Yuta Endo, Kuninori Kobayashi, Haruna Shibo, Makoto Amanuma, Shigehide Kuhara
{"title":"Using Dictionary Matching to Improve the Accuracy of MOLLI Myocardial T1 Analysis and Measurements of Heart Rate Variability.","authors":"Yuta Endo, Kuninori Kobayashi, Haruna Shibo, Makoto Amanuma, Shigehide Kuhara","doi":"10.2463/mrms.tn.2022-0013","DOIUrl":null,"url":null,"abstract":"<p><p>We analyzed modified Look-Locker inversion recovery (MOLLI) T1 measurements by applying a dictionary matching strategy and aimed to acquire T1 measurements more accurately than those acquired by the conventional three-parameter matching analysis. We particularly clarified the robustness of this method for measuring heart rate (HR) variability. A phantom experiment using a 3T MRI system was performed for various HRs. The ideal MOLLI signal corresponding to the scan parameter in the MRI experiment was simulated over a wide range of T1 values according to the dictionary. The unknown T1 values were determined by finding the simulated signals in the dictionary corresponding to the measured signals using pattern matching. The measured T1 values showed that the proposed analysis improved the accuracy of T1 measurements compared to those acquired by traditional analysis by up to 10%. In addition, the variability of measurements at several HRs was reduced by up to 100 ms.</p>","PeriodicalId":18119,"journal":{"name":"Magnetic Resonance in Medical Sciences","volume":"22 3","pages":"389-399"},"PeriodicalIF":2.5000,"publicationDate":"2023-07-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://ftp.ncbi.nlm.nih.gov/pub/pmc/oa_pdf/48/ad/mrms-22-389.PMC10449558.pdf","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Magnetic Resonance in Medical Sciences","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.2463/mrms.tn.2022-0013","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0
Abstract
We analyzed modified Look-Locker inversion recovery (MOLLI) T1 measurements by applying a dictionary matching strategy and aimed to acquire T1 measurements more accurately than those acquired by the conventional three-parameter matching analysis. We particularly clarified the robustness of this method for measuring heart rate (HR) variability. A phantom experiment using a 3T MRI system was performed for various HRs. The ideal MOLLI signal corresponding to the scan parameter in the MRI experiment was simulated over a wide range of T1 values according to the dictionary. The unknown T1 values were determined by finding the simulated signals in the dictionary corresponding to the measured signals using pattern matching. The measured T1 values showed that the proposed analysis improved the accuracy of T1 measurements compared to those acquired by traditional analysis by up to 10%. In addition, the variability of measurements at several HRs was reduced by up to 100 ms.
期刊介绍:
Magnetic Resonance in Medical Sciences (MRMS or Magn
Reson Med Sci) is an international journal pursuing the
publication of original articles contributing to the progress
of magnetic resonance in the field of biomedical sciences
including technical developments and clinical applications.
MRMS is an official journal of the Japanese Society for
Magnetic Resonance in Medicine (JSMRM).