{"title":"Evaluation and Improvement of the Accuracy of TEM Transmission Cells with Internal Field Measurements","authors":"K. Jokela, L. Puranen","doi":"10.1109/EUMA.1983.333312","DOIUrl":null,"url":null,"abstract":"Two square TEM transmission cells with side heights 30 cm and 50 cm were designed for the calibration of RF-hazard monitors. The error due to reflections was reduced from ±10 % to ±2 % (input VSWR < 1.2) by lengthening straight sections and measuring the longitudinal E-field patterns at calibration frequencies. Comparison measurements with small probes showed that the maximum difference between the cells was 7 % except some sharp resonance points caused by TEM harmonics and TE01 mode. The vertical field non-uniformity was also examined.","PeriodicalId":105436,"journal":{"name":"1983 13th European Microwave Conference","volume":"70 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"1983-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"3","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"1983 13th European Microwave Conference","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/EUMA.1983.333312","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 3
Abstract
Two square TEM transmission cells with side heights 30 cm and 50 cm were designed for the calibration of RF-hazard monitors. The error due to reflections was reduced from ±10 % to ±2 % (input VSWR < 1.2) by lengthening straight sections and measuring the longitudinal E-field patterns at calibration frequencies. Comparison measurements with small probes showed that the maximum difference between the cells was 7 % except some sharp resonance points caused by TEM harmonics and TE01 mode. The vertical field non-uniformity was also examined.