{"title":"Josephon Array Voltage Calibration System: Operational Use and Verification","authors":"R. Steiner, B. Field","doi":"10.1109/CPEM.1988.671140","DOIUrl":null,"url":null,"abstract":"A New Josephson Array System Now Maintains The U. S. Legal Volt On A Weekly Basis. The Accuracy Of The I-volt Level Of This System Was Verified To Within 0.03 Ppm In Multiple Tests Against Two Independent Resistive Dividers. The Precision In The Verification Was -limited By Uncertainties In The Divider Instruments.","PeriodicalId":326579,"journal":{"name":"1988 Conference on Precision Electromagnetic Measurements","volume":"42 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"1988-06-07","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"4","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"1988 Conference on Precision Electromagnetic Measurements","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/CPEM.1988.671140","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 4
Abstract
A New Josephson Array System Now Maintains The U. S. Legal Volt On A Weekly Basis. The Accuracy Of The I-volt Level Of This System Was Verified To Within 0.03 Ppm In Multiple Tests Against Two Independent Resistive Dividers. The Precision In The Verification Was -limited By Uncertainties In The Divider Instruments.