Evaluation of long-term robustness of the International System of Reference (SIR) used in radionuclide metrology based on the meta-analysis of its machine-readable database
Romain Maximilien Coulon, Carine Michotte, Vincent Gressier
{"title":"Evaluation of long-term robustness of the International System of Reference (SIR) used in radionuclide metrology based on the meta-analysis of its machine-readable database","authors":"Romain Maximilien Coulon, Carine Michotte, Vincent Gressier","doi":"10.1088/1681-7575/ad0a36","DOIUrl":null,"url":null,"abstract":"Abstract In support of the CIPM MRA, the reference system in radionuclide metrology (SIR) is a unique experimental facility that has been able to provide continuous key comparison results for 72 radionuclides for 46 years. These measurement data are now available in a machine-readable format, which allows statistical analysis of these measurement results to be performed efficiently. Such a meta-analysis was conducted to verify whether or not the accumulated data confirmed the bold assumption made in the 1970s of the long-term robustness of the system over decades of operation. It was notably assumed that variation in the properties of the standard solution to be measured or the choice of the 226 Ra reference source would have no impact on the measurement.&#xD;For the majority of radionuclides measured by the SIR, the non-parametric significance tests of the null hypothesis carried out in this study gave no evidence, at the 95 % confidence level, that the above statement should be reconsidered. However, this meta-analysis revealed a significant impact of solution density in the case of 109 Cd, corroborating previous observations made in the case of 241 Am where the robustness of an ionization chamber with respect to the density variation of the solution does not appear to be perfect.&#xD;The extension of the SIR by a liquid scintillation system is under development. It will improve the robustness of international comparison results for those borderline cases where radionuclides emit only low-energy x-rays or gamma rays (< 100 keV).","PeriodicalId":18444,"journal":{"name":"Metrologia","volume":"8 12","pages":"0"},"PeriodicalIF":2.1000,"publicationDate":"2023-11-07","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Metrologia","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1088/1681-7575/ad0a36","RegionNum":3,"RegionCategory":"工程技术","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"INSTRUMENTS & INSTRUMENTATION","Score":null,"Total":0}
引用次数: 0
Abstract
Abstract In support of the CIPM MRA, the reference system in radionuclide metrology (SIR) is a unique experimental facility that has been able to provide continuous key comparison results for 72 radionuclides for 46 years. These measurement data are now available in a machine-readable format, which allows statistical analysis of these measurement results to be performed efficiently. Such a meta-analysis was conducted to verify whether or not the accumulated data confirmed the bold assumption made in the 1970s of the long-term robustness of the system over decades of operation. It was notably assumed that variation in the properties of the standard solution to be measured or the choice of the 226 Ra reference source would have no impact on the measurement.
For the majority of radionuclides measured by the SIR, the non-parametric significance tests of the null hypothesis carried out in this study gave no evidence, at the 95 % confidence level, that the above statement should be reconsidered. However, this meta-analysis revealed a significant impact of solution density in the case of 109 Cd, corroborating previous observations made in the case of 241 Am where the robustness of an ionization chamber with respect to the density variation of the solution does not appear to be perfect.
The extension of the SIR by a liquid scintillation system is under development. It will improve the robustness of international comparison results for those borderline cases where radionuclides emit only low-energy x-rays or gamma rays (< 100 keV).
期刊介绍:
Published 6 times per year, Metrologia covers the fundamentals of measurements, particularly those dealing with the seven base units of the International System of Units (metre, kilogram, second, ampere, kelvin, candela, mole) or proposals to replace them.
The journal also publishes papers that contribute to the solution of difficult measurement problems and improve the accuracy of derived units and constants that are of fundamental importance to physics.
In addition to regular papers, the journal publishes review articles, issues devoted to single topics of timely interest and occasional conference proceedings. Letters to the Editor and Short Communications (generally three pages or less) are also considered.