Accuracy of two osmometers on standard samples: electrical impedance technique and freezing point depression technique

C. García-Resúa, Hugo Pena-Verdeal, M. Miñones, Jorge Gilino, M. .. Giraldez, E. Yebra-Pimentel
{"title":"Accuracy of two osmometers on standard samples: electrical impedance technique and freezing point depression technique","authors":"C. García-Resúa, Hugo Pena-Verdeal, M. Miñones, Jorge Gilino, M. .. Giraldez, E. Yebra-Pimentel","doi":"10.1117/12.2030428","DOIUrl":null,"url":null,"abstract":"High tear fluid osmolarity is a feature common to all types of dry eye. This study was designed to establish the accuracy of two osmometers, a freezing point depression osmometer (Fiske 110) and an electrical impedance osmometer (TearLab™) by using standard samples. To assess the accuracy of the measurements provided by the two instruments we used 5 solutions of known osmolarity/osmolality; 50, 290 and 850 mOsm/kg and 292 and 338 mOsm/L. Fiske 110 is designed to be used in samples of 20 μl, so measurements were made on 1:9, 1:4, 1:1 and 1:0 dilutions of the standards. Tear Lab is addressed to be used in tear film and only a sample of 0.05 μl is required, so no dilutions were employed. Due to the smaller measurement range of the TearLab, the 50 and 850 mOsm/kg standards were not included. 20 measurements per standard sample were used and differences with the reference value was analysed by one sample t-test. Fiske 110 showed that osmolarity measurements differed statistically from standard values except those recorded for 290 mOsm/kg standard diluted 1:1 (p = 0.309), the 292 mOsm/L H2O sample (1:1) and 338 mOsm/L H2O standard (1:4). The more diluted the sample, the higher the error rate. For the TearLab measurements, one-sample t-test indicated that all determinations differed from the theoretical values (p = 0.001), though differences were always small. For undiluted solutions, Fiske 110 shows similar performance than TearLab. However, for the diluted standards, Fiske 110 worsens.","PeriodicalId":135913,"journal":{"name":"Iberoamerican Meeting of Optics and the Latin American Meeting of Optics, Lasers and Their Applications","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2013-11-18","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"2","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Iberoamerican Meeting of Optics and the Latin American Meeting of Optics, Lasers and Their Applications","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1117/12.2030428","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 2

Abstract

High tear fluid osmolarity is a feature common to all types of dry eye. This study was designed to establish the accuracy of two osmometers, a freezing point depression osmometer (Fiske 110) and an electrical impedance osmometer (TearLab™) by using standard samples. To assess the accuracy of the measurements provided by the two instruments we used 5 solutions of known osmolarity/osmolality; 50, 290 and 850 mOsm/kg and 292 and 338 mOsm/L. Fiske 110 is designed to be used in samples of 20 μl, so measurements were made on 1:9, 1:4, 1:1 and 1:0 dilutions of the standards. Tear Lab is addressed to be used in tear film and only a sample of 0.05 μl is required, so no dilutions were employed. Due to the smaller measurement range of the TearLab, the 50 and 850 mOsm/kg standards were not included. 20 measurements per standard sample were used and differences with the reference value was analysed by one sample t-test. Fiske 110 showed that osmolarity measurements differed statistically from standard values except those recorded for 290 mOsm/kg standard diluted 1:1 (p = 0.309), the 292 mOsm/L H2O sample (1:1) and 338 mOsm/L H2O standard (1:4). The more diluted the sample, the higher the error rate. For the TearLab measurements, one-sample t-test indicated that all determinations differed from the theoretical values (p = 0.001), though differences were always small. For undiluted solutions, Fiske 110 shows similar performance than TearLab. However, for the diluted standards, Fiske 110 worsens.
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
两种渗透仪在标准样品上的精度:电阻抗技术和凝固点降低技术
高泪液渗透压是所有类型的干眼症的共同特征。本研究旨在通过使用标准样品建立两种渗透仪的准确性,一种是冰点下降渗透仪(Fiske 110),另一种是电阻抗渗透仪(TearLab™)。为了评估这两种仪器提供的测量结果的准确性,我们使用了5种已知的渗透压/渗透压溶液;50、290、850 mOsm/kg和292、338 mOsm/L。Fiske 110设计用于20 μl的样品,因此在1:9,1:4,1:1和1:0的稀释度下进行测量。泪液实验室被指定用于泪膜,只需要0.05 μl的样品,所以没有使用稀释。由于TearLab的测量范围较小,因此没有包括50和850 mOsm/kg标准。每个标准样本使用20个测量值,并通过一个样本t检验分析与参考值的差异。Fiske 110表明,除稀释比例为1:1的290 mOsm/kg标准品(p = 0.309)、稀释比例为1:1的292 mOsm/L H2O样品和稀释比例为1:4的338 mOsm/L H2O标准品外,渗透压测量值与标准值有统计学差异。样品越稀释,错误率越高。对于TearLab测量,单样本t检验表明所有测定值都与理论值不同(p = 0.001),尽管差异总是很小。对于未稀释溶液,Fiske 110的性能与TearLab相似。然而,对于稀释后的标准,Fiske 110恶化了。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
Remote access to an interferometric fringes stabilization active system via RENATA Optical design of a Coudé-Train for a stable and efficient simultaneous feeding of the ESPRESSO spectrograph from the four VLT telescopes Configurable multipulsing of a MOPA pulsed fiber laser with applications in materials processing New method for sub-structured Ronchi rulings generation and his irradiance profile Photorefractive moiré-like patterns with different variation directions for multi-projection in profilometer applications
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1