分流装置对胎儿后交通动脉瘤的血流动力学改变:使用 CFD 进行模拟研究,比较超越流线型、管道柔性和管桥装置

IF 3 3区 医学 Q2 CLINICAL NEUROLOGY Journal of Neuroradiology Pub Date : 2024-02-01 DOI:10.1016/j.neurad.2023.07.002
Yu Fu , Xin Bian , Rong Zou , Rongbo Jin , Xiaochang Leng , Feng Fan , Sen Wei , Xuan Cui , Jianping Xiang , Sheng Guan
{"title":"分流装置对胎儿后交通动脉瘤的血流动力学改变:使用 CFD 进行模拟研究,比较超越流线型、管道柔性和管桥装置","authors":"Yu Fu ,&nbsp;Xin Bian ,&nbsp;Rong Zou ,&nbsp;Rongbo Jin ,&nbsp;Xiaochang Leng ,&nbsp;Feng Fan ,&nbsp;Sen Wei ,&nbsp;Xuan Cui ,&nbsp;Jianping Xiang ,&nbsp;Sheng Guan","doi":"10.1016/j.neurad.2023.07.002","DOIUrl":null,"url":null,"abstract":"<div><h3>Purpose</h3><p>Traditional flow diverters (FDs) for treating aneurysms at the fetal posterior communicating artery<span> are unsatisfactory. Surpass Streamline is a novel FD with different mesh characteristics; however, the outcomes for such aneurysms remain unclear. This study aimed to compare hemodynamic alterations induced by Surpass Streamline, Pipeline Flex, and Tubridge devices and explore possible strategies for aneurysms at the fetal posterior communicating artery.</span></p></div><div><h3>Methods</h3><p>Two simulated aneurysms (Case 1, Case 2) were constructed from digital subtraction angiography (DSA). The three FDs were virtually deployed, and hemodynamic analysis based on computational fluid dynamics was performed. Hemodynamic parameters, including the sac-averaged velocity magnitude (Velocity), high-flow volume (HFV), and wall shear stress (WSS), were compared between each FD and the untreated model (control). Surpass Streamline was performed in real life for two aneurysms and the clinical outcomes were collected for analysis.</p></div><div><h3>Results</h3><p>Compared to the control, the Surpass resulted in the most significant reduction in flow. In Case 1, the Velocity, HFV, and WSS were reduced by 51.6%, 78.1%, and 64.3%, respectively. In Case 2, the Velocity, HFV, and WSS were reduced by 48.0%, 81.1%, and 65.3%, respectively. Tubridge showed slightly larger changes in hemodynamic parameters than Pipeline. In addition, our analysis suggested that metal coverage was correlated with the WSS, Velocity, and HFV. The postoperative DSA showed that the aneurysm was nearly occluded in Case 1 and decreased in Case 2.</p></div><div><h3>Conclusion</h3><p>Compared to that with the Pipeline and Tubridge, the Surpass resulted in the greatest reduction in hemodynamic parameters and might be effective for aneurysms at the fetal posterior communicating artery. Virtual FD deployment and computational fluid dynamics analysis may be used to predict the treatment outcomes.</p></div>","PeriodicalId":50115,"journal":{"name":"Journal of Neuroradiology","volume":"51 1","pages":"Pages 74-81"},"PeriodicalIF":3.0000,"publicationDate":"2024-02-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Hemodynamic alterations of flow diverters on aneurysms at the fetal posterior communicating artery: A simulation study using CFD to compare the surpass streamline, pipeline flex, and tubridge devices\",\"authors\":\"Yu Fu ,&nbsp;Xin Bian ,&nbsp;Rong Zou ,&nbsp;Rongbo Jin ,&nbsp;Xiaochang Leng ,&nbsp;Feng Fan ,&nbsp;Sen Wei ,&nbsp;Xuan Cui ,&nbsp;Jianping Xiang ,&nbsp;Sheng Guan\",\"doi\":\"10.1016/j.neurad.2023.07.002\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Purpose</h3><p>Traditional flow diverters (FDs) for treating aneurysms at the fetal posterior communicating artery<span> are unsatisfactory. Surpass Streamline is a novel FD with different mesh characteristics; however, the outcomes for such aneurysms remain unclear. This study aimed to compare hemodynamic alterations induced by Surpass Streamline, Pipeline Flex, and Tubridge devices and explore possible strategies for aneurysms at the fetal posterior communicating artery.</span></p></div><div><h3>Methods</h3><p>Two simulated aneurysms (Case 1, Case 2) were constructed from digital subtraction angiography (DSA). The three FDs were virtually deployed, and hemodynamic analysis based on computational fluid dynamics was performed. Hemodynamic parameters, including the sac-averaged velocity magnitude (Velocity), high-flow volume (HFV), and wall shear stress (WSS), were compared between each FD and the untreated model (control). Surpass Streamline was performed in real life for two aneurysms and the clinical outcomes were collected for analysis.</p></div><div><h3>Results</h3><p>Compared to the control, the Surpass resulted in the most significant reduction in flow. In Case 1, the Velocity, HFV, and WSS were reduced by 51.6%, 78.1%, and 64.3%, respectively. In Case 2, the Velocity, HFV, and WSS were reduced by 48.0%, 81.1%, and 65.3%, respectively. Tubridge showed slightly larger changes in hemodynamic parameters than Pipeline. In addition, our analysis suggested that metal coverage was correlated with the WSS, Velocity, and HFV. The postoperative DSA showed that the aneurysm was nearly occluded in Case 1 and decreased in Case 2.</p></div><div><h3>Conclusion</h3><p>Compared to that with the Pipeline and Tubridge, the Surpass resulted in the greatest reduction in hemodynamic parameters and might be effective for aneurysms at the fetal posterior communicating artery. Virtual FD deployment and computational fluid dynamics analysis may be used to predict the treatment outcomes.</p></div>\",\"PeriodicalId\":50115,\"journal\":{\"name\":\"Journal of Neuroradiology\",\"volume\":\"51 1\",\"pages\":\"Pages 74-81\"},\"PeriodicalIF\":3.0000,\"publicationDate\":\"2024-02-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Neuroradiology\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S0150986123002213\",\"RegionNum\":3,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"CLINICAL NEUROLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Neuroradiology","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0150986123002213","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"CLINICAL NEUROLOGY","Score":null,"Total":0}
引用次数: 0

摘要

目的传统的血流分流器(FD)治疗胎儿后交通动脉瘤的效果并不理想。Surpass Streamline 是一种新型的血流分流器,具有不同的网状特征;但对此类动脉瘤的治疗效果仍不清楚。本研究旨在比较 Surpass Streamline、Pipeline Flex 和 Tubridge 装置引起的血流动力学改变,并探索治疗胎儿后交通动脉瘤的可能策略。虚拟部署了三个 FD,并根据计算流体动力学进行了血流动力学分析。比较了每个 FD 与未处理模型(对照组)之间的血流动力学参数,包括囊平均速度幅值(Velocity)、高流量(HFV)和壁剪应力(WSS)。结果与对照组相比,Surpass 使血流明显减少。在病例 1 中,流速、HFV 和 WSS 分别降低了 51.6%、78.1% 和 64.3%。在病例 2 中,流速、高频变压和 WSS 分别降低了 48.0%、81.1% 和 65.3%。管桥的血液动力学参数变化略大于管道。此外,我们的分析表明,金属覆盖与 WSS、Velocity 和 HFV 相关。结论与 Pipeline 和 Tubridge 相比,Surpass 能最大程度地降低血流动力学参数,可能对胎儿后交通动脉瘤有效。虚拟 FD 部署和计算流体动力学分析可用于预测治疗效果。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Hemodynamic alterations of flow diverters on aneurysms at the fetal posterior communicating artery: A simulation study using CFD to compare the surpass streamline, pipeline flex, and tubridge devices

Purpose

Traditional flow diverters (FDs) for treating aneurysms at the fetal posterior communicating artery are unsatisfactory. Surpass Streamline is a novel FD with different mesh characteristics; however, the outcomes for such aneurysms remain unclear. This study aimed to compare hemodynamic alterations induced by Surpass Streamline, Pipeline Flex, and Tubridge devices and explore possible strategies for aneurysms at the fetal posterior communicating artery.

Methods

Two simulated aneurysms (Case 1, Case 2) were constructed from digital subtraction angiography (DSA). The three FDs were virtually deployed, and hemodynamic analysis based on computational fluid dynamics was performed. Hemodynamic parameters, including the sac-averaged velocity magnitude (Velocity), high-flow volume (HFV), and wall shear stress (WSS), were compared between each FD and the untreated model (control). Surpass Streamline was performed in real life for two aneurysms and the clinical outcomes were collected for analysis.

Results

Compared to the control, the Surpass resulted in the most significant reduction in flow. In Case 1, the Velocity, HFV, and WSS were reduced by 51.6%, 78.1%, and 64.3%, respectively. In Case 2, the Velocity, HFV, and WSS were reduced by 48.0%, 81.1%, and 65.3%, respectively. Tubridge showed slightly larger changes in hemodynamic parameters than Pipeline. In addition, our analysis suggested that metal coverage was correlated with the WSS, Velocity, and HFV. The postoperative DSA showed that the aneurysm was nearly occluded in Case 1 and decreased in Case 2.

Conclusion

Compared to that with the Pipeline and Tubridge, the Surpass resulted in the greatest reduction in hemodynamic parameters and might be effective for aneurysms at the fetal posterior communicating artery. Virtual FD deployment and computational fluid dynamics analysis may be used to predict the treatment outcomes.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Journal of Neuroradiology
Journal of Neuroradiology 医学-核医学
CiteScore
6.10
自引率
5.70%
发文量
142
审稿时长
6-12 weeks
期刊介绍: The Journal of Neuroradiology is a peer-reviewed journal, publishing worldwide clinical and basic research in the field of diagnostic and Interventional neuroradiology, translational and molecular neuroimaging, and artificial intelligence in neuroradiology. The Journal of Neuroradiology considers for publication articles, reviews, technical notes and letters to the editors (correspondence section), provided that the methodology and scientific content are of high quality, and that the results will have substantial clinical impact and/or physiological importance.
期刊最新文献
Impaired iron metabolism and cerebral perfusion patterns in unilateral middle cerebral artery stenosis or occlusion: Insights from quantitative susceptibility mapping Deep learning segmentation-based bone removal from computed tomography of the brain improves subdural hematoma detection Intra-operative use of Augmented Reality for 3D visualisation of rotational angiography data: Feasibility and workflow demonstration using a PCOM aneurysm case Editorial board Contents
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1