Optical quality in vitro and in vivo of an extended depth-of-focus intraocular lens with isofocal design.

IF 2.4 3区 医学 Q2 OPHTHALMOLOGY Graefe’s Archive for Clinical and Experimental Ophthalmology Pub Date : 2024-12-01 Epub Date: 2024-07-24 DOI:10.1007/s00417-024-06590-y
Sara Fernández-Núñez, Lidia Pérez-Sanz, Jose Antonio Gómez-Pedrero, María García-Montero, César Albarrán-Diego, Nuria Garzón
{"title":"Optical quality in vitro and in vivo of an extended depth-of-focus intraocular lens with isofocal design.","authors":"Sara Fernández-Núñez, Lidia Pérez-Sanz, Jose Antonio Gómez-Pedrero, María García-Montero, César Albarrán-Diego, Nuria Garzón","doi":"10.1007/s00417-024-06590-y","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>The aim of this study is to compare optical quality results obtained in laboratory analysis (in vitro) versus clinical data (in vivo).</p><p><strong>Methods: </strong>The optical quality of ISOPure intraocular lens was assessed both in vitro and in vivo using the modulation transfer function (MTF) for 3.0 and 4.5 mm pupil diameters. In vitro measurements were obtained using deflectometer NIMO TRF1504, while in vivo measurements were taken with OPD-Scan III in a set of patients implanted with this lens. Ray tracing techniques were used to determine the MTF and area under MTF curve (MTFa) from the measured wavefront for the isolated lens and for the whole eye.</p><p><strong>Results: </strong>The MTF of the isolated lens obtained under both in vitro and in vivo conditions showed comparable results for both pupil sizes. However, differences were found when comparing the MTF of the whole eye with the lens implanted versus the MTF measured in vitro for 4.5 mm pupil size. Also, the MTFa defocus curve was compared with the defocus curve measured in vivo.</p><p><strong>Conclusion: </strong>The defocus curve from the in vivo study aligns closely with the MTFa of the in vitro model, with a useful defocus range of 0.40D. Thus, it is possible to anticipate the visual results of the implanted isofocal lens by using measurements on an optical bench and conducting optical simulations.</p>","PeriodicalId":12795,"journal":{"name":"Graefe’s Archive for Clinical and Experimental Ophthalmology","volume":" ","pages":"3905-3913"},"PeriodicalIF":2.4000,"publicationDate":"2024-12-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Graefe’s Archive for Clinical and Experimental Ophthalmology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1007/s00417-024-06590-y","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/7/24 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: The aim of this study is to compare optical quality results obtained in laboratory analysis (in vitro) versus clinical data (in vivo).

Methods: The optical quality of ISOPure intraocular lens was assessed both in vitro and in vivo using the modulation transfer function (MTF) for 3.0 and 4.5 mm pupil diameters. In vitro measurements were obtained using deflectometer NIMO TRF1504, while in vivo measurements were taken with OPD-Scan III in a set of patients implanted with this lens. Ray tracing techniques were used to determine the MTF and area under MTF curve (MTFa) from the measured wavefront for the isolated lens and for the whole eye.

Results: The MTF of the isolated lens obtained under both in vitro and in vivo conditions showed comparable results for both pupil sizes. However, differences were found when comparing the MTF of the whole eye with the lens implanted versus the MTF measured in vitro for 4.5 mm pupil size. Also, the MTFa defocus curve was compared with the defocus curve measured in vivo.

Conclusion: The defocus curve from the in vivo study aligns closely with the MTFa of the in vitro model, with a useful defocus range of 0.40D. Thus, it is possible to anticipate the visual results of the implanted isofocal lens by using measurements on an optical bench and conducting optical simulations.

Abstract Image

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
采用等焦设计的延焦深度眼内透镜的体外和体内光学质量。
目的:本研究旨在比较实验室分析(体外)与临床数据(体内)得出的光学质量结果:方法:使用调制传递函数(MTF)对 3.0 毫米和 4.5 毫米瞳孔直径的 ISOPure 眼内透镜的光学质量进行了体外和体内评估。体外测量使用偏转仪 NIMO TRF1504,体内测量使用 OPD-Scan III,测量对象为植入该镜片的一组患者。采用光线跟踪技术,根据测量的波面确定离体晶状体和整个眼球的 MTF 和 MTF 曲线下面积(MTFa):在体外和体内条件下获得的离体晶状体的 MTF 对两种瞳孔大小都显示出相似的结果。然而,在 4.5 毫米瞳孔大小的情况下,将植入晶状体的全眼 MTF 与体外测量的 MTF 进行比较时,发现两者存在差异。此外,还将 MTFa 散焦曲线与体外测量的散焦曲线进行了比较:结论:体内研究的散焦曲线与体外模型的 MTFa 非常吻合,有用的散焦范围为 0.40D。因此,通过在光学工作台上进行测量和光学模拟,可以预测植入式等焦距透镜的视觉效果。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
CiteScore
5.40
自引率
7.40%
发文量
398
审稿时长
3 months
期刊介绍: Graefe''s Archive for Clinical and Experimental Ophthalmology is a distinguished international journal that presents original clinical reports and clini-cally relevant experimental studies. Founded in 1854 by Albrecht von Graefe to serve as a source of useful clinical information and a stimulus for discussion, the journal has published articles by leading ophthalmologists and vision research scientists for more than a century. With peer review by an international Editorial Board and prompt English-language publication, Graefe''s Archive provides rapid dissemination of clinical and clinically related experimental information.
期刊最新文献
Emerging clinical evidence of a dual role for Ang-2 and VEGF-A blockade with faricimab in retinal diseases. Evaluating anterior segment stability and corneal endothelium after prolene gonioscopy assisted transluminal trabeculotomy (GATT) in open-angle glaucoma. Effects of ambient atmospheric pressure on intraocular pressure measured using a Goldman applanation tonometer in normal eyes under ordinary conditions. Enhancing diabetic retinopathy and macular edema detection through multi scale feature fusion using deep learning model. Primary angle-closed diseases recognition through artificial intelligence-based anterior segment-optical coherence tomography imaging.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1