{"title":"用于无菌重组连接到液体袋的小药瓶的封闭系统传输装置的容器封闭完整性测试和工艺验证。","authors":"Roland B van den Berg, Kübra Akgöl, Eleonora L Swart, Bastiaan Nuijen, Mirjam Crul","doi":"10.1136/ejhpharm-2022-003604","DOIUrl":null,"url":null,"abstract":"<p><strong>Objectives: </strong>The closure integrity and process validation of closed system transfer devices (CSTDs) should be confirmed before implementation in clinical settings. We aimed to investigate the closure integrity and validate the aseptic procedure of two types of CSTDs by using a combination of the dye ingress test and a media fill test.</p><p><strong>Methods: </strong>The dye ingress test with methylene blue was used for both CSTDs with 10 samples of drug vials of three brands. A media fill test was performed with both CSTDs (300 samples per CSTD, 150 carried out in a safety cabinet and 150 under non-classified environmental conditions).</p><p><strong>Results: </strong>In all samples of both CSTDs, methylene blue was absent after visual inspection and spectrophotometric analysis. The nutrient media of one sample with CSTD A and none of the CSTD B samples were contaminated when reconstituted in a GMP grade A environment. Under non-classified environmental conditions, one sample of CSTD A and two samples of CSTD B were contaminated.</p><p><strong>Conclusions: </strong>Both CSTDs connected to the drug vials met the terms of closure integrity by using the dye ingress test. The aseptic procedure of CSTD B was validated with the media fill test when reconstituted in a GMP grade A environment, but failed for CSTD A. Both CSTDs failed the media fill test when reconstituted under non-classified environmental conditions.</p>","PeriodicalId":12050,"journal":{"name":"European journal of hospital pharmacy : science and practice","volume":" ","pages":"358-362"},"PeriodicalIF":1.6000,"publicationDate":"2024-06-21","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11265557/pdf/","citationCount":"0","resultStr":"{\"title\":\"Container closure integrity testing and process validation of closed system transfer devices for aseptic reconstitution of drug vials connected to fluid bags.\",\"authors\":\"Roland B van den Berg, Kübra Akgöl, Eleonora L Swart, Bastiaan Nuijen, Mirjam Crul\",\"doi\":\"10.1136/ejhpharm-2022-003604\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Objectives: </strong>The closure integrity and process validation of closed system transfer devices (CSTDs) should be confirmed before implementation in clinical settings. We aimed to investigate the closure integrity and validate the aseptic procedure of two types of CSTDs by using a combination of the dye ingress test and a media fill test.</p><p><strong>Methods: </strong>The dye ingress test with methylene blue was used for both CSTDs with 10 samples of drug vials of three brands. A media fill test was performed with both CSTDs (300 samples per CSTD, 150 carried out in a safety cabinet and 150 under non-classified environmental conditions).</p><p><strong>Results: </strong>In all samples of both CSTDs, methylene blue was absent after visual inspection and spectrophotometric analysis. The nutrient media of one sample with CSTD A and none of the CSTD B samples were contaminated when reconstituted in a GMP grade A environment. Under non-classified environmental conditions, one sample of CSTD A and two samples of CSTD B were contaminated.</p><p><strong>Conclusions: </strong>Both CSTDs connected to the drug vials met the terms of closure integrity by using the dye ingress test. The aseptic procedure of CSTD B was validated with the media fill test when reconstituted in a GMP grade A environment, but failed for CSTD A. Both CSTDs failed the media fill test when reconstituted under non-classified environmental conditions.</p>\",\"PeriodicalId\":12050,\"journal\":{\"name\":\"European journal of hospital pharmacy : science and practice\",\"volume\":\" \",\"pages\":\"358-362\"},\"PeriodicalIF\":1.6000,\"publicationDate\":\"2024-06-21\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11265557/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"European journal of hospital pharmacy : science and practice\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1136/ejhpharm-2022-003604\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"PHARMACOLOGY & PHARMACY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"European journal of hospital pharmacy : science and practice","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1136/ejhpharm-2022-003604","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"PHARMACOLOGY & PHARMACY","Score":null,"Total":0}
引用次数: 0
摘要
目的:在临床应用之前,应确认封闭系统转移装置(CSTD)的封闭完整性和流程验证。我们的目的是通过结合使用染料进入测试和介质填充测试,研究两种 CSTD 的封闭完整性并验证其无菌程序:方法:使用亚甲基蓝对两种 CSTD 进行染料侵入测试,测试对象为三种品牌的 10 个药瓶样品。对两个 CSTD 都进行了介质填充试验(每个 CSTD 300 个样品,其中 150 个在安全柜中进行,150 个在非机密环境条件下进行):结果:经目测和分光光度分析,两种 CSTD 的所有样品中都不含亚甲基蓝。在 GMP A 级环境中重组时,一个 CSTD A 样品和一个 CSTD B 样品的营养介质均未受到污染。在非分级环境条件下,一个 CSTD A 样品和两个 CSTD B 样品受到污染:通过染料渗入测试,与药瓶相连的两个 CSTD 都符合封闭完整性的要求。在 GMP A 级环境中重组时,CSTD B 的无菌程序通过了介质填充测试验证,但 CSTD A 的无菌程序未通过测试。
Container closure integrity testing and process validation of closed system transfer devices for aseptic reconstitution of drug vials connected to fluid bags.
Objectives: The closure integrity and process validation of closed system transfer devices (CSTDs) should be confirmed before implementation in clinical settings. We aimed to investigate the closure integrity and validate the aseptic procedure of two types of CSTDs by using a combination of the dye ingress test and a media fill test.
Methods: The dye ingress test with methylene blue was used for both CSTDs with 10 samples of drug vials of three brands. A media fill test was performed with both CSTDs (300 samples per CSTD, 150 carried out in a safety cabinet and 150 under non-classified environmental conditions).
Results: In all samples of both CSTDs, methylene blue was absent after visual inspection and spectrophotometric analysis. The nutrient media of one sample with CSTD A and none of the CSTD B samples were contaminated when reconstituted in a GMP grade A environment. Under non-classified environmental conditions, one sample of CSTD A and two samples of CSTD B were contaminated.
Conclusions: Both CSTDs connected to the drug vials met the terms of closure integrity by using the dye ingress test. The aseptic procedure of CSTD B was validated with the media fill test when reconstituted in a GMP grade A environment, but failed for CSTD A. Both CSTDs failed the media fill test when reconstituted under non-classified environmental conditions.
期刊介绍:
European Journal of Hospital Pharmacy (EJHP) offers a high quality, peer-reviewed platform for the publication of practical and innovative research which aims to strengthen the profile and professional status of hospital pharmacists. EJHP is committed to being the leading journal on all aspects of hospital pharmacy, thereby advancing the science, practice and profession of hospital pharmacy. The journal aims to become a major source for education and inspiration to improve practice and the standard of patient care in hospitals and related institutions worldwide.