Response function measurement for a non-destructive gas-sheet beam profile monitor.

IF 1.3 4区 工程技术 Q3 INSTRUMENTS & INSTRUMENTATION Review of Scientific Instruments Pub Date : 2024-12-01 DOI:10.1063/5.0234781
Ippei Yamada, Junichiro Kamiya
{"title":"Response function measurement for a non-destructive gas-sheet beam profile monitor.","authors":"Ippei Yamada, Junichiro Kamiya","doi":"10.1063/5.0234781","DOIUrl":null,"url":null,"abstract":"<p><p>A gas-sheet beam profile monitor enabling non-destructive two-dimensional profile measurements of a high-intensity beam by capturing an image of a beam-induced fluorescence was developed. For quantitative profile measurements, the monitor's response function comprising, e.g., the gas sheet density distribution and the detector's sensitivity distributions must be experimentally clarified because the monitor output is a converted profile with the response function. A response function measurement method was devised based on the beam-profile-measurement method formula of the monitor. The response function was obtained by injecting a thin electron beam into the developed monitor and scanning its center position in transverse. The measured response function was evaluated by the J-PARC 3 MeV, 60 mA H- beam profile measurement. The 2-D beam profile was successfully reconstructed with the measured response function within the 2.74% residual of the least-squares method and 6.01% experimental statistic deviation. The projected 1-D profiles agreed well with those measured using a wire-scanning-type profile monitor.</p>","PeriodicalId":21111,"journal":{"name":"Review of Scientific Instruments","volume":"95 12","pages":""},"PeriodicalIF":1.3000,"publicationDate":"2024-12-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Review of Scientific Instruments","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1063/5.0234781","RegionNum":4,"RegionCategory":"工程技术","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"INSTRUMENTS & INSTRUMENTATION","Score":null,"Total":0}
引用次数: 0

Abstract

A gas-sheet beam profile monitor enabling non-destructive two-dimensional profile measurements of a high-intensity beam by capturing an image of a beam-induced fluorescence was developed. For quantitative profile measurements, the monitor's response function comprising, e.g., the gas sheet density distribution and the detector's sensitivity distributions must be experimentally clarified because the monitor output is a converted profile with the response function. A response function measurement method was devised based on the beam-profile-measurement method formula of the monitor. The response function was obtained by injecting a thin electron beam into the developed monitor and scanning its center position in transverse. The measured response function was evaluated by the J-PARC 3 MeV, 60 mA H- beam profile measurement. The 2-D beam profile was successfully reconstructed with the measured response function within the 2.74% residual of the least-squares method and 6.01% experimental statistic deviation. The projected 1-D profiles agreed well with those measured using a wire-scanning-type profile monitor.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
非破坏性气体片束剖面监测仪的响应函数测量。
开发了一种气体片光束剖面监测仪,通过捕获光束诱导荧光的图像,可以对高强度光束进行非破坏性二维剖面测量。对于定量剖面测量,监测器的响应函数包括,例如气体片密度分布和探测器的灵敏度分布,必须通过实验来澄清,因为监测器的输出是带有响应函数的转换剖面。根据监测仪的波束剖面测量方法公式,设计了响应函数测量方法。通过向所研制的监测仪注入细电子束,并在横向上扫描其中心位置,得到了响应函数。通过J-PARC 3 MeV, 60 mA的H束剖面测量来评估所测得的响应函数。在最小二乘法残差为2.74%、实验统计偏差为6.01%的范围内,成功地重建了二维光束轮廓。投影的1-D剖面与使用线扫描型剖面监视器测量的结果非常吻合。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
Review of Scientific Instruments
Review of Scientific Instruments 工程技术-物理:应用
CiteScore
3.00
自引率
12.50%
发文量
758
审稿时长
2.6 months
期刊介绍: Review of Scientific Instruments, is committed to the publication of advances in scientific instruments, apparatuses, and techniques. RSI seeks to meet the needs of engineers and scientists in physics, chemistry, and the life sciences.
期刊最新文献
Automated high-resolution 3D inspection methods for sealant applications in aerospace based on line structured light. Cryogenic front-end circuit for capacitive sensing in superconducting gravimeters. Development of a 300 kV/3 kHz nanosecond pulse generator using semiconductor opening switches. Measurement and characterization of internal delamination defects in CFRP based on line laser thermography frequency domain analysis. An efficiency improvement method for high-voltage nanosecond pulse spiral generator based on optimized voltage wave propagation process.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1