首页 > 最新文献

Magnetic Resonance in Medical Sciences最新文献

英文 中文
Novel T1 Analysis Method to Address Reduced Measurement Accuracy Due to Irregular Heart Rate Variability in Myocardial T1 Mapping Using Polarity-corrected Inversion Time Preparation. 一种新的T1分析方法,用于解决使用极性校正反演时间制备的心肌T1制图中不规则心率变异性导致测量精度降低的问题。
IF 2.5 3区 医学 Q2 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2025-01-01 Epub Date: 2023-09-04 DOI: 10.2463/mrms.mp.2023-0029
Yuta Endo, Sanae Takahashi, Haruna Shibo, Makoto Amanuma, Kuninori Kobayashi, Shigehide Kuhara

Purpose: Polarity-corrected inversion time preparation (PCTIP), a myocardial T1 mapping technique, is expected to reduce measurement underestimation in the modified Look-Locker inversion recover method. However, measurement precision is reduced, especially for heart rate variability. We devised an analysis using a recurrence formula to overcome this problem and showed that it improved the measurement accuracy, especially at high heart rates. Therefore, this study aimed to determine the effect of this analysis on the accuracy and precision of T1 measurements for irregular heart rate variability.

Methods: A PCTIP scan using a 3T MRI scanner was performed in phantom experiment. We generated the simulated R-waves required for electrocardiogram (ECG)-gated acquisition using a signal generator set to 30 combinations. T1 map was generated using the signal train of the PCTIP images by nonlinear curve fitting using conventional and recurrence formulas. Accuracy against reference T1 and precision of heart rate variability were evaluated. To evaluate the fitting accuracy of both analyses, the relative fitting error was calculated.

Results: For the longer T1, the fitting error was larger than the short T1, with the conventional analysis showing 10.1±2.0%. The recurrence formula analysis showed a small fitting error less than 1%, which was consistent for all heart rate variability patterns. In the conventional analysis, the accuracy, especially for longer T1, showed a large underestimation of the measurements and poor linearity. However, in the recurrence formula analysis, the accuracy improved at a long T1, and linearity also improved. The Bland-Altman plot showed that it varied greatly depending on the heart rate variability pattern for the longer T1 in the conventional analysis, whereas the recurrence formula analysis suppressed this variation.

Conclusion: T1 analysis of PCTIP using the recurrence formula analysis achieved accurate and precise T1 measurements, even for irregular heart rate variability.

目的:极性校正反演时间制备(PCTIP)是一种心肌T1定位技术,有望减少改进的Look-Locker反演恢复方法中的测量低估。然而,测量精度降低,特别是心率变异性。我们设计了一个使用递归公式的分析来克服这个问题,并表明它提高了测量精度,特别是在高心率时。因此,本研究旨在确定该分析对不规则心率变异性T1测量的准确性和精密度的影响。方法:采用3T MRI扫描进行幻像实验。我们使用一组30种组合的信号发生器产生了心电图门控采集所需的模拟r波。利用PCTIP图像的信号序列,采用常规公式和递归公式进行非线性曲线拟合,生成T1图。对照参考T1和心率变异性的准确性进行了评估。为了评价两种分析的拟合精度,计算了相对拟合误差。结果:对于较长的T1,拟合误差大于较短的T1,常规分析误差为10.1±2.0%。递归公式分析显示,拟合误差小于1%,这与所有心率变异性模式一致。在传统的分析中,精度,特别是对于较长的T1,显示出测量值的严重低估和较差的线性。然而,在递归式分析中,在较长的T1下精度有所提高,线性度也有所提高。Bland-Altman图显示,在常规分析中,随着T1的延长,心率变异性模式的变化很大,而递归公式分析则抑制了这种变化。结论:使用递推公式分析PCTIP的T1分析可以获得准确和精确的T1测量,即使对于不规则心率变异性也是如此。
{"title":"Novel T1 Analysis Method to Address Reduced Measurement Accuracy Due to Irregular Heart Rate Variability in Myocardial T1 Mapping Using Polarity-corrected Inversion Time Preparation.","authors":"Yuta Endo, Sanae Takahashi, Haruna Shibo, Makoto Amanuma, Kuninori Kobayashi, Shigehide Kuhara","doi":"10.2463/mrms.mp.2023-0029","DOIUrl":"10.2463/mrms.mp.2023-0029","url":null,"abstract":"<p><strong>Purpose: </strong>Polarity-corrected inversion time preparation (PCTIP), a myocardial T1 mapping technique, is expected to reduce measurement underestimation in the modified Look-Locker inversion recover method. However, measurement precision is reduced, especially for heart rate variability. We devised an analysis using a recurrence formula to overcome this problem and showed that it improved the measurement accuracy, especially at high heart rates. Therefore, this study aimed to determine the effect of this analysis on the accuracy and precision of T1 measurements for irregular heart rate variability.</p><p><strong>Methods: </strong>A PCTIP scan using a 3T MRI scanner was performed in phantom experiment. We generated the simulated R-waves required for electrocardiogram (ECG)-gated acquisition using a signal generator set to 30 combinations. T1 map was generated using the signal train of the PCTIP images by nonlinear curve fitting using conventional and recurrence formulas. Accuracy against reference T1 and precision of heart rate variability were evaluated. To evaluate the fitting accuracy of both analyses, the relative fitting error was calculated.</p><p><strong>Results: </strong>For the longer T1, the fitting error was larger than the short T1, with the conventional analysis showing 10.1±2.0%. The recurrence formula analysis showed a small fitting error less than 1%, which was consistent for all heart rate variability patterns. In the conventional analysis, the accuracy, especially for longer T1, showed a large underestimation of the measurements and poor linearity. However, in the recurrence formula analysis, the accuracy improved at a long T1, and linearity also improved. The Bland-Altman plot showed that it varied greatly depending on the heart rate variability pattern for the longer T1 in the conventional analysis, whereas the recurrence formula analysis suppressed this variation.</p><p><strong>Conclusion: </strong>T1 analysis of PCTIP using the recurrence formula analysis achieved accurate and precise T1 measurements, even for irregular heart rate variability.</p>","PeriodicalId":18119,"journal":{"name":"Magnetic Resonance in Medical Sciences","volume":" ","pages":"1-9"},"PeriodicalIF":2.5,"publicationDate":"2025-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11733503/pdf/","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"10500239","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"OA","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Numerical and Clinical Evaluation of the Robustness of Open-source Networks for Parallel MR Imaging Reconstruction. 并行磁共振成像重建开源网络的鲁棒性数值和临床评估。
IF 2.5 3区 医学 Q2 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2024-10-01 Epub Date: 2023-07-28 DOI: 10.2463/mrms.mp.2023-0031
Naoto Fujita, Suguru Yokosawa, Toru Shirai, Yasuhiko Terada

Purpose: Deep neural networks (DNNs) for MRI reconstruction often require large datasets for training. Still, in clinical settings, the domains of datasets are diverse, and how robust DNNs are to domain differences between training and testing datasets has been an open question. Here, we numerically and clinically evaluate the generalization of the reconstruction networks across various domains under clinically practical conditions and provide practical guidance on what points to consider when selecting models for clinical application.

Methods: We compare the reconstruction performance between four network models: U-Net, the deep cascade of convolutional neural networks (DC-CNNs), Hybrid Cascade, and variational network (VarNet). We used the public multicoil dataset fastMRI for training and testing and performed a single-domain test, where the domains of the dataset used for training and testing were the same, and cross-domain tests, where the source and target domains were different. We conducted a single-domain test (Experiment 1) and cross-domain tests (Experiments 2-4), focusing on six factors (the number of images, sampling pattern, acceleration factor, noise level, contrast, and anatomical structure) both numerically and clinically.

Results: U-Net had lower performance than the three model-based networks and was less robust to domain shifts between training and testing datasets. VarNet had the highest performance and robustness among the three model-based networks, followed by Hybrid Cascade and DC-CNN. Especially, VarNet showed high performance even with a limited number of training images (200 images/10 cases). U-Net was more robust to domain shifts concerning noise level than the other model-based networks. Hybrid Cascade showed slightly better performance and robustness than DC-CNN, except for robustness to noise-level domain shifts. The results of the clinical evaluations generally agreed with the results of the quantitative metrics.

Conclusion: In this study, we numerically and clinically evaluated the robustness of the publicly available networks using the multicoil data. Therefore, this study provided practical guidance for clinical applications.

目的:用于磁共振成像重建的深度神经网络(DNN)通常需要大型数据集进行训练。然而,在临床环境中,数据集的领域多种多样,DNN 对训练数据集和测试数据集之间的领域差异的鲁棒性如何一直是个未决问题。在此,我们从数值和临床角度评估了重建网络在临床实际条件下对不同领域的泛化能力,并就临床应用中选择模型时应考虑的要点提供了实际指导:我们比较了四种网络模型的重建性能:方法:我们比较了四种网络模型的重建性能:U-Net、深度级联卷积神经网络(DC-CNNs)、混合级联和变异网络(VarNet)。我们使用公开的多卷积数据集 fastMRI 进行训练和测试,并进行了单域测试(用于训练和测试的数据集域相同)和跨域测试(源域和目标域不同)。我们进行了单域测试(实验 1)和跨域测试(实验 2-4),重点关注六个因素(图像数量、采样模式、加速因子、噪声水平、对比度和解剖结构)的数值和临床表现:结果:U-Net 的性能低于三个基于模型的网络,而且对训练和测试数据集之间的领域转移的鲁棒性较差。在三种基于模型的网络中,VarNet 的性能和鲁棒性最高,其次是混合级联和 DC-CNN。特别是,VarNet 即使在训练图像数量有限(200 张图像/10 个案例)的情况下也表现出了很高的性能。与其他基于模型的网络相比,U-Net 对噪声水平的域偏移具有更强的鲁棒性。混合级联的性能和鲁棒性略优于 DC-CNN,但对噪声级域偏移的鲁棒性除外。临床评估结果与定量指标结果基本一致:在这项研究中,我们利用多扰数据对公开网络的鲁棒性进行了数值和临床评估。因此,本研究为临床应用提供了实际指导。
{"title":"Numerical and Clinical Evaluation of the Robustness of Open-source Networks for Parallel MR Imaging Reconstruction.","authors":"Naoto Fujita, Suguru Yokosawa, Toru Shirai, Yasuhiko Terada","doi":"10.2463/mrms.mp.2023-0031","DOIUrl":"10.2463/mrms.mp.2023-0031","url":null,"abstract":"<p><strong>Purpose: </strong>Deep neural networks (DNNs) for MRI reconstruction often require large datasets for training. Still, in clinical settings, the domains of datasets are diverse, and how robust DNNs are to domain differences between training and testing datasets has been an open question. Here, we numerically and clinically evaluate the generalization of the reconstruction networks across various domains under clinically practical conditions and provide practical guidance on what points to consider when selecting models for clinical application.</p><p><strong>Methods: </strong>We compare the reconstruction performance between four network models: U-Net, the deep cascade of convolutional neural networks (DC-CNNs), Hybrid Cascade, and variational network (VarNet). We used the public multicoil dataset fastMRI for training and testing and performed a single-domain test, where the domains of the dataset used for training and testing were the same, and cross-domain tests, where the source and target domains were different. We conducted a single-domain test (Experiment 1) and cross-domain tests (Experiments 2-4), focusing on six factors (the number of images, sampling pattern, acceleration factor, noise level, contrast, and anatomical structure) both numerically and clinically.</p><p><strong>Results: </strong>U-Net had lower performance than the three model-based networks and was less robust to domain shifts between training and testing datasets. VarNet had the highest performance and robustness among the three model-based networks, followed by Hybrid Cascade and DC-CNN. Especially, VarNet showed high performance even with a limited number of training images (200 images/10 cases). U-Net was more robust to domain shifts concerning noise level than the other model-based networks. Hybrid Cascade showed slightly better performance and robustness than DC-CNN, except for robustness to noise-level domain shifts. The results of the clinical evaluations generally agreed with the results of the quantitative metrics.</p><p><strong>Conclusion: </strong>In this study, we numerically and clinically evaluated the robustness of the publicly available networks using the multicoil data. Therefore, this study provided practical guidance for clinical applications.</p>","PeriodicalId":18119,"journal":{"name":"Magnetic Resonance in Medical Sciences","volume":" ","pages":"460-478"},"PeriodicalIF":2.5,"publicationDate":"2024-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11447470/pdf/","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"9897830","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"OA","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Diagnostic Utility of an Adjusted DWI Lexicon Using Multiple b-values to Evaluate Breast Lesions in Combination with BI-RADS. 使用多个 b 值评估乳腺病变的调整 DWI 词典与 BI-RADS 的诊断效用。
IF 2.5 3区 医学 Q2 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2024-10-01 Epub Date: 2023-05-26 DOI: 10.2463/mrms.mp.2022-0056
Aika Okazawa, Mami Iima, Masako Kataoka, Ryosuke Okumura, Sachiko Takahara, Tomotaka Noda, Taro Nishi, Takayoshi Ishimori, Yuji Nakamoto

Purpose: We aimed to investigate the diagnostic feasibility of an adjusted diffusion-weighted imaging (DWI) lexicon using multiple b values to assess breast lesions according to DWI-based breast imaging reporting and data system (BI-RADS).

Methods: This Institutional Review Board (IRB)-approved prospective study included 127 patients with suspected breast cancer. Breast MRI was performed using a 3T scanner. Breast DW images were acquired using five b-values of 0, 200, 800, 1000, and 1500 s/mm2 (5b-value DWI) on 3T MRI. Two readers independently assessed lesion characteristics and normal breast tissue using DWI alone (5b-value DWI and 2b-value DWI with b = 0 and 800 s/mm2) according to DWI-based BI-RADS and in combination with the standard dynamic contrast-enhanced images (combined MRI). Interobserver and intermethod agreements were assessed using kappa statistics. The specificity and sensitivity of lesion classification were evaluated.

Results: Ninety-five breast lesions (39 malignant and 56 benign) were evaluated. Interobserver agreement for lesion assessment on 5b-value DWI was very good (k ≥ 0.82) for DWI-based BI-RADS categories, lesion type, and mass characteristics; good (k = 0.75) in breast composition; and moderate (k ≥ 0.44) in background parenchymal signal (BPS) and non-mass distribution. Intermethod agreement between assessments performed using either 5b-value DWI or combined MRI was good-to-moderate (k = 0.52-0.67) for lesion type; moderate (k = 0.49-0.59) for DWI-based BI-RADS category and mass characteristics; and fair (k = 0.25-0.40) for mass shape, BPS, and breast composition. The sensitivity and positive predictive values (PPVs) for 5b-value DWI were 79.5%, 84.6% and 60.8%, 61.1% for each reader, respectively; 74.4%, 74.4% and 63.0%, 61.7% for 2b-value DWI; and 97.4%, 97.4% and 73.1%, 76.0% for combined MRI. The specificity and negative predictive values (NPVs) were 64.3%, 62.5% and 81.8%, 85.4% for 5b-value DWI; 69.6%, 67.9% and 79.6%, 79.2% for 2b-value DWI; and 75.0%, 78.6% and 97.7%, 97.8% for combined MRI.

Conclusion: Good observer agreement was observed in the 5b-value DWI. The 5b-value DWI based on multiple b-values might have the potential to complement the 2b-value DWI; however, their diagnostic performance tended to be inferior to that of combined MRI for the characterization of breast tumors.

目的:我们旨在研究根据基于 DWI 的乳腺成像报告和数据系统(BI-RADS),使用多 b 值评估乳腺病变的调整后扩散加权成像(DWI)词典的诊断可行性:这项经机构审查委员会(IRB)批准的前瞻性研究纳入了127名疑似乳腺癌患者。使用 3T 扫描仪进行乳腺 MRI 扫描。在 3T 磁共振成像上使用 0、200、800、1000 和 1500 s/mm2 五个 b 值(5b 值 DWI)采集乳腺 DW 图像。两名阅读者根据基于 DWI 的 BI-RADS 并结合标准动态对比增强图像(联合 MRI),使用单独的 DWI(5b 值 DWI 和 b = 0 和 800 s/mm2 的 2b 值 DWI)独立评估病变特征和正常乳腺组织。使用卡帕统计评估了观察者之间和方法之间的一致性。对病变分类的特异性和敏感性进行了评估:结果:共评估了 95 例乳腺病变(39 例恶性,56 例良性)。在基于 DWI 的 BI-RADS 分类、病变类型和肿块特征方面,5b 值 DWI 病变评估的观察者间一致性非常好(k ≥ 0.82);在乳腺组成方面,一致性良好(k = 0.75);在背景实质信号(BPS)和非肿块分布方面,一致性中等(k ≥ 0.44)。在病变类型方面,使用 5b 值 DWI 或联合 MRI 进行的评估之间的方法间一致性为良好至中等(k = 0.52-0.67);在基于 DWI 的 BI-RADS 类别和肿块特征方面为中等(k = 0.49-0.59);在肿块形状、BPS 和乳腺组成方面为一般(k = 0.25-0.40)。每位读者对 5b 值 DWI 的敏感性和阳性预测值(PPV)分别为 79.5%、84.6% 和 60.8%、61.1%;对 2b 值 DWI 的敏感性和阳性预测值(PPV)分别为 74.4%、74.4% 和 63.0%、61.7%;对联合 MRI 的敏感性和阳性预测值(PPV)分别为 97.4%、97.4% 和 73.1%、76.0%。5b值DWI的特异性和阴性预测值(NPV)分别为64.3%、62.5%和81.8%、85.4%;2b值DWI的特异性和阴性预测值(NPV)分别为69.6%、67.9%和79.6%、79.2%;联合MRI的特异性和阴性预测值(NPV)分别为75.0%、78.6%和97.7%、97.8%:结论:在 5b 值 DWI 中观察到了良好的观察者一致性。基于多个 b 值的 5b 值 DWI 可能具有补充 2b 值 DWI 的潜力;但是,就乳腺肿瘤的特征描述而言,它们的诊断性能往往不如联合 MRI。
{"title":"Diagnostic Utility of an Adjusted DWI Lexicon Using Multiple b-values to Evaluate Breast Lesions in Combination with BI-RADS.","authors":"Aika Okazawa, Mami Iima, Masako Kataoka, Ryosuke Okumura, Sachiko Takahara, Tomotaka Noda, Taro Nishi, Takayoshi Ishimori, Yuji Nakamoto","doi":"10.2463/mrms.mp.2022-0056","DOIUrl":"10.2463/mrms.mp.2022-0056","url":null,"abstract":"<p><strong>Purpose: </strong>We aimed to investigate the diagnostic feasibility of an adjusted diffusion-weighted imaging (DWI) lexicon using multiple b values to assess breast lesions according to DWI-based breast imaging reporting and data system (BI-RADS).</p><p><strong>Methods: </strong>This Institutional Review Board (IRB)-approved prospective study included 127 patients with suspected breast cancer. Breast MRI was performed using a 3T scanner. Breast DW images were acquired using five b-values of 0, 200, 800, 1000, and 1500 s/mm<sup>2</sup> (5b-value DWI) on 3T MRI. Two readers independently assessed lesion characteristics and normal breast tissue using DWI alone (5b-value DWI and 2b-value DWI with b = 0 and 800 s/mm<sup>2</sup>) according to DWI-based BI-RADS and in combination with the standard dynamic contrast-enhanced images (combined MRI). Interobserver and intermethod agreements were assessed using kappa statistics. The specificity and sensitivity of lesion classification were evaluated.</p><p><strong>Results: </strong>Ninety-five breast lesions (39 malignant and 56 benign) were evaluated. Interobserver agreement for lesion assessment on 5b-value DWI was very good (k ≥ 0.82) for DWI-based BI-RADS categories, lesion type, and mass characteristics; good (k = 0.75) in breast composition; and moderate (k ≥ 0.44) in background parenchymal signal (BPS) and non-mass distribution. Intermethod agreement between assessments performed using either 5b-value DWI or combined MRI was good-to-moderate (k = 0.52-0.67) for lesion type; moderate (k = 0.49-0.59) for DWI-based BI-RADS category and mass characteristics; and fair (k = 0.25-0.40) for mass shape, BPS, and breast composition. The sensitivity and positive predictive values (PPVs) for 5b-value DWI were 79.5%, 84.6% and 60.8%, 61.1% for each reader, respectively; 74.4%, 74.4% and 63.0%, 61.7% for 2b-value DWI; and 97.4%, 97.4% and 73.1%, 76.0% for combined MRI. The specificity and negative predictive values (NPVs) were 64.3%, 62.5% and 81.8%, 85.4% for 5b-value DWI; 69.6%, 67.9% and 79.6%, 79.2% for 2b-value DWI; and 75.0%, 78.6% and 97.7%, 97.8% for combined MRI.</p><p><strong>Conclusion: </strong>Good observer agreement was observed in the 5b-value DWI. The 5b-value DWI based on multiple b-values might have the potential to complement the 2b-value DWI; however, their diagnostic performance tended to be inferior to that of combined MRI for the characterization of breast tumors.</p>","PeriodicalId":18119,"journal":{"name":"Magnetic Resonance in Medical Sciences","volume":" ","pages":"438-448"},"PeriodicalIF":2.5,"publicationDate":"2024-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11447468/pdf/","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"9526055","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"OA","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Nonlinear Gradient Field Mapping Using a Spherical Grid Phantom for 3 and 7 Tesla MR Imaging Systems Equipped with High-performance Gradient Coils. 为配备高性能梯度线圈的 3 特斯拉和 7 特斯拉磁共振成像系统使用球形网格模型进行非线性梯度场绘图。
IF 2.5 3区 医学 Q2 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2024-10-01 Epub Date: 2023-09-09 DOI: 10.2463/mrms.tn.2023-0063
Ryoichi Kose, Katsumi Kose, Koji Fujimoto, Tomohisa Okada, Daiki Tamada, Utaroh Motosugi

Recent high-performance gradient coils are fabricated mainly at the expense of spatial linearity. In this study, we measured the spatial nonlinearity of the magnetic field generated by the gradient coils of two MRI systems with high-performance gradient coils. The nonlinearity of the gradient fields was measured using 3D gradient echo sequences and a spherical phantom with a built-in lattice structure. The spatial variation of the gradient field was approximated to the 3rd order polynomials. The coefficients of the polynomials were calculated using the steepest descent method. The geometric distortion of the acquired 3D MR images was corrected using the polynomials and compared with the 3D images corrected using the harmonic functions provided by the MRI venders. As a result, it was found that the nonlinearity correction formulae provided by the vendors were insufficient and needed to be verified or corrected using a geometric phantom such as used in this study.

最近制造的高性能梯度线圈主要以牺牲空间线性为代价。在这项研究中,我们测量了两个配备高性能梯度线圈的磁共振成像系统的梯度线圈所产生磁场的空间非线性。我们使用三维梯度回波序列和内置晶格结构的球形模型测量了梯度场的非线性。梯度场的空间变化近似于三阶多项式。多项式系数采用最陡下降法计算。获得的三维磁共振图像的几何失真使用多项式进行了校正,并与使用磁共振成像供应商提供的谐函数校正的三维图像进行了比较。结果发现,供应商提供的非线性校正公式不够充分,需要使用本研究中使用的几何模型进行验证或校正。
{"title":"Nonlinear Gradient Field Mapping Using a Spherical Grid Phantom for 3 and 7 Tesla MR Imaging Systems Equipped with High-performance Gradient Coils.","authors":"Ryoichi Kose, Katsumi Kose, Koji Fujimoto, Tomohisa Okada, Daiki Tamada, Utaroh Motosugi","doi":"10.2463/mrms.tn.2023-0063","DOIUrl":"10.2463/mrms.tn.2023-0063","url":null,"abstract":"<p><p>Recent high-performance gradient coils are fabricated mainly at the expense of spatial linearity. In this study, we measured the spatial nonlinearity of the magnetic field generated by the gradient coils of two MRI systems with high-performance gradient coils. The nonlinearity of the gradient fields was measured using 3D gradient echo sequences and a spherical phantom with a built-in lattice structure. The spatial variation of the gradient field was approximated to the 3rd order polynomials. The coefficients of the polynomials were calculated using the steepest descent method. The geometric distortion of the acquired 3D MR images was corrected using the polynomials and compared with the 3D images corrected using the harmonic functions provided by the MRI venders. As a result, it was found that the nonlinearity correction formulae provided by the vendors were insufficient and needed to be verified or corrected using a geometric phantom such as used in this study.</p>","PeriodicalId":18119,"journal":{"name":"Magnetic Resonance in Medical Sciences","volume":" ","pages":"525-536"},"PeriodicalIF":2.5,"publicationDate":"2024-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11447462/pdf/","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"10553809","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"OA","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Evaluating T1, T2 Relaxation, and Proton Density in Normal Brain Using Synthetic MRI with Fast Imaging Protocol. 使用快速成像协议合成磁共振成像评估正常大脑的 T1、T2 弛豫和质子密度
IF 2.5 3区 医学 Q2 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2024-10-01 Epub Date: 2023-09-08 DOI: 10.2463/mrms.tn.2022-0161
Zuofeng Zheng, Yawen Liu, Hongxia Yin, Pengling Ren, Tingting Zhang, Jiafei Yang, Zhenchang Wang

Synthetic MRI is being increasingly used for the quantification of brain longitudinal relaxation time (T1), transverse relaxation time (T2), and proton density (PD) values. However, the effect of fast imaging protocols on these quantitative values has not been fully estimated. The purpose of this study was to investigate the effect of fast scan parameters on T1, T2, and PD measured with a multi-dynamic multi-echo (MDME) sequence of normal brain at 3.0T. Thirty-four volunteers were scanned using 3 MDME sequences with different scan times (named Fast, 2 min, 29 sec; Routine, 4 min, 07 sec; and Research, 7 min, 46 sec, respectively). The measured T1, T2, and PD in 18 volumes of interest (VOI) of brain were compared between the 3 sequences using rank sum test, t test, coefficients of variation (CVs) analysis, correlation analysis, and Bland-Altman analysis. We found that even though T1, T2, and PD were significantly different between the 3 sequences in most of the brain regions, the intersequence CVs were relatively low and linear correlation were high. Bland-Altman plots showed that most of the values fall within the 95% prediction limits. We concluded that fast imaging protocols of MDME sequence used in our study can potentially be used for quantitative evaluation of brain tissues. Since changing scan parameters can affect the measured T1, T2, and PD values, it is necessary to use consistent scan parameter for comparing or following up cases quantitatively.

合成 MRI 越来越多地被用于量化大脑纵向弛豫时间(T1)、横向弛豫时间(T2)和质子密度(PD)值。然而,快速成像方案对这些定量值的影响尚未得到充分估计。本研究旨在研究快速扫描参数对 3.0T 下正常大脑多动态多回波(MDME)序列测量的 T1、T2 和 PD 的影响。研究人员使用 3 个扫描时间不同的 MDME 序列(分别为快速,2 分 29 秒;常规,4 分 07 秒;研究,7 分 46 秒)对 34 名志愿者进行了扫描。使用秩和检验、t检验、变异系数(CVs)分析、相关分析和布兰-阿尔特曼(Bland-Altman)分析比较了三种序列在大脑18个感兴趣容积(VOI)中测得的T1、T2和PD。我们发现,尽管 3 个序列的 T1、T2 和 PD 在大多数脑区有显著差异,但序列间变异系数相对较低,线性相关性较高。Bland-Altman图显示,大多数数值都在95%的预测范围内。我们的结论是,研究中使用的 MDME 序列快速成像方案可用于脑组织的定量评估。由于改变扫描参数会影响测量的 T1、T2 和 PD 值,因此有必要使用一致的扫描参数对病例进行定量比较或随访。
{"title":"Evaluating T1, T2 Relaxation, and Proton Density in Normal Brain Using Synthetic MRI with Fast Imaging Protocol.","authors":"Zuofeng Zheng, Yawen Liu, Hongxia Yin, Pengling Ren, Tingting Zhang, Jiafei Yang, Zhenchang Wang","doi":"10.2463/mrms.tn.2022-0161","DOIUrl":"10.2463/mrms.tn.2022-0161","url":null,"abstract":"<p><p>Synthetic MRI is being increasingly used for the quantification of brain longitudinal relaxation time (T1), transverse relaxation time (T2), and proton density (PD) values. However, the effect of fast imaging protocols on these quantitative values has not been fully estimated. The purpose of this study was to investigate the effect of fast scan parameters on T1, T2, and PD measured with a multi-dynamic multi-echo (MDME) sequence of normal brain at 3.0T. Thirty-four volunteers were scanned using 3 MDME sequences with different scan times (named Fast, 2 min, 29 sec; Routine, 4 min, 07 sec; and Research, 7 min, 46 sec, respectively). The measured T1, T2, and PD in 18 volumes of interest (VOI) of brain were compared between the 3 sequences using rank sum test, t test, coefficients of variation (CVs) analysis, correlation analysis, and Bland-Altman analysis. We found that even though T1, T2, and PD were significantly different between the 3 sequences in most of the brain regions, the intersequence CVs were relatively low and linear correlation were high. Bland-Altman plots showed that most of the values fall within the 95% prediction limits. We concluded that fast imaging protocols of MDME sequence used in our study can potentially be used for quantitative evaluation of brain tissues. Since changing scan parameters can affect the measured T1, T2, and PD values, it is necessary to use consistent scan parameter for comparing or following up cases quantitatively.</p>","PeriodicalId":18119,"journal":{"name":"Magnetic Resonance in Medical Sciences","volume":" ","pages":"514-524"},"PeriodicalIF":2.5,"publicationDate":"2024-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11447464/pdf/","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"10256637","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"OA","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
A Versatile MR Elastography Research Tool with a Modified Motion Signal-to-noise Ratio Approach. 采用修正运动信噪比方法的多功能磁共振弹性成像研究工具
IF 2.5 3区 医学 Q2 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2024-10-01 Epub Date: 2023-04-12 DOI: 10.2463/mrms.mp.2022-0149
Daiki Ito, Tetsushi Habe, Tomokazu Numano, Shigeo Okuda, Shigeyoshi Soga, Masahiro Jinzaki

Purpose: This study aimed to facilitate research progress in MR elastography (MRE) by providing a versatile and convenient application for MRE reconstruction, namely the MRE research tool (MRE-rTool). It can be used for a series of MRE image analyses, including phase unwrapping, arbitrary bandpass and directional filtering, noise assessment of the wave propagation image (motion SNR), and reconstruction of the elastogram in both 2D and 3D MRE acquisitions. To reinforce the versatility of MRE-rTool, the conventional method of motion SNR was modified into a new method that reflects the effects of image filtering.

Methods: MRE tests of the phantom and liver were performed using different estimation algorithms for stiffness value (algebraic inversion of the differential equation [AIDE], local frequency estimation [LFE] in MRE-rTool, and multimodel direct inversion [MMDI] in clinical reconstruction) and acquiring dimensions (2D and 3D acquisitions). This study also tested the accuracy of masking low SNR regions using modified and conventional motion SNR under various mechanical vibration powers.

Results: The stiffness values estimated using AIDE/LFE in MRE-rTool were comparable to that of MMDI (phantom, 3.71 ± 0.74, 3.60 ± 0.32, and 3.60 ± 0.54 kPa in AIDE, LFE, and MMDI; liver, 2.26 ± 0.31, 2.74 ± 0.16, and 2.21 ± 0.26 kPa in AIDE, LFE, and MMDI). The stiffness value in 3D acquisition was independent of the direction of the motion-encoding gradient and was more accurate than that of 2D acquisition. The masking of low SNR regions using the modified motion SNR worked better than that in the conventional motion SNR for each vibration power, especially when using a directional filter.

Conclusion: The performance of MRE-rTool on test data reached the level required in clinical MRE studies. MRE-rTool has the potential to facilitate MRE research, contribute to the future development of MRE, and has been freely released online.

目的:本研究旨在通过提供一个多功能、便捷的磁共振弹性成像重建应用程序,即磁共振弹性成像研究工具(MRE-rTool),促进磁共振弹性成像(MRE)的研究进展。该工具可用于一系列 MRE 图像分析,包括相位解包、任意带通和方向滤波、波传播图像的噪声评估(运动信噪比)以及二维和三维 MRE 采集的弹性图重建。为了加强 MRE-rTool 的多功能性,将传统的运动 SNR 方法修改为一种新方法,以反映图像滤波的效果:方法:使用不同的刚度值估计算法(微分方程代数反演[AIDE]、MRE-rTool 中的局部频率估计[LFE]和临床重建中的多模型直接反演[MMDI])和采集维度(二维和三维采集)对模型和肝脏进行了 MRE 测试。本研究还测试了在不同机械振动功率下使用修正和传统运动 SNR 屏蔽低 SNR 区域的准确性:在MRE-rTool中使用AIDE/LFE估算的硬度值与MMDI相当(在AIDE、LFE和MMDI中,模型的硬度值分别为3.71±0.74、3.60±0.32和3.60±0.54 kPa;在AIDE、LFE和MMDI中,肝脏的硬度值分别为2.26±0.31、2.74±0.16和2.21±0.26 kPa)。三维采集的硬度值与运动编码梯度的方向无关,比二维采集更准确。使用修正运动信噪比对低信噪比区域进行屏蔽的效果优于使用传统运动信噪比对各种振动功率进行屏蔽的效果,尤其是在使用定向滤波器时:结论:MRE-rTool 在测试数据上的表现达到了临床 MRE 研究的要求。MRE-rTool 有潜力促进 MRE 研究,为 MRE 的未来发展做出贡献,并已在网上免费发布。
{"title":"A Versatile MR Elastography Research Tool with a Modified Motion Signal-to-noise Ratio Approach.","authors":"Daiki Ito, Tetsushi Habe, Tomokazu Numano, Shigeo Okuda, Shigeyoshi Soga, Masahiro Jinzaki","doi":"10.2463/mrms.mp.2022-0149","DOIUrl":"10.2463/mrms.mp.2022-0149","url":null,"abstract":"<p><strong>Purpose: </strong>This study aimed to facilitate research progress in MR elastography (MRE) by providing a versatile and convenient application for MRE reconstruction, namely the MRE research tool (MRE-rTool). It can be used for a series of MRE image analyses, including phase unwrapping, arbitrary bandpass and directional filtering, noise assessment of the wave propagation image (motion SNR), and reconstruction of the elastogram in both 2D and 3D MRE acquisitions. To reinforce the versatility of MRE-rTool, the conventional method of motion SNR was modified into a new method that reflects the effects of image filtering.</p><p><strong>Methods: </strong>MRE tests of the phantom and liver were performed using different estimation algorithms for stiffness value (algebraic inversion of the differential equation [AIDE], local frequency estimation [LFE] in MRE-rTool, and multimodel direct inversion [MMDI] in clinical reconstruction) and acquiring dimensions (2D and 3D acquisitions). This study also tested the accuracy of masking low SNR regions using modified and conventional motion SNR under various mechanical vibration powers.</p><p><strong>Results: </strong>The stiffness values estimated using AIDE/LFE in MRE-rTool were comparable to that of MMDI (phantom, 3.71 ± 0.74, 3.60 ± 0.32, and 3.60 ± 0.54 kPa in AIDE, LFE, and MMDI; liver, 2.26 ± 0.31, 2.74 ± 0.16, and 2.21 ± 0.26 kPa in AIDE, LFE, and MMDI). The stiffness value in 3D acquisition was independent of the direction of the motion-encoding gradient and was more accurate than that of 2D acquisition. The masking of low SNR regions using the modified motion SNR worked better than that in the conventional motion SNR for each vibration power, especially when using a directional filter.</p><p><strong>Conclusion: </strong>The performance of MRE-rTool on test data reached the level required in clinical MRE studies. MRE-rTool has the potential to facilitate MRE research, contribute to the future development of MRE, and has been freely released online.</p>","PeriodicalId":18119,"journal":{"name":"Magnetic Resonance in Medical Sciences","volume":" ","pages":"417-427"},"PeriodicalIF":2.5,"publicationDate":"2024-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11447463/pdf/","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"9644589","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"OA","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Blockage of CSF Outflow in Rats after Deep Cervical Lymph Node Ligation Observed Using Gd-based MR Imaging. 利用钆基磁共振成像观察深颈部淋巴结结扎后大鼠脑脊液流出受阻情况
IF 2.5 3区 医学 Q2 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2024-10-01 Epub Date: 2023-05-30 DOI: 10.2463/mrms.mp.2023-0023
Naoya Kinota, Hiroyuki Kameda, Bai Xiawei, Takaaki Fujii, Daisuke Kato, Bunya Takahashi, Ryo Morita, Daisuke Abo, Ryusei Majima, Hiroshi Ishii, Kazuyuki Minowa, Kohsuke Kudo

Purpose: To investigate whether deep cervical lymph node (DCLN) ligation alters intracranial cerebrospinal fluid (CSF) tracer dynamics and outflow using a rat model with intrathecal dynamic contrast-enhanced (DCE) MRI.

Methods: Six bilateral DCLN-ligated and six sham-operated rats were subjected to DCE MRI with Gd-BTDO3A, and dynamic T1-weighted images were acquired. ROIs were collected from the CSF at the C1 level (CSF_C1), CSF between the olfactory bulbs (CSF_OB), CSF at the pituitary recess (CSF_PitR), and CSF at the pineal recess (CSF_PinR), upper nasal turbinate (UNT), olfactory bulbs, cerebrum, and the jugular region. Time-intensity curves were evaluated, and the maximum slope, peak timing, peak signal ratio, and elimination half-life for the four CSF ROIs and UNT were calculated and compared.

Results: Delayed tracer arrival in the rostral CSF space and the nasal cavity with tracer retention in the ventral CSF space were observed in the ligation group. The maximum slopes were smaller in the ligation group at UNT (sham: 0.075 ± 0.0061, ligation: 0.044 ± 0.0086/min, P = 0.011). A significant difference was not detected in peak timings. The peak signal ratio values were lower in the ligation group at UNT (sham: 2.12 ± 0.19, ligation: 1.72 ± 0.11, P = 0.011). The elimination half-life was delayed in the ligation group at CSF_C1 (sham: 30.5 ± 2.70, ligation: 44.4 ± 12.6 min, P = 0.043), CSF_OB (sham: 30.2 ± 2.67, ligation: 44.8 ± 7.47 min, P = 0.021), and CSF_PitR (sham: 30.2 ± 2.49, ligation: 41.3 ± 7.57 min, P = 0.021).

Conclusion: The DCLN ligation in rats blocked CSF outflow into the nasal cavity and caused CSF retention.

目的:通过大鼠鞘内动态对比增强(DCE)磁共振成像(MRI)研究颈深淋巴结(DCLN)结扎是否会改变颅内脑脊液(CSF)示踪剂的动态变化和流出:对 6 只双侧 DCLN 结扎大鼠和 6 只假手术大鼠进行 Gd-BTDO3A 的 DCE MRI 检查,并获取动态 T1 加权图像。从C1水平的CSF(CSF_C1)、嗅球之间的CSF(CSF_OB)、垂体凹陷处的CSF(CSF_PitR)、松果体凹陷处的CSF(CSF_PinR)、上鼻甲(UNT)、嗅球、大脑和颈静脉区域采集ROI。评估了时间强度曲线,计算并比较了四个 CSF ROI 和 UNT 的最大斜率、峰值时间、峰值信号比和消除半衰期:结果:结扎组观察到示踪剂延迟到达喙侧 CSF 间隙和鼻腔,并在腹侧 CSF 间隙滞留。结扎组在 UNT 时的最大斜率较小(假性:0.075 ± 0.0061,结扎:0.044 ± 0.0086/分钟,P = 0.011)。在峰值时间上未发现明显差异。结扎组的 UNT 峰值信号比值较低(假体:2.12 ± 0.19,结扎:1.72 ± 0.11,P = 0.011)。结扎组在 CSF_C1(假:30.5 ± 2.70,结扎:44.4 ± 12.6 分钟,P = 0.043)、CSF_OB(假:30.2 ± 2.67,结扎:44.8 ± 7.47 分钟,P = 0.021)和 CSF_PitR (假:30.2 ± 2.49,结扎:41.3 ± 7.57 分钟,P = 0.021)处的消除半衰期延迟:结论:对大鼠进行 DCLN 结扎可阻止 CSF 流出鼻腔并导致 CSF 滞留。
{"title":"Blockage of CSF Outflow in Rats after Deep Cervical Lymph Node Ligation Observed Using Gd-based MR Imaging.","authors":"Naoya Kinota, Hiroyuki Kameda, Bai Xiawei, Takaaki Fujii, Daisuke Kato, Bunya Takahashi, Ryo Morita, Daisuke Abo, Ryusei Majima, Hiroshi Ishii, Kazuyuki Minowa, Kohsuke Kudo","doi":"10.2463/mrms.mp.2023-0023","DOIUrl":"10.2463/mrms.mp.2023-0023","url":null,"abstract":"<p><strong>Purpose: </strong>To investigate whether deep cervical lymph node (DCLN) ligation alters intracranial cerebrospinal fluid (CSF) tracer dynamics and outflow using a rat model with intrathecal dynamic contrast-enhanced (DCE) MRI.</p><p><strong>Methods: </strong>Six bilateral DCLN-ligated and six sham-operated rats were subjected to DCE MRI with Gd-BTDO3A, and dynamic T1-weighted images were acquired. ROIs were collected from the CSF at the C1 level (CSF_C1), CSF between the olfactory bulbs (CSF_OB), CSF at the pituitary recess (CSF_PitR), and CSF at the pineal recess (CSF_PinR), upper nasal turbinate (UNT), olfactory bulbs, cerebrum, and the jugular region. Time-intensity curves were evaluated, and the maximum slope, peak timing, peak signal ratio, and elimination half-life for the four CSF ROIs and UNT were calculated and compared.</p><p><strong>Results: </strong>Delayed tracer arrival in the rostral CSF space and the nasal cavity with tracer retention in the ventral CSF space were observed in the ligation group. The maximum slopes were smaller in the ligation group at UNT (sham: 0.075 ± 0.0061, ligation: 0.044 ± 0.0086/min, P = 0.011). A significant difference was not detected in peak timings. The peak signal ratio values were lower in the ligation group at UNT (sham: 2.12 ± 0.19, ligation: 1.72 ± 0.11, P = 0.011). The elimination half-life was delayed in the ligation group at CSF_C1 (sham: 30.5 ± 2.70, ligation: 44.4 ± 12.6 min, P = 0.043), CSF_OB (sham: 30.2 ± 2.67, ligation: 44.8 ± 7.47 min, P = 0.021), and CSF_PitR (sham: 30.2 ± 2.49, ligation: 41.3 ± 7.57 min, P = 0.021).</p><p><strong>Conclusion: </strong>The DCLN ligation in rats blocked CSF outflow into the nasal cavity and caused CSF retention.</p>","PeriodicalId":18119,"journal":{"name":"Magnetic Resonance in Medical Sciences","volume":" ","pages":"449-459"},"PeriodicalIF":2.5,"publicationDate":"2024-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11447471/pdf/","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"9553292","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"OA","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Comparison of Slab and Pencil Beam Labeling in Spin-labeled MR Imaging for Pancreatic Juice Flow Visualization. 自旋标记磁共振成像中胰腺汁流可视化的平板标记与铅笔束标记的比较
IF 2.5 3区 医学 Q2 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2024-10-01 Epub Date: 2023-06-21 DOI: 10.2463/mrms.tn.2022-0132
Hideki Hoshi, Reiji Sugita, Yoshio Machida

The usefulness of a highly targeted pencil beam (PB) label was compared with the commonly used slab label for direct visualization of pancreaticobiliary reflux using the time-spatial labeling inversion pulse (time-SLIP) technique. Signal profiles of flow phantom images obtained with a 1.5T MRI were analyzed. Both labels had similar labeling capabilities, but the edge characteristics of the PB label were blunt. Next, sixty-eight patients were classified into two groups according to the angle of the pancreaticobiliary ducts, and the displacement of the pancreatic ducts in respiratory fluctuation was measured. The results were approximately 7 mm in both groups. The blunt edge characteristics of the PB label suggest that it is robust to respiratory fluctuations. The overall labeling ability of the PB is comparable to that of the slab. In the larger angle of pancreaticobiliary ducts, the PB label may be able to label the pancreatic duct more selectively.

在使用时间空间标记反转脉冲(time-SLIP)技术直接观察胰胆管反流时,比较了高靶向性铅笔束(PB)标记与常用的平板标记的实用性。对使用 1.5T 磁共振成像技术获得的血流模型图像的信号剖面进行了分析。两种标签的标记能力相似,但 PB 标签的边缘特征较钝。接着,根据胰胆管的角度将 68 名患者分为两组,并测量了胰管在呼吸波动中的位移。结果两组患者的胰管位移量均约为 7 毫米。胰胆管标记的钝边特征表明它对呼吸波动有很强的抵抗力。PB 的整体标记能力与平板相当。在胰胆管角度较大的情况下,PB 标签可能能更有选择性地标记胰管。
{"title":"Comparison of Slab and Pencil Beam Labeling in Spin-labeled MR Imaging for Pancreatic Juice Flow Visualization.","authors":"Hideki Hoshi, Reiji Sugita, Yoshio Machida","doi":"10.2463/mrms.tn.2022-0132","DOIUrl":"10.2463/mrms.tn.2022-0132","url":null,"abstract":"<p><p>The usefulness of a highly targeted pencil beam (PB) label was compared with the commonly used slab label for direct visualization of pancreaticobiliary reflux using the time-spatial labeling inversion pulse (time-SLIP) technique. Signal profiles of flow phantom images obtained with a 1.5T MRI were analyzed. Both labels had similar labeling capabilities, but the edge characteristics of the PB label were blunt. Next, sixty-eight patients were classified into two groups according to the angle of the pancreaticobiliary ducts, and the displacement of the pancreatic ducts in respiratory fluctuation was measured. The results were approximately 7 mm in both groups. The blunt edge characteristics of the PB label suggest that it is robust to respiratory fluctuations. The overall labeling ability of the PB is comparable to that of the slab. In the larger angle of pancreaticobiliary ducts, the PB label may be able to label the pancreatic duct more selectively.</p>","PeriodicalId":18119,"journal":{"name":"Magnetic Resonance in Medical Sciences","volume":" ","pages":"502-513"},"PeriodicalIF":2.5,"publicationDate":"2024-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11447465/pdf/","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"9671398","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"OA","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Microstructural Abnormalities in the Contralateral Normal-appearing White Matter of Glioblastoma Patients Evaluated with Advanced Diffusion Imaging. 用高级弥散成像评估胶质母细胞瘤患者对侧正常白质的微结构异常。
IF 2.5 3区 医学 Q2 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2024-10-01 Epub Date: 2023-08-01 DOI: 10.2463/mrms.mp.2023-0054
Barış Genç, Kerim Aslan, Ali Özçağlayan, Lütfi İncesu

Purpose: Glioblastoma patients develop recurrence in the opposite hemisphere far from the primary tumor site even after complete resection. This is one of the main reasons for short disease survival. Our aim in this study is to detect microstructural changes in the contralateral hemisphere of glioblastoma patients using different diffusion models with the fully automated tract-based spatial statistics (TBSS) method.

Methods: Fourteen right-sided and eleven left-sided glioblastoma patients without any treatment and eighteen age- and gender-matched controls were included in the study. Multi-shell diffusion weighted images were created with a 3T MRI device. After various preprocessing steps, images of fractional anisotropy (FA), mean diffusivity (MD), axial diffusivity (AD), radial diffusivity (RD), axial kurtosis (AK), mean kurtosis (MK), radial kurtosis (RK), intracellular volume fraction (ICVF), orientation dispersion index (ODI), and isotropic water fraction (ISO) were obtained. TBSS was used to compare diffusion tensor imaging, diffusion kurtosis imaging, and neurite orientation dispersion and density imaging parameters of right- and left-sided glioblastoma patients with the control group for the contralateral hemisphere.

Results: Both right-sided and left-sided glioblastoma patients have shown an increase in MD and ODI in the contralateral hemisphere. While right-sided glioblastoma patients showed an increase in RD, AD, and ISO in a more limited area in the contralateral hemisphere, left-sided glioblastoma patients showed an increase in MK and AK. FA, ICVF, and RK did not show any difference in both groups.

Conclusion: There are microstructural changes in the contralateral hemisphere in glioblastoma patients, and these changes differ between right-sided and left-sided glioblastoma patients.

目的:胶质母细胞瘤患者即使在完全切除肿瘤后,仍会在远离原发肿瘤部位的对侧半球复发。这是导致患者生存期短的主要原因之一。本研究的目的是利用不同的弥散模型和全自动基于束的空间统计(TBSS)方法检测胶质母细胞瘤患者对侧半球的微结构变化:研究对象包括 14 名未接受任何治疗的右侧和 11 名左侧胶质母细胞瘤患者,以及 18 名年龄和性别匹配的对照组患者。使用 3T 磁共振成像设备生成多壳弥散加权图像。经过各种预处理步骤后,获得了分数各向异性(FA)、平均扩散率(MD)、轴向扩散率(AD)、径向扩散率(RD)、轴向峰度(AK)、平均峰度(MK)、径向峰度(RK)、细胞内体积分数(ICVF)、方向弥散指数(ODI)和各向同性水分数(ISO)的图像。使用 TBSS 比较右侧和左侧胶质母细胞瘤患者与对侧半球对照组的弥散张量成像、弥散峰度成像、神经元取向弥散和密度成像参数:结果:右侧和左侧胶质母细胞瘤患者对侧半球的MD和ODI均有所增加。右侧胶质母细胞瘤患者的 RD、AD 和 ISO 在对侧半球更有限的区域内增加,而左侧胶质母细胞瘤患者的 MK 和 AK 增加。两组患者的FA、ICVF和RK没有任何差异:结论:胶质母细胞瘤患者的对侧大脑半球存在微结构变化,右侧和左侧胶质母细胞瘤患者的这些变化有所不同。
{"title":"Microstructural Abnormalities in the Contralateral Normal-appearing White Matter of Glioblastoma Patients Evaluated with Advanced Diffusion Imaging.","authors":"Barış Genç, Kerim Aslan, Ali Özçağlayan, Lütfi İncesu","doi":"10.2463/mrms.mp.2023-0054","DOIUrl":"10.2463/mrms.mp.2023-0054","url":null,"abstract":"<p><strong>Purpose: </strong>Glioblastoma patients develop recurrence in the opposite hemisphere far from the primary tumor site even after complete resection. This is one of the main reasons for short disease survival. Our aim in this study is to detect microstructural changes in the contralateral hemisphere of glioblastoma patients using different diffusion models with the fully automated tract-based spatial statistics (TBSS) method.</p><p><strong>Methods: </strong>Fourteen right-sided and eleven left-sided glioblastoma patients without any treatment and eighteen age- and gender-matched controls were included in the study. Multi-shell diffusion weighted images were created with a 3T MRI device. After various preprocessing steps, images of fractional anisotropy (FA), mean diffusivity (MD), axial diffusivity (AD), radial diffusivity (RD), axial kurtosis (AK), mean kurtosis (MK), radial kurtosis (RK), intracellular volume fraction (ICVF), orientation dispersion index (ODI), and isotropic water fraction (ISO) were obtained. TBSS was used to compare diffusion tensor imaging, diffusion kurtosis imaging, and neurite orientation dispersion and density imaging parameters of right- and left-sided glioblastoma patients with the control group for the contralateral hemisphere.</p><p><strong>Results: </strong>Both right-sided and left-sided glioblastoma patients have shown an increase in MD and ODI in the contralateral hemisphere. While right-sided glioblastoma patients showed an increase in RD, AD, and ISO in a more limited area in the contralateral hemisphere, left-sided glioblastoma patients showed an increase in MK and AK. FA, ICVF, and RK did not show any difference in both groups.</p><p><strong>Conclusion: </strong>There are microstructural changes in the contralateral hemisphere in glioblastoma patients, and these changes differ between right-sided and left-sided glioblastoma patients.</p>","PeriodicalId":18119,"journal":{"name":"Magnetic Resonance in Medical Sciences","volume":" ","pages":"479-486"},"PeriodicalIF":2.5,"publicationDate":"2024-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11447469/pdf/","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"9926992","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"OA","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
Increased Parenchymal Free Water May Be Decreased by Revascularization Surgery in Patients with Moyamoya Disease. Moyamoya病患者血管重建手术可减少实质游离水的增加
IF 2.5 3区 医学 Q2 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Pub Date : 2024-10-01 Epub Date: 2023-04-18 DOI: 10.2463/mrms.mp.2022-0146
Shoko Hara, Masaaki Hori, Koji Kamagata, Christina Andica, Motoki Inaji, Yoji Tanaka, Shigeki Aoki, Tadashi Nariai, Taketoshi Maehara

Purpose: Moyamoya disease (MMD) is a cerebrovascular disease associated with steno-occlusive changes in the arteries of the circle of Willis and with hemodynamic impairment. Previous studies have reported that parenchymal extracellular free water levels may be increased and the number of neurites may be decreased in patients with MMD. The aim of the present study was to investigate the postoperative changes in parenchymal free water and neurites and their relationship with cognitive improvement.

Methods: Multi-shell diffusion MRI (neurite orientation dispersion and density imaging and free water imaging using a bi-tensor model) was performed in 15 hemispheres of 13 adult patients with MMD (11 female, mean age 37.9 years) who had undergone revascularization surgery as well as age- and sex-matched normal controls. Parameter maps of free water and free-water-eliminated neurites were created, and the regional parameter values were compared among controls, patients before surgery, and patients after surgery.

Results: The anterior and middle cerebral artery territories of patients showed higher preoperative free water levels (P ≤ 0.007) and lower postoperative free water levels (P ≤ 0.001) than those of normal controls. The change in the dispersion of the white matter in the anterior cerebral artery territory correlated with cognitive improvement (r = -0.75; P = 0.004).

Conclusion: Our study suggests that increased parenchymal free water levels decreased after surgery and that postoperative changes in neurite parameters are related to postoperative cognitive improvement in adult patients with MMD. Diffusion analytical methods separately calculating free water and neurites may be useful for unraveling the pathophysiology of chronic ischemia and the postoperative changes that occur after revascularization surgery in this disease population.

目的:莫亚莫亚病(MMD)是一种脑血管疾病,与威利斯圈动脉狭窄闭塞性病变和血液动力学损伤有关。以往的研究表明,MMD 患者的实质细胞外游离水含量可能会增加,神经元数量可能会减少。本研究旨在探讨术后实质游离水和神经元的变化及其与认知能力改善的关系:方法:对接受血管重建手术的13名成年多发性硬化症患者(11名女性,平均年龄37.9岁)的15个半球以及年龄和性别匹配的正常对照组进行了多壳体弥散核磁共振成像(神经元取向弥散和密度成像以及使用双张量模型的游离水成像)。绘制了游离水和游离水消除神经元的参数图,并比较了对照组、手术前患者和手术后患者的区域参数值:结果:与正常对照组相比,患者大脑前动脉和大脑中动脉区域的术前游离水含量更高(P≤0.007),术后游离水含量更低(P≤0.001)。大脑前动脉区域白质弥散度的变化与认知能力的改善相关(r = -0.75;P = 0.004):我们的研究表明,成年多发性硬化症患者术后实质游离水含量的增加会减少,术后神经元参数的变化与术后认知能力的改善有关。分别计算游离水和神经元的扩散分析方法可能有助于揭示慢性缺血的病理生理学以及该疾病人群血管重建手术后发生的术后变化。
{"title":"Increased Parenchymal Free Water May Be Decreased by Revascularization Surgery in Patients with Moyamoya Disease.","authors":"Shoko Hara, Masaaki Hori, Koji Kamagata, Christina Andica, Motoki Inaji, Yoji Tanaka, Shigeki Aoki, Tadashi Nariai, Taketoshi Maehara","doi":"10.2463/mrms.mp.2022-0146","DOIUrl":"10.2463/mrms.mp.2022-0146","url":null,"abstract":"<p><strong>Purpose: </strong>Moyamoya disease (MMD) is a cerebrovascular disease associated with steno-occlusive changes in the arteries of the circle of Willis and with hemodynamic impairment. Previous studies have reported that parenchymal extracellular free water levels may be increased and the number of neurites may be decreased in patients with MMD. The aim of the present study was to investigate the postoperative changes in parenchymal free water and neurites and their relationship with cognitive improvement.</p><p><strong>Methods: </strong>Multi-shell diffusion MRI (neurite orientation dispersion and density imaging and free water imaging using a bi-tensor model) was performed in 15 hemispheres of 13 adult patients with MMD (11 female, mean age 37.9 years) who had undergone revascularization surgery as well as age- and sex-matched normal controls. Parameter maps of free water and free-water-eliminated neurites were created, and the regional parameter values were compared among controls, patients before surgery, and patients after surgery.</p><p><strong>Results: </strong>The anterior and middle cerebral artery territories of patients showed higher preoperative free water levels (P ≤ 0.007) and lower postoperative free water levels (P ≤ 0.001) than those of normal controls. The change in the dispersion of the white matter in the anterior cerebral artery territory correlated with cognitive improvement (r = -0.75; P = 0.004).</p><p><strong>Conclusion: </strong>Our study suggests that increased parenchymal free water levels decreased after surgery and that postoperative changes in neurite parameters are related to postoperative cognitive improvement in adult patients with MMD. Diffusion analytical methods separately calculating free water and neurites may be useful for unraveling the pathophysiology of chronic ischemia and the postoperative changes that occur after revascularization surgery in this disease population.</p>","PeriodicalId":18119,"journal":{"name":"Magnetic Resonance in Medical Sciences","volume":" ","pages":"405-416"},"PeriodicalIF":2.5,"publicationDate":"2024-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11447467/pdf/","citationCount":null,"resultStr":null,"platform":"Semanticscholar","paperid":"9790547","PeriodicalName":null,"FirstCategoryId":null,"ListUrlMain":null,"RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":"OA","EPubDate":null,"PubModel":null,"JCR":null,"JCRName":null,"Score":null,"Total":0}
引用次数: 0
期刊
Magnetic Resonance in Medical Sciences
全部 Acc. Chem. Res. ACS Applied Bio Materials ACS Appl. Electron. Mater. ACS Appl. Energy Mater. ACS Appl. Mater. Interfaces ACS Appl. Nano Mater. ACS Appl. Polym. Mater. ACS BIOMATER-SCI ENG ACS Catal. ACS Cent. Sci. ACS Chem. Biol. ACS Chemical Health & Safety ACS Chem. Neurosci. ACS Comb. Sci. ACS Earth Space Chem. ACS Energy Lett. ACS Infect. Dis. ACS Macro Lett. ACS Mater. Lett. ACS Med. Chem. Lett. ACS Nano ACS Omega ACS Photonics ACS Sens. ACS Sustainable Chem. Eng. ACS Synth. Biol. Anal. Chem. BIOCHEMISTRY-US Bioconjugate Chem. BIOMACROMOLECULES Chem. Res. Toxicol. Chem. Rev. Chem. Mater. CRYST GROWTH DES ENERG FUEL Environ. Sci. Technol. Environ. Sci. Technol. Lett. Eur. J. Inorg. Chem. IND ENG CHEM RES Inorg. Chem. J. Agric. Food. Chem. J. Chem. Eng. Data J. Chem. Educ. J. Chem. Inf. Model. J. Chem. Theory Comput. J. Med. Chem. J. Nat. Prod. J PROTEOME RES J. Am. Chem. Soc. LANGMUIR MACROMOLECULES Mol. Pharmaceutics Nano Lett. Org. Lett. ORG PROCESS RES DEV ORGANOMETALLICS J. Org. Chem. J. Phys. Chem. J. Phys. Chem. A J. Phys. Chem. B J. Phys. Chem. C J. Phys. Chem. Lett. Analyst Anal. Methods Biomater. Sci. Catal. Sci. Technol. Chem. Commun. Chem. Soc. Rev. CHEM EDUC RES PRACT CRYSTENGCOMM Dalton Trans. Energy Environ. Sci. ENVIRON SCI-NANO ENVIRON SCI-PROC IMP ENVIRON SCI-WAT RES Faraday Discuss. Food Funct. Green Chem. Inorg. Chem. Front. Integr. Biol. J. Anal. At. Spectrom. J. Mater. Chem. A J. Mater. Chem. B J. Mater. Chem. C Lab Chip Mater. Chem. Front. Mater. Horiz. MEDCHEMCOMM Metallomics Mol. Biosyst. Mol. Syst. Des. Eng. Nanoscale Nanoscale Horiz. Nat. Prod. Rep. New J. Chem. Org. Biomol. Chem. Org. Chem. Front. PHOTOCH PHOTOBIO SCI PCCP Polym. Chem.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1