{"title":"乳房切除术后放疗:栓剂厚度和照射技术对皮肤剂量的影响。","authors":"Frédéric A. Miéville, Nicolas Pitteloud, Vérane Achard, Giorgio Lamanna, Olivier Pisaturo, Pierre-Alain Tercier, Abdelkarim S. Allal","doi":"10.1016/j.zemedi.2023.03.004","DOIUrl":null,"url":null,"abstract":"<div><h3>Purpose</h3><div>To determine 10 MV IMRT and VMAT based protocols with a daily bolus targeting a skin dose of 45 Gy in order to replace the 6 MV tangential fields with a 5 mm thick bolus on alternate days method for post-mastectomy radiotherapy.</div></div><div><h3>Method</h3><div>We measured the mean surface dose along the chest wall PTV as a function of different bolus thicknesses for sliding window IMRT and VMAT plans. We analyzed surface dose profiles and dose homogeneities and compared them to our standard 6 MV strategy. All measurements were performed on a thorax phantom with Gafchromic films while dosimetric plans were computed using the Acuros XB algorithm (Varian).</div></div><div><h3>Results</h3><div>We obtained the best compromise between measured surface dose (mean dose and homogeneity) and skin toxicity threshold obtained from the literature using a daily 3 mm thick bolus. Mean surface doses were 91.4 ± 2.8% [85.7% – 95.4%] and 92.2 ± 2.3% [85.6% – 95.2%] of the prescribed dose with IMRT and VMAT techniques, respectively. Our standard 6 MV alternate days 5 mm thick bolus leads to 89.0 ± 3.7% [83.6% – 95.5%]. Mean dose differences between measured and TPS results were < 3.2% for depths as low as 2 mm depth.</div></div><div><h3>Conclusion</h3><div>10 MV IMRT-based protocols with a daily 3 mm thick bolus produce a surface dose comparable to the standard 6 MV 5 mm thick bolus on alternate days method but with an improved surface dose homogeneity. This allows for a better control of skin toxicity and target volume coverage.</div></div>","PeriodicalId":54397,"journal":{"name":"Zeitschrift fur Medizinische Physik","volume":"34 4","pages":"Pages 542-554"},"PeriodicalIF":2.4000,"publicationDate":"2024-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Post-mastectomy radiotherapy: Impact of bolus thickness and irradiation technique on skin dose\",\"authors\":\"Frédéric A. Miéville, Nicolas Pitteloud, Vérane Achard, Giorgio Lamanna, Olivier Pisaturo, Pierre-Alain Tercier, Abdelkarim S. Allal\",\"doi\":\"10.1016/j.zemedi.2023.03.004\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Purpose</h3><div>To determine 10 MV IMRT and VMAT based protocols with a daily bolus targeting a skin dose of 45 Gy in order to replace the 6 MV tangential fields with a 5 mm thick bolus on alternate days method for post-mastectomy radiotherapy.</div></div><div><h3>Method</h3><div>We measured the mean surface dose along the chest wall PTV as a function of different bolus thicknesses for sliding window IMRT and VMAT plans. We analyzed surface dose profiles and dose homogeneities and compared them to our standard 6 MV strategy. All measurements were performed on a thorax phantom with Gafchromic films while dosimetric plans were computed using the Acuros XB algorithm (Varian).</div></div><div><h3>Results</h3><div>We obtained the best compromise between measured surface dose (mean dose and homogeneity) and skin toxicity threshold obtained from the literature using a daily 3 mm thick bolus. Mean surface doses were 91.4 ± 2.8% [85.7% – 95.4%] and 92.2 ± 2.3% [85.6% – 95.2%] of the prescribed dose with IMRT and VMAT techniques, respectively. Our standard 6 MV alternate days 5 mm thick bolus leads to 89.0 ± 3.7% [83.6% – 95.5%]. Mean dose differences between measured and TPS results were < 3.2% for depths as low as 2 mm depth.</div></div><div><h3>Conclusion</h3><div>10 MV IMRT-based protocols with a daily 3 mm thick bolus produce a surface dose comparable to the standard 6 MV 5 mm thick bolus on alternate days method but with an improved surface dose homogeneity. This allows for a better control of skin toxicity and target volume coverage.</div></div>\",\"PeriodicalId\":54397,\"journal\":{\"name\":\"Zeitschrift fur Medizinische Physik\",\"volume\":\"34 4\",\"pages\":\"Pages 542-554\"},\"PeriodicalIF\":2.4000,\"publicationDate\":\"2024-11-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Zeitschrift fur Medizinische Physik\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S0939388923000417\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Zeitschrift fur Medizinische Physik","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0939388923000417","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
Post-mastectomy radiotherapy: Impact of bolus thickness and irradiation technique on skin dose
Purpose
To determine 10 MV IMRT and VMAT based protocols with a daily bolus targeting a skin dose of 45 Gy in order to replace the 6 MV tangential fields with a 5 mm thick bolus on alternate days method for post-mastectomy radiotherapy.
Method
We measured the mean surface dose along the chest wall PTV as a function of different bolus thicknesses for sliding window IMRT and VMAT plans. We analyzed surface dose profiles and dose homogeneities and compared them to our standard 6 MV strategy. All measurements were performed on a thorax phantom with Gafchromic films while dosimetric plans were computed using the Acuros XB algorithm (Varian).
Results
We obtained the best compromise between measured surface dose (mean dose and homogeneity) and skin toxicity threshold obtained from the literature using a daily 3 mm thick bolus. Mean surface doses were 91.4 ± 2.8% [85.7% – 95.4%] and 92.2 ± 2.3% [85.6% – 95.2%] of the prescribed dose with IMRT and VMAT techniques, respectively. Our standard 6 MV alternate days 5 mm thick bolus leads to 89.0 ± 3.7% [83.6% – 95.5%]. Mean dose differences between measured and TPS results were < 3.2% for depths as low as 2 mm depth.
Conclusion
10 MV IMRT-based protocols with a daily 3 mm thick bolus produce a surface dose comparable to the standard 6 MV 5 mm thick bolus on alternate days method but with an improved surface dose homogeneity. This allows for a better control of skin toxicity and target volume coverage.
期刊介绍:
Zeitschrift fur Medizinische Physik (Journal of Medical Physics) is an official organ of the German and Austrian Society of Medical Physic and the Swiss Society of Radiobiology and Medical Physics.The Journal is a platform for basic research and practical applications of physical procedures in medical diagnostics and therapy. The articles are reviewed following international standards of peer reviewing.
Focuses of the articles are:
-Biophysical methods in radiation therapy and nuclear medicine
-Dosimetry and radiation protection
-Radiological diagnostics and quality assurance
-Modern imaging techniques, such as computed tomography, magnetic resonance imaging, positron emission tomography
-Ultrasonography diagnostics, application of laser and UV rays
-Electronic processing of biosignals
-Artificial intelligence and machine learning in medical physics
In the Journal, the latest scientific insights find their expression in the form of original articles, reviews, technical communications, and information for the clinical practice.