Effect of exogenous lipids contamination on blood gas analysis

G. Lippi, L. Pighi, G. Salvagno, Elena Tiziani, Maria Elena Castellini, Roberta Ferraro, Brandon M. Henry
{"title":"Effect of exogenous lipids contamination on blood gas analysis","authors":"G. Lippi, L. Pighi, G. Salvagno, Elena Tiziani, Maria Elena Castellini, Roberta Ferraro, Brandon M. Henry","doi":"10.1515/almed-2024-0043","DOIUrl":null,"url":null,"abstract":"\n \n \n The purpose of this study was to investigate the effects of contamination of venous blood with a lipid-containing solution on parameters measured by a modern blood gas analyzer.\n \n \n \n We collected venous blood from 17 healthcare workers (46 ± 11 years; 53 % women) into three blood gas syringes containing 0 , 5 and 10 % lipid-containing solution. Blood gas analysis was performed within 15 min from sample collection on GEM Premier 5000, while triglycerides and serum indices were assays on Roche COBAS C702.\n \n \n \n Triglycerides concentration increased from 1.0 ± 0.3 mmol/L in the uncontaminated blood gas syringe, to 39.4 ± 7.8 and 65.3 ± 14.4 mmol/L (both p<0.001) in syringes with 5 and 10 % final lipid contamination. The lipemic and hemolysis indices increased accordingly. Statistically significant variation was noted for all analytes except hematocrit and COHb in the syringe with 5 % lipids, while only COHb did not vary in the syringe with 10 % lipids. Significant increases were observed from 5 % lipid contamination for pO2, SO2 and lactate, while the values of pH, pCO2, sodium, potassium, chloride, ionized calcium, glucose, hematocrit (10 % contamination), hemoglobin and MetHB decreased. All these changes except lactate and CoHb exceeded their relative performance specifications.\n \n \n \n Artifactual hyperlipidemia caused by contamination with exogenous lipids can have a clinically significant impact on blood gas analysis. Manufacturers of blood gas analyzers must be persuaded to develop new instruments equipped with serum indices.\n","PeriodicalId":502333,"journal":{"name":"Advances in Laboratory Medicine / Avances en Medicina de Laboratorio","volume":"102 6","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2024-05-07","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Advances in Laboratory Medicine / Avances en Medicina de Laboratorio","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1515/almed-2024-0043","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

The purpose of this study was to investigate the effects of contamination of venous blood with a lipid-containing solution on parameters measured by a modern blood gas analyzer. We collected venous blood from 17 healthcare workers (46 ± 11 years; 53 % women) into three blood gas syringes containing 0 , 5 and 10 % lipid-containing solution. Blood gas analysis was performed within 15 min from sample collection on GEM Premier 5000, while triglycerides and serum indices were assays on Roche COBAS C702. Triglycerides concentration increased from 1.0 ± 0.3 mmol/L in the uncontaminated blood gas syringe, to 39.4 ± 7.8 and 65.3 ± 14.4 mmol/L (both p<0.001) in syringes with 5 and 10 % final lipid contamination. The lipemic and hemolysis indices increased accordingly. Statistically significant variation was noted for all analytes except hematocrit and COHb in the syringe with 5 % lipids, while only COHb did not vary in the syringe with 10 % lipids. Significant increases were observed from 5 % lipid contamination for pO2, SO2 and lactate, while the values of pH, pCO2, sodium, potassium, chloride, ionized calcium, glucose, hematocrit (10 % contamination), hemoglobin and MetHB decreased. All these changes except lactate and CoHb exceeded their relative performance specifications. Artifactual hyperlipidemia caused by contamination with exogenous lipids can have a clinically significant impact on blood gas analysis. Manufacturers of blood gas analyzers must be persuaded to develop new instruments equipped with serum indices.
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
外源性脂质污染对血气分析的影响
本研究旨在探讨静脉血被含脂溶液污染对现代血气分析仪测量参数的影响。 我们采集了 17 名医护人员(46 ± 11 岁;53 % 为女性)的静脉血,分别注入含有 0 %、5 % 和 10 % 含脂溶液的三个血气分析注射器中。血气分析在样本采集后 15 分钟内通过 GEM Premier 5000 进行,甘油三酯和血清指数则通过罗氏 COBAS C702 进行检测。 甘油三酯浓度从未受损伤的血气注射器中的 1.0 ± 0.3 mmol/L,上升到最终脂质污染为 5% 和 10% 的注射器中的 39.4 ± 7.8 mmol/L 和 65.3 ± 14.4 mmol/L(均 p<0.001)。血脂和溶血指数也相应增加。在脂质含量为 5% 的注射器中,除血细胞比容和 COHb 外,所有分析物均有明显的统计学变化,而在脂质含量为 10% 的注射器中,只有 COHb 没有变化。从 5% 的脂质污染开始,pO2、SO2 和乳酸盐的值显著增加,而 pH、pCO2、钠、钾、氯、离子钙、葡萄糖、血细胞比容(10% 污染)、血红蛋白和甲氧基肾上腺素的值则下降。除乳酸和 CoHb 外,所有这些变化都超过了其相对性能指标。 外源性脂质污染造成的人为高脂血症会对血气分析产生重大临床影响。必须说服血气分析仪制造商开发配备血清指数的新仪器。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
Recuperación de espermatozoides de la orina en hombres con eyaculación retrógrada Investigating the incremental value of urine sediment reporting in emergency medicine with a Sysmex UN urinalysis system Surgimiento y evolución de los sistemas de normalización para laboratorios clínicos Interference by vitamin B12-macrocomplexes: towards an effective detection and correct interpretation of hypo- and hypervitaminemia Mass spectrometry in clinical protein laboratories
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1