In vitro skin puncture methodology for material characterization

IF 1.7 4区 医学 Q3 ENGINEERING, BIOMEDICAL Medical Engineering & Physics Pub Date : 2024-06-26 DOI:10.1016/j.medengphy.2024.104199
Joseph LeSueur , Carolyn Hampton , Michael Kleinberger , William Dzwierzynski , Frank A. Pintar
{"title":"In vitro skin puncture methodology for material characterization","authors":"Joseph LeSueur ,&nbsp;Carolyn Hampton ,&nbsp;Michael Kleinberger ,&nbsp;William Dzwierzynski ,&nbsp;Frank A. Pintar","doi":"10.1016/j.medengphy.2024.104199","DOIUrl":null,"url":null,"abstract":"<div><p>Quantifying the mechanical behavior of skin has been foundational in applications of cosmetics, surgical techniques, forensic science, and protective clothing development. However, previous puncture studies have lacked consistent and physiological boundary conditions of skin. To determine natural skin tension, excision of in situ porcine skin resulted in significantly different diameter reduction (shrinkage) in leg (19.5 %) and abdominal skin (38.4 %) compared to flank skin (28.5 %) (<em>p</em> = 0.047). To examine effects of initial tension and pre-conditioning, five conditions of initial tension (as percentage of diameter increase) and pre-conditioning were tested in quasistatic puncture with a 5 mm spherical impactor using an electrohydraulic load frame and custom clamping apparatus. Samples with less than 5 % initial tension resulted in significantly greater (<em>p</em> = 0.011) force at failure (279.2 N) compared to samples with greater than 25 % initial tension (195.1 N). Eight pre-conditioning cycles of 15 mm displacement reduced hysteresis by 45 %. The coefficient of variance was substantially reduced for force, force normalized by cutis thickness, displacement, stiffness, and strain energy up to 46 %. Pre-conditioned samples at physiological initial tension (14–25 %) resulted in significantly greater (<em>p</em> = 0.03) normalized forces at failure (278.3 N/mm) compared to non-conditioned samples of the same initial tension (234.4 N/mm). Pre-conditioned samples with 14–25 % initial tension, representing physiological boundary conditions, resulted in the most appropriate failure thresholds with the least variation. For in vitro puncture studies, the magnitude of applied initial tension should be defined based on anatomical location, through a shrinkage experimentation, to match natural tension of skin. Characterizing the biological behavior and tolerances of skin may be utilized in finite element models to aid in protective clothing development and forensic science analyses.</p></div>","PeriodicalId":49836,"journal":{"name":"Medical Engineering & Physics","volume":"130 ","pages":"Article 104199"},"PeriodicalIF":1.7000,"publicationDate":"2024-06-26","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Medical Engineering & Physics","FirstCategoryId":"5","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S1350453324001000","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"ENGINEERING, BIOMEDICAL","Score":null,"Total":0}
引用次数: 0

Abstract

Quantifying the mechanical behavior of skin has been foundational in applications of cosmetics, surgical techniques, forensic science, and protective clothing development. However, previous puncture studies have lacked consistent and physiological boundary conditions of skin. To determine natural skin tension, excision of in situ porcine skin resulted in significantly different diameter reduction (shrinkage) in leg (19.5 %) and abdominal skin (38.4 %) compared to flank skin (28.5 %) (p = 0.047). To examine effects of initial tension and pre-conditioning, five conditions of initial tension (as percentage of diameter increase) and pre-conditioning were tested in quasistatic puncture with a 5 mm spherical impactor using an electrohydraulic load frame and custom clamping apparatus. Samples with less than 5 % initial tension resulted in significantly greater (p = 0.011) force at failure (279.2 N) compared to samples with greater than 25 % initial tension (195.1 N). Eight pre-conditioning cycles of 15 mm displacement reduced hysteresis by 45 %. The coefficient of variance was substantially reduced for force, force normalized by cutis thickness, displacement, stiffness, and strain energy up to 46 %. Pre-conditioned samples at physiological initial tension (14–25 %) resulted in significantly greater (p = 0.03) normalized forces at failure (278.3 N/mm) compared to non-conditioned samples of the same initial tension (234.4 N/mm). Pre-conditioned samples with 14–25 % initial tension, representing physiological boundary conditions, resulted in the most appropriate failure thresholds with the least variation. For in vitro puncture studies, the magnitude of applied initial tension should be defined based on anatomical location, through a shrinkage experimentation, to match natural tension of skin. Characterizing the biological behavior and tolerances of skin may be utilized in finite element models to aid in protective clothing development and forensic science analyses.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
表征材料特性的体外皮肤穿刺方法
量化皮肤的机械行为对化妆品、外科技术、法医学和防护服开发等应用具有重要意义。然而,以往的穿刺研究缺乏一致的皮肤生理边界条件。为了确定皮肤的自然张力,原位切除猪皮导致腿部皮肤(19.5%)和腹部皮肤(38.4%)的直径缩小(收缩)与侧腹皮肤(28.5%)相比有显著差异(p = 0.047)。为了研究初始张力和预处理的影响,使用电动液压负载框架和定制夹具,在 5 毫米球形冲击器的准静态穿刺中测试了五种初始张力(直径增加百分比)和预处理条件。与初始张力大于 25% 的样品(195.1 N)相比,初始张力小于 5% 的样品的破坏力(279.2 N)明显更大(p = 0.011)。15 毫米位移的八个预调周期将滞后减少了 45%。力、按切口厚度归一化的力、位移、刚度和应变能的方差系数大幅降低了 46%。与相同初始张力(234.4 牛顿/毫米)的非预处理样品相比,生理初始张力(14-25%)的预处理样品在破坏时的归一化力(278.3 牛顿/毫米)明显更大(p = 0.03)。预调节样品的初始张力为 14-25%,代表了生理边界条件,结果是最合适的破坏阈值,变化最小。在体外穿刺研究中,应根据解剖位置,通过收缩实验来确定施加的初始张力大小,以符合皮肤的自然张力。表征皮肤的生物行为和容差可用于有限元模型,以帮助防护服开发和法医学分析。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
Medical Engineering & Physics
Medical Engineering & Physics 工程技术-工程:生物医学
CiteScore
4.30
自引率
4.50%
发文量
172
审稿时长
3.0 months
期刊介绍: Medical Engineering & Physics provides a forum for the publication of the latest developments in biomedical engineering, and reflects the essential multidisciplinary nature of the subject. The journal publishes in-depth critical reviews, scientific papers and technical notes. Our focus encompasses the application of the basic principles of physics and engineering to the development of medical devices and technology, with the ultimate aim of producing improvements in the quality of health care.Topics covered include biomechanics, biomaterials, mechanobiology, rehabilitation engineering, biomedical signal processing and medical device development. Medical Engineering & Physics aims to keep both engineers and clinicians abreast of the latest applications of technology to health care.
期刊最新文献
A patient-matched prosthesis for thumb amputations: Design, mechanical and functional evaluation Influence of surface type on outdoor gait parameters measured using an In-Shoe Motion Sensor System Assessment of pre- and post-operative gait dynamics in total knee arthroplasty by a wearable capture system A novel 3D lightweight model for COVID-19 lung CT Lesion Segmentation ResGloTBNet: An interpretable deep residual network with global long-range dependency for tuberculosis screening of sputum smear microscopy images
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1