Results of the proficiency test for the Ka CMC of the SSDL-ININ Mexico, for diagnostic radiology IEC 61267:2005 and radiation protection ISO 4037:2019 X-ray beam qualities.
José Trinidad Álvarez Romero, Gonzalo Walwyn Salas, María de Los Angeles Montes Rodríguez, Mario Raymundo Cabrera Vertti, Andy Luis Romero Acosta
{"title":"Results of the proficiency test for the Ka CMC of the SSDL-ININ Mexico, for diagnostic radiology IEC 61267:2005 and radiation protection ISO 4037:2019 X-ray beam qualities.","authors":"José Trinidad Álvarez Romero, Gonzalo Walwyn Salas, María de Los Angeles Montes Rodríguez, Mario Raymundo Cabrera Vertti, Andy Luis Romero Acosta","doi":"10.1093/rpd/ncae208","DOIUrl":null,"url":null,"abstract":"<p><p>The present work describes the results for the bilateral comparison between the Secondary Standards Dosimetry Laboratories of the Instituto Nacional de Investigaciones Nucleares (SSDL-ININ), Mexico, and the pilot laboratory, the Centro de Protección e Higiene de las Radiaciones (SSDL-CPHR), Cuba, for the realization of the air kerma (${\\boldsymbol{K}}_{\\boldsymbol{a}}$) quantity to: (i) diagnostic X-ray (DXR) IEC 61267:2005 beam qualities: RQR 5 RQR 9, RQT 8, and RQT 9 and (ii) radiation protection (RP) ISO 4037:2019 X-ray beam qualities: N60, N100, and W60, as well S-Cs for gamma radiation. The matched and characterized reference fields are realized at SSDL-ININ with a Toshiba E7252FX X-ray tube. In all cases, the ${\\boldsymbol{K}}_{\\boldsymbol{a}}$ measurements are traceable to PTB, except for S-Cs, traceable to ININ. The SSDL-CPHR realizes its reference radiation beam qualities also as matched reference fields with a Pantak HF160C X-ray tube. For the S-Cs beam quality, the CPHR uses a Buchler OB6 irradiator; the ININ, a Siemens Cesagammatron. The transfer ionization chambers calibrated in the RQR and RQT qualities are an Exradin A650/DO70312; for the RQT qualities for CT, a PTW 30009/0655 and RC3CT/9217 were calibrated; and for the RP X-ray qualities, the Exradin A650/DO70312 and a PTW 32002/073 were calibrated. The ratio ${\\boldsymbol{R}}_{\\boldsymbol{k}}$ of the resulting calibration coefficients ${\\boldsymbol{N}}_{{\\boldsymbol{K}}_{\\boldsymbol{a}}}$ of each SSDL of the RQR qualities are consistent within the $\\boldsymbol{U}\\left(\\boldsymbol{k}=\\mathbf{2}\\right)\\%$ ≤ 2.3%; for the RQT qualities, the differences are less than the $\\boldsymbol{U}\\left(\\boldsymbol{k}=\\mathbf{2}\\right)\\%$ ≤ 3.6%; and for the RP X-ray qualities, within the order of $\\boldsymbol{U}\\left(\\boldsymbol{k}=\\mathbf{2}\\right)\\%$ ≤ 3.5%. For the S-Cs quality, the ${\\boldsymbol{R}}_{\\boldsymbol{k}}$ is in the order of $\\boldsymbol{U}\\left(\\boldsymbol{k}=\\mathbf{2}\\right)$ ≤ 3.0%. It should be noted that the ININ U's are generally larger due to the failure of the ${\\boldsymbol{k}}_{\\boldsymbol{TP}}$ correction factor, a brief discussion is given here. The ${\\boldsymbol{E}}_{\\boldsymbol{n}}$ score from ISO 17043 provides statistical support to this comparison.</p>","PeriodicalId":20795,"journal":{"name":"Radiation protection dosimetry","volume":" ","pages":""},"PeriodicalIF":0.8000,"publicationDate":"2024-12-03","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Radiation protection dosimetry","FirstCategoryId":"93","ListUrlMain":"https://doi.org/10.1093/rpd/ncae208","RegionNum":4,"RegionCategory":"环境科学与生态学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"ENVIRONMENTAL SCIENCES","Score":null,"Total":0}
引用次数: 0
Abstract
The present work describes the results for the bilateral comparison between the Secondary Standards Dosimetry Laboratories of the Instituto Nacional de Investigaciones Nucleares (SSDL-ININ), Mexico, and the pilot laboratory, the Centro de Protección e Higiene de las Radiaciones (SSDL-CPHR), Cuba, for the realization of the air kerma (${\boldsymbol{K}}_{\boldsymbol{a}}$) quantity to: (i) diagnostic X-ray (DXR) IEC 61267:2005 beam qualities: RQR 5 RQR 9, RQT 8, and RQT 9 and (ii) radiation protection (RP) ISO 4037:2019 X-ray beam qualities: N60, N100, and W60, as well S-Cs for gamma radiation. The matched and characterized reference fields are realized at SSDL-ININ with a Toshiba E7252FX X-ray tube. In all cases, the ${\boldsymbol{K}}_{\boldsymbol{a}}$ measurements are traceable to PTB, except for S-Cs, traceable to ININ. The SSDL-CPHR realizes its reference radiation beam qualities also as matched reference fields with a Pantak HF160C X-ray tube. For the S-Cs beam quality, the CPHR uses a Buchler OB6 irradiator; the ININ, a Siemens Cesagammatron. The transfer ionization chambers calibrated in the RQR and RQT qualities are an Exradin A650/DO70312; for the RQT qualities for CT, a PTW 30009/0655 and RC3CT/9217 were calibrated; and for the RP X-ray qualities, the Exradin A650/DO70312 and a PTW 32002/073 were calibrated. The ratio ${\boldsymbol{R}}_{\boldsymbol{k}}$ of the resulting calibration coefficients ${\boldsymbol{N}}_{{\boldsymbol{K}}_{\boldsymbol{a}}}$ of each SSDL of the RQR qualities are consistent within the $\boldsymbol{U}\left(\boldsymbol{k}=\mathbf{2}\right)\%$ ≤ 2.3%; for the RQT qualities, the differences are less than the $\boldsymbol{U}\left(\boldsymbol{k}=\mathbf{2}\right)\%$ ≤ 3.6%; and for the RP X-ray qualities, within the order of $\boldsymbol{U}\left(\boldsymbol{k}=\mathbf{2}\right)\%$ ≤ 3.5%. For the S-Cs quality, the ${\boldsymbol{R}}_{\boldsymbol{k}}$ is in the order of $\boldsymbol{U}\left(\boldsymbol{k}=\mathbf{2}\right)$ ≤ 3.0%. It should be noted that the ININ U's are generally larger due to the failure of the ${\boldsymbol{k}}_{\boldsymbol{TP}}$ correction factor, a brief discussion is given here. The ${\boldsymbol{E}}_{\boldsymbol{n}}$ score from ISO 17043 provides statistical support to this comparison.
期刊介绍:
Radiation Protection Dosimetry covers all aspects of personal and environmental dosimetry and monitoring, for both ionising and non-ionising radiations. This includes biological aspects, physical concepts, biophysical dosimetry, external and internal personal dosimetry and monitoring, environmental and workplace monitoring, accident dosimetry, and dosimetry related to the protection of patients. Particular emphasis is placed on papers covering the fundamentals of dosimetry; units, radiation quantities and conversion factors. Papers covering archaeological dating are included only if the fundamental measurement method or technique, such as thermoluminescence, has direct application to personal dosimetry measurements. Papers covering the dosimetric aspects of radon or other naturally occurring radioactive materials and low level radiation are included. Animal experiments and ecological sample measurements are not included unless there is a significant relevant content reason.