Characterisation Covid-19 IgG Antibody Kinetics Post Infection and Vaccination using the Boditech iCHROMA™ Fluorescence Immunoassay (FIA) Method

Bolodeoku J
{"title":"Characterisation Covid-19 IgG Antibody Kinetics Post Infection and Vaccination using the Boditech iCHROMA™ Fluorescence Immunoassay (FIA) Method","authors":"Bolodeoku J","doi":"10.23880/aii-16000163","DOIUrl":null,"url":null,"abstract":"Background: Coronavirus disease 2019 (COVID-19) is an emerging threat affecting millions of people worldwide. This study aims to assess the serologic profiles and time kinetics of antibodies against severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) in patients with COVID-19 using the Boditech iCHROMA fluorescence immunoassay (FIA) method. Methods: Samples were collected from 43 subjects and analyzed for IgG antibodies against SARS-CoV-2 using the Boditech iCHROMA FIA. The iCHROMA processes the signal using a cut off index of 0.9 – 1.1, results <0.9 are interpreted as negative, results between 0.9 and 1.1 are interpreted as indeterminate and results >1.1 are interpreted as positive. Longitudinal samples were collected on days between day 1- day 100 post onset of symptoms in 1 subject, single samples collected on days 40-90 post infection in 30 subjects, and single samples collected at 3 time points (pre vaccination, 8-18 days post vaccination and 20-28 days post vaccination) collected from 12 subjects non-infected and previously infected. Results: The Covid-19 IgG concentration was detected (positive and cut off index above 1.1) on day 11 and remained consistently elevated for another 100 days of measurement in the sera of the single infected subject. The IgG cut off indices between day 20 and 100 ranged between 26.8 and 46.7, with a mean of 36.5. In the sera of the 30 patients infected with Covid-19 collected 40 – 90 days post infection, IgG antibodies were detected in all the samples. The IgG cut off indices ranged between 14.0 and 32.60, with a mean of 20.65. Of the twelve subjects who were vaccinated, eight had not been previously infected, while 4 had been previously infected. All 4 (100%) previously infected subjects seroconverted by the first time point (8-18 days post vaccination), whilst only 1/8 (12.5%) of the non-infected subjects seroconverted by the first time point (8-18 days post vaccination). By the second time point (20 - 28 days post vaccination), 2/12 subjects (17%) had failed to seroconvert, whilst 10/12 (83%) of the subjects had seroconverted. In addition, a previously infected subject whose IgG antibodies had dropped to a very low-level, after receiving the first dose of vaccination his IgG antibodies had gone back to his previous antibody levels at the first time point (8-18 days post vaccination). Conclusion: The Boditech iCHROMA FIA method for antibody testing is useful in detecting SARS-CoV-2 in a variety of situations at different time points but has its limitations with regards to determining the difference between previously infected and vaccinated individuals.","PeriodicalId":409855,"journal":{"name":"Annals of Immunology &amp; Immunotherapy","volume":"29 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"1900-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Annals of Immunology &amp; Immunotherapy","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.23880/aii-16000163","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Background: Coronavirus disease 2019 (COVID-19) is an emerging threat affecting millions of people worldwide. This study aims to assess the serologic profiles and time kinetics of antibodies against severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) in patients with COVID-19 using the Boditech iCHROMA fluorescence immunoassay (FIA) method. Methods: Samples were collected from 43 subjects and analyzed for IgG antibodies against SARS-CoV-2 using the Boditech iCHROMA FIA. The iCHROMA processes the signal using a cut off index of 0.9 – 1.1, results <0.9 are interpreted as negative, results between 0.9 and 1.1 are interpreted as indeterminate and results >1.1 are interpreted as positive. Longitudinal samples were collected on days between day 1- day 100 post onset of symptoms in 1 subject, single samples collected on days 40-90 post infection in 30 subjects, and single samples collected at 3 time points (pre vaccination, 8-18 days post vaccination and 20-28 days post vaccination) collected from 12 subjects non-infected and previously infected. Results: The Covid-19 IgG concentration was detected (positive and cut off index above 1.1) on day 11 and remained consistently elevated for another 100 days of measurement in the sera of the single infected subject. The IgG cut off indices between day 20 and 100 ranged between 26.8 and 46.7, with a mean of 36.5. In the sera of the 30 patients infected with Covid-19 collected 40 – 90 days post infection, IgG antibodies were detected in all the samples. The IgG cut off indices ranged between 14.0 and 32.60, with a mean of 20.65. Of the twelve subjects who were vaccinated, eight had not been previously infected, while 4 had been previously infected. All 4 (100%) previously infected subjects seroconverted by the first time point (8-18 days post vaccination), whilst only 1/8 (12.5%) of the non-infected subjects seroconverted by the first time point (8-18 days post vaccination). By the second time point (20 - 28 days post vaccination), 2/12 subjects (17%) had failed to seroconvert, whilst 10/12 (83%) of the subjects had seroconverted. In addition, a previously infected subject whose IgG antibodies had dropped to a very low-level, after receiving the first dose of vaccination his IgG antibodies had gone back to his previous antibody levels at the first time point (8-18 days post vaccination). Conclusion: The Boditech iCHROMA FIA method for antibody testing is useful in detecting SARS-CoV-2 in a variety of situations at different time points but has its limitations with regards to determining the difference between previously infected and vaccinated individuals.
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
使用Boditech iCHROMA™荧光免疫分析(FIA)方法表征感染和接种后Covid-19 IgG抗体动力学
背景:2019冠状病毒病(COVID-19)是影响全球数百万人的新威胁。本研究旨在利用Boditech iCHROMA荧光免疫分析法(FIA)评估COVID-19患者抗严重急性呼吸综合征冠状病毒2 (SARS-CoV-2)抗体的血清学特征和时间动力学。方法:采集43例样本,采用Boditech iCHROMA FIA检测SARS-CoV-2 IgG抗体。iCHROMA使用截断指数0.9 - 1.1处理信号,结果1.1被解释为正。纵向样本采集于1名受试者出现症状后第1天至第100天之间,30名受试者在感染后第40-90天采集单样本,12名未感染和既往感染受试者在接种前、接种后8-18天和接种后20-28天采集单样本。结果:在第11天检测到单个感染者血清中Covid-19 IgG浓度(阳性且切断指数大于1.1),并在随后的100天测量中持续升高。第20 ~ 100天IgG切断指数为26.8 ~ 46.7,平均值为36.5。在感染后40 ~ 90天采集的30例新冠肺炎患者血清中均检测到IgG抗体。IgG切断指数范围为14.0 ~ 32.60,平均值为20.65。在接种疫苗的12名受试者中,有8人以前没有感染过,而4人以前感染过。所有4名(100%)先前感染的受试者在第一个时间点(接种疫苗后8-18天)血清转化,而未感染的受试者中只有1/8(12.5%)在第一个时间点(接种疫苗后8-18天)血清转化。到第二个时间点(接种后20 - 28天),2/12(17%)的受试者血清转化失败,10/12(83%)的受试者血清转化成功。此外,先前感染的受试者,其IgG抗体已降至极低水平,在接受第一次疫苗接种后,其IgG抗体已恢复到第一个时间点(疫苗接种后8-18天)的抗体水平。结论:Boditech iCHROMA FIA抗体检测方法可用于多种情况下不同时间点的SARS-CoV-2检测,但在确定既往感染者和接种者之间的差异方面存在局限性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
COVID-19 and Long Covid as Complex, Multi-Organs Involvement and Multi-System Disease Immunological Disorders in Adult Patients with Trichophytia and Microsporia of the Genital Area, Tajikistan How is the Mechanism of Immunodeficiency Caused by Metallic Implants and What Can they Cause when Inside the Patient's Body? Co-occurrence of Rheumatoid Arthritis and Pulmonary Tuberculosis: A Diagnostic and Therapeutic Challenge The Passing Pandemic? Perspective and Projections for COVID-19
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1