Comparison of the effects of laser systems and cold atmospheric plasma on the surface roughness and shear bond strength of flexible hybrid ceramics.

IF 2.1 4区 医学 Q3 ENGINEERING, BIOMEDICAL Lasers in Medical Science Pub Date : 2024-11-13 DOI:10.1007/s10103-024-04238-1
Seda Üstün Aladağ, Elif Aydoğan Ayaz
{"title":"Comparison of the effects of laser systems and cold atmospheric plasma on the surface roughness and shear bond strength of flexible hybrid ceramics.","authors":"Seda Üstün Aladağ, Elif Aydoğan Ayaz","doi":"10.1007/s10103-024-04238-1","DOIUrl":null,"url":null,"abstract":"<p><p>The aim of this study was to compare the effects of laser systems and cold atmospheric plasma (CAP) on the surface roughness (Ra) and shear bond strength (SBS) of flexible hybrid ceramics (FHCs). Eighty FHC samples were divided into 5 groups to be subjected to surface treatments (hydrofluoric acid (HFA), HFA + 5 W Er: YAG laser (HFA + 5WE), HFA + 3 W Er: YAG laser (HFA + 3WE), HFA + ultrafast fiber laser (HFA + FL), and CAP). The Er: YAG laser (AT Fidelis Plus III, Fotona, Slovenia) was operated with a 0.9 mm diameter tip, delivering 250 mJ and 150 mJ per pulse, with output powers of 5 W and 3 W, and fluences of 23,623 J/cm² and 14,157 J/cm², respectively, at a frequency of 20 Hz and a pulse duration of 600 µs for 30 s. The FL (FiberLAST Inc., Turkey) was applied with a 7 mm spot size, 1 mJ pulse energy, 20 W output power, 100 kHz repetition rate, ultrashort pulse length (100 ns), and the fluence of 1,820 J/cm² for 7 s. The CAP (PiezoBrush PZ3, Relyon Plasma, Germany) applied to the surfaces for 80 s at a treatment speed of 5 cm<sup>2</sup>/s and 100% plasma power. After the Ra values were measured, each group was divided into 2 subgroups to be cemented with total-etch (TE) and self-adhesive (SA) resin cements. The SBS of all the samples was measured. One sample was randomly selected from each group, and the fractured surfaces were examined by SEM analysis. The significance level was at P < 0.05. HFA + FL had the highest Ra values, and there was no significant difference between HFA + FL and HFA + 5WE (P > 0.05). There was no difference in Ra between CAP and HFA (P > 0.05). The highest SBS values were observed for the samples cemented with TE after HFA + FL. The difference between the SBS values of the groups cemented with TE after CAP, HFA + 5WE, HFA + 3WE, and HFA + FL was not significant (P < 0.05). The use of FL may be a promising method to improve SBS without causing thermal damage when using TE or SA cements. CAP can be recommended as a practical and safe treatment with SA cement because it is suitable for chairside use with a handheld device and does not change the physical properties of the material.</p>","PeriodicalId":17978,"journal":{"name":"Lasers in Medical Science","volume":null,"pages":null},"PeriodicalIF":2.1000,"publicationDate":"2024-11-13","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Lasers in Medical Science","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1007/s10103-024-04238-1","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"ENGINEERING, BIOMEDICAL","Score":null,"Total":0}
引用次数: 0

Abstract

The aim of this study was to compare the effects of laser systems and cold atmospheric plasma (CAP) on the surface roughness (Ra) and shear bond strength (SBS) of flexible hybrid ceramics (FHCs). Eighty FHC samples were divided into 5 groups to be subjected to surface treatments (hydrofluoric acid (HFA), HFA + 5 W Er: YAG laser (HFA + 5WE), HFA + 3 W Er: YAG laser (HFA + 3WE), HFA + ultrafast fiber laser (HFA + FL), and CAP). The Er: YAG laser (AT Fidelis Plus III, Fotona, Slovenia) was operated with a 0.9 mm diameter tip, delivering 250 mJ and 150 mJ per pulse, with output powers of 5 W and 3 W, and fluences of 23,623 J/cm² and 14,157 J/cm², respectively, at a frequency of 20 Hz and a pulse duration of 600 µs for 30 s. The FL (FiberLAST Inc., Turkey) was applied with a 7 mm spot size, 1 mJ pulse energy, 20 W output power, 100 kHz repetition rate, ultrashort pulse length (100 ns), and the fluence of 1,820 J/cm² for 7 s. The CAP (PiezoBrush PZ3, Relyon Plasma, Germany) applied to the surfaces for 80 s at a treatment speed of 5 cm2/s and 100% plasma power. After the Ra values were measured, each group was divided into 2 subgroups to be cemented with total-etch (TE) and self-adhesive (SA) resin cements. The SBS of all the samples was measured. One sample was randomly selected from each group, and the fractured surfaces were examined by SEM analysis. The significance level was at P < 0.05. HFA + FL had the highest Ra values, and there was no significant difference between HFA + FL and HFA + 5WE (P > 0.05). There was no difference in Ra between CAP and HFA (P > 0.05). The highest SBS values were observed for the samples cemented with TE after HFA + FL. The difference between the SBS values of the groups cemented with TE after CAP, HFA + 5WE, HFA + 3WE, and HFA + FL was not significant (P < 0.05). The use of FL may be a promising method to improve SBS without causing thermal damage when using TE or SA cements. CAP can be recommended as a practical and safe treatment with SA cement because it is suitable for chairside use with a handheld device and does not change the physical properties of the material.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
比较激光系统和冷大气等离子体对柔性混合陶瓷表面粗糙度和剪切结合强度的影响。
本研究旨在比较激光系统和冷大气等离子体(CAP)对柔性混合陶瓷(FHC)表面粗糙度(Ra)和剪切结合强度(SBS)的影响。80 个 FHC 样品被分成 5 组,分别进行表面处理(氢氟酸 (HFA)、HFA + 5 W Er: YAG 激光器 (HFA + 5WE)、HFA + 3 W Er: YAG 激光器 (HFA + 3WE)、HFA + 超快光纤激光器 (HFA + FL) 和 CAP)。Er: YAG 激光器(AT Fidelis Plus III,斯洛文尼亚 Fotona 公司)使用 0.9 毫米直径的尖端,每个脉冲输出 250 mJ 和 150 mJ,输出功率分别为 5 W 和 3 W,流量分别为 23,623 J/cm² 和 14,157 J/cm²,频率为 20 Hz,脉冲持续时间为 600 µs,持续 30 秒、CAP(PiezoBrush PZ3,Relyon Plasma,德国)以 5 cm2/s 的处理速度和 100% 的等离子功率对表面进行了 80 秒的处理。测量 Ra 值后,将每组样品分成两组,分别使用全蚀(TE)和自粘(SA)树脂水门汀进行粘接。测量所有样品的 SBS。从每组中随机抽取一个样本,用扫描电镜分析检查断裂表面。显著性水平为 P 0.05)。CAP 和 HFA 的 Ra 值没有差异(P > 0.05)。在 HFA + FL 之后用 TE 固结的样品的 SBS 值最高。在 CAP、HFA + 5WE、HFA + 3WE 和 HFA + FL 之后使用 TE 粘接的各组之间的 SBS 值差异不显著(P<0.05)。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
Lasers in Medical Science
Lasers in Medical Science 医学-工程:生物医学
CiteScore
4.50
自引率
4.80%
发文量
192
审稿时长
3-8 weeks
期刊介绍: Lasers in Medical Science (LIMS) has established itself as the leading international journal in the rapidly expanding field of medical and dental applications of lasers and light. It provides a forum for the publication of papers on the technical, experimental, and clinical aspects of the use of medical lasers, including lasers in surgery, endoscopy, angioplasty, hyperthermia of tumors, and photodynamic therapy. In addition to medical laser applications, LIMS presents high-quality manuscripts on a wide range of dental topics, including aesthetic dentistry, endodontics, orthodontics, and prosthodontics. The journal publishes articles on the medical and dental applications of novel laser technologies, light delivery systems, sensors to monitor laser effects, basic laser-tissue interactions, and the modeling of laser-tissue interactions. Beyond laser applications, LIMS features articles relating to the use of non-laser light-tissue interactions.
期刊最新文献
Effect of photobiomodulation and corticopuncture methods on tooth displacement and gene expression: animal study. Diabetic rats skin wounds treated with heterologous fibrin sealant followed by photobiomodulation therapy. Holmium laser enucleation of the prostate: Modified urethral mucosa pre-dissected technique versus traditional three-lobe technique-A retrospective study. Influence of photobiomodulation and radiofrequency on the healing of pressure lesions in mice. Influence of photobiomodulation on energy contribution in normoxia and hypoxia conditions in amateur runners: a single-blinded and randomized crossover pilot study.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1