Assessing the feasibility of handheld scanning technologies in neonatal intensive care: Trueness, acceptability, and suitability for personalised medical devices

IF 2.6 3区 医学 Q2 CRITICAL CARE MEDICINE Australian Critical Care Pub Date : 2024-12-11 DOI:10.1016/j.aucc.2024.09.012
Deanne August RN, PhD , Isabel Byram BBus, BFA , David Forrestal BEng, PhD , Mathilde Desselle MEng, PhD , Nathan Stevenson BEng, PhD , Kartik Iyer BEng, PhD , Mark W. Davies MBBS, PhD , Katherine White MBBS , Linda Cobbald BN, PGCertNg , Lynette Chapple BN, GradDipNg , Kellie McGrory CN, CertAdvNg , Margaret McLean BN , Stephanie Hall Bn, MNursPrac , Brittany Schoenmaker BN, MClinNurs , Jackie Clement BN, GradDipNg , Melissa M. Lai MBBS, PhD
{"title":"Assessing the feasibility of handheld scanning technologies in neonatal intensive care: Trueness, acceptability, and suitability for personalised medical devices","authors":"Deanne August RN, PhD ,&nbsp;Isabel Byram BBus, BFA ,&nbsp;David Forrestal BEng, PhD ,&nbsp;Mathilde Desselle MEng, PhD ,&nbsp;Nathan Stevenson BEng, PhD ,&nbsp;Kartik Iyer BEng, PhD ,&nbsp;Mark W. Davies MBBS, PhD ,&nbsp;Katherine White MBBS ,&nbsp;Linda Cobbald BN, PGCertNg ,&nbsp;Lynette Chapple BN, GradDipNg ,&nbsp;Kellie McGrory CN, CertAdvNg ,&nbsp;Margaret McLean BN ,&nbsp;Stephanie Hall Bn, MNursPrac ,&nbsp;Brittany Schoenmaker BN, MClinNurs ,&nbsp;Jackie Clement BN, GradDipNg ,&nbsp;Melissa M. Lai MBBS, PhD","doi":"10.1016/j.aucc.2024.09.012","DOIUrl":null,"url":null,"abstract":"<div><h3>Background</h3><div>Nasal continuous positive airway pressure (CPAP) injuries are common for premature infants. Clinical use of three-dimensional (3D) scanning is established in adult medicine, but the possibilities in neonatal care are still emerging. Custom printed CPAP devices have the potential to reduce injuries and disfigurement in this vulnerable population.</div></div><div><h3>Aim</h3><div>We sought to identify the most feasible portable 3D scanner for use in the neonatal intensive care environment towards the development of custom-fitting CPAP devices for premature infants.</div></div><div><h3>Methods</h3><div>Four handheld 3D scanners were assessed and compared, Artec Leo, Revopoint POP 2, iPad Pro/Metascan, and iPhone/Scandy Pro. Trained neonatal clinicians (medical and nursing) undertook mock scans in a simulated neonatal intensive care environment.</div></div><div><h3>Results</h3><div>Sixty scans were performed by 13 neonatal clinicians (four medical/nurse practitioners and nine nurses). The median mean absolute error was 0.21 mm (interquartile range [IQR]: 0.19–0.26), 0.17 mm (IQR: 0.15–0.21), and 1.08 mm (IQR: 1.0–1.63) for Artec Leo, Revopoint POP 2, and Scandy Pro, respectively. Scan times were the quickest for Artec Leo at 22.9 sec (IQR: 18.5–27), followed by Revopoint POP 2 at 25.2 sec (IQR: 22–34.4). Artec Leo was rated most expensive, but Revopoint POP 2 was rated more ergonomic. Both app-based 3D scanners (Metascan and Scandy Pro) presented data security issues.</div></div><div><h3>Conclusions</h3><div>Artec Leo and Revopoint POP 2 were identified as most feasible for use to perform 3D scans on premature infants in the neonatal intensive care environment.</div></div>","PeriodicalId":51239,"journal":{"name":"Australian Critical Care","volume":"38 2","pages":"Article 101127"},"PeriodicalIF":2.6000,"publicationDate":"2024-12-11","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Australian Critical Care","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S1036731424002789","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"CRITICAL CARE MEDICINE","Score":null,"Total":0}
引用次数: 0

Abstract

Background

Nasal continuous positive airway pressure (CPAP) injuries are common for premature infants. Clinical use of three-dimensional (3D) scanning is established in adult medicine, but the possibilities in neonatal care are still emerging. Custom printed CPAP devices have the potential to reduce injuries and disfigurement in this vulnerable population.

Aim

We sought to identify the most feasible portable 3D scanner for use in the neonatal intensive care environment towards the development of custom-fitting CPAP devices for premature infants.

Methods

Four handheld 3D scanners were assessed and compared, Artec Leo, Revopoint POP 2, iPad Pro/Metascan, and iPhone/Scandy Pro. Trained neonatal clinicians (medical and nursing) undertook mock scans in a simulated neonatal intensive care environment.

Results

Sixty scans were performed by 13 neonatal clinicians (four medical/nurse practitioners and nine nurses). The median mean absolute error was 0.21 mm (interquartile range [IQR]: 0.19–0.26), 0.17 mm (IQR: 0.15–0.21), and 1.08 mm (IQR: 1.0–1.63) for Artec Leo, Revopoint POP 2, and Scandy Pro, respectively. Scan times were the quickest for Artec Leo at 22.9 sec (IQR: 18.5–27), followed by Revopoint POP 2 at 25.2 sec (IQR: 22–34.4). Artec Leo was rated most expensive, but Revopoint POP 2 was rated more ergonomic. Both app-based 3D scanners (Metascan and Scandy Pro) presented data security issues.

Conclusions

Artec Leo and Revopoint POP 2 were identified as most feasible for use to perform 3D scans on premature infants in the neonatal intensive care environment.
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
评估手持式扫描技术在新生儿重症监护中的可行性:个性化医疗设备的真实性、可接受性和适用性。
背景:鼻腔持续气道正压通气(CPAP)损伤在早产儿中很常见。临床使用三维(3D)扫描是建立在成人医学,但在新生儿护理的可能性仍在出现。定制打印CPAP设备有可能减少这一弱势群体的伤害和毁容。目的:我们试图确定最可行的便携式3D扫描仪用于新生儿重症监护环境,以开发适合早产儿的定制CPAP设备。方法:对Artec Leo、revpoint POP 2、iPad Pro/Metascan和iPhone/ scany Pro四种手持式3D扫描仪进行评估和比较。训练有素的新生儿临床医生(医疗和护理)在模拟的新生儿重症监护环境中进行模拟扫描。结果:由13名新生儿临床医生(4名医疗/护士从业人员和9名护士)进行了60次扫描。Artec Leo、Revopoint POP 2和scany Pro的平均绝对误差中位数分别为0.21 mm(四分位间距[IQR]: 0.19-0.26)、0.17 mm(四分位间距[IQR]: 0.15-0.21)和1.08 mm(四分位间距[IQR]: 1.0-1.63)。扫描时间最快的是Artec Leo,为22.9秒(IQR: 18.5-27),其次是revpoint POP 2,为25.2秒(IQR: 22-34.4)。Artec Leo被评为最昂贵的,但revpoint POP 2被评为更符合人体工程学。基于应用程序的3D扫描仪(Metascan和scany Pro)都存在数据安全问题。结论:在新生儿重症监护环境下,Artec Leo和revpoint POP 2被认为是对早产儿进行3D扫描最可行的。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
Australian Critical Care
Australian Critical Care NURSING-NURSING
CiteScore
4.90
自引率
9.10%
发文量
148
审稿时长
>12 weeks
期刊介绍: Australian Critical Care is the official journal of the Australian College of Critical Care Nurses (ACCCN). It is a bi-monthly peer-reviewed journal, providing clinically relevant research, reviews and articles of interest to the critical care community. Australian Critical Care publishes peer-reviewed scholarly papers that report research findings, research-based reviews, discussion papers and commentaries which are of interest to an international readership of critical care practitioners, educators, administrators and researchers. Interprofessional articles are welcomed.
期刊最新文献
Wellbeing as perceived and experienced by intensive care unit nurses: An interpretive qualitative analysis Employer-provided wellbeing support for nurses working in intensive care units: A national cross-sectional study Understanding crisis needs among family caregivers of patients in critical care: A qualitative analysis Antidepressant use, but not polypharmacy, is associated with worse outcomes after in-hospital cardiac arrest in older people “Because I couldn't understand and respond”: A mixed-method study examining the impact of language barriers on patient experiences of intensive care unit outreach team care
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1