output of the machines within the kVp range of 60 to 120 falls between 13.52 x 10-3 and 161.26 x 10-3 with a linear relationship between the kVp and tube output in all the hospitals. It also showed exposure parameters’ operation to be within tolerance limits except in a single hospital with aspects of out-of-range performance.
Conclusion: All equipment under study functioned within safe limits but further investigation and corrective action is required in one of the centres.