{"title":"A digital four-arm bridge for the comparison of resistance with capacitance","authors":"Yicheng Wang and Stephan Schlamminger","doi":"10.1088/1681-7575/ad7590","DOIUrl":null,"url":null,"abstract":"We have built and demonstrated a digital four-arm bridge for the comparison of resistance with capacitance. The digital four-arm bridge mimics the classical quad bridge in the digital domain with three balances: the source balance, the detector balance, as well as the main balance. Due to correlation, the required precision of the source voltages is only of the order of the square root of the ultimate bridge precision. For the comparison of a 100 kΩ resistor with a 1 nF capacitor near 1592 Hz, the combined standard uncertainty is .","PeriodicalId":18444,"journal":{"name":"Metrologia","volume":"25 1","pages":""},"PeriodicalIF":2.1000,"publicationDate":"2024-09-10","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Metrologia","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1088/1681-7575/ad7590","RegionNum":3,"RegionCategory":"工程技术","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"INSTRUMENTS & INSTRUMENTATION","Score":null,"Total":0}
引用次数: 0
Abstract
We have built and demonstrated a digital four-arm bridge for the comparison of resistance with capacitance. The digital four-arm bridge mimics the classical quad bridge in the digital domain with three balances: the source balance, the detector balance, as well as the main balance. Due to correlation, the required precision of the source voltages is only of the order of the square root of the ultimate bridge precision. For the comparison of a 100 kΩ resistor with a 1 nF capacitor near 1592 Hz, the combined standard uncertainty is .
期刊介绍:
Published 6 times per year, Metrologia covers the fundamentals of measurements, particularly those dealing with the seven base units of the International System of Units (metre, kilogram, second, ampere, kelvin, candela, mole) or proposals to replace them.
The journal also publishes papers that contribute to the solution of difficult measurement problems and improve the accuracy of derived units and constants that are of fundamental importance to physics.
In addition to regular papers, the journal publishes review articles, issues devoted to single topics of timely interest and occasional conference proceedings. Letters to the Editor and Short Communications (generally three pages or less) are also considered.