Rebecca J Wilson, Marcy Bhandari, Jane A Dickerson, Lisa M Johnson
{"title":"Validation and performance of microvue sC5b-9 plus ELISA on the Dynex DS2 platform.","authors":"Rebecca J Wilson, Marcy Bhandari, Jane A Dickerson, Lisa M Johnson","doi":"10.1016/j.cca.2025.120127","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>The complement membrane attack complex involves C5b-mediated assembly of C6-C9 polymers to form pores in cell membranes during complement activation. Inactive complexes can become soluble C5b-9 (sC5b-9) when they bind to Protein S. Elevated sC5b-9 levels are associated with increased risk of hematopoietic stem cell transplant-associated thrombotic microangiopathy (TA-TMA), a serious condition which can be improved with eculizumab therapy. Early detection of TA-TMA is essential for improving patient outcomes and optimizing the use of this costly treatment. We assessed the Quidel Microvue sC5b-9 Plus Enzyme Immunoassay on the Dynex-DS2 platform to screen for patients at risk of TA-TMA.</p><p><strong>Methods: </strong>EDTA plasma samples were collected from bone marrow transplant (BMT) patients and others not at risk for TA-TMA. Assay validation included correlation with another laboratory, precision, linearity, and hemolysis interference. Additionally, clinical accuracy was assessed through retrospective patient data analyses.</p><p><strong>Results: </strong>The assay showed acceptable intra- and interday precision, with less than 13 % variation. Linearity ranged from 80 to 1600 ng/mL, and there was no hemolysis interference up to 100 mg/dL. Clinical data revealed that monitoring sC5b-9 levels could detect significant increases indicative of TA-TMA, facilitating timely eculizumab intervention. Analytically, changes in sC5b-9 by 2- to 3-fold were significant for patient monitoring of TA-TMA. Lastly, a retrospective analysis on utility of the assay demonstrated effective utilization at our institution.</p><p><strong>Conclusion: </strong>The Quidel Microvue sC5b-9 Plus Enzyme Immunoassay demonstrated good analytical and clinical performance in screening patients with increased risk of TA-TMA.</p>","PeriodicalId":10205,"journal":{"name":"Clinica Chimica Acta","volume":" ","pages":"120127"},"PeriodicalIF":3.2000,"publicationDate":"2025-01-07","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Clinica Chimica Acta","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1016/j.cca.2025.120127","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Background: The complement membrane attack complex involves C5b-mediated assembly of C6-C9 polymers to form pores in cell membranes during complement activation. Inactive complexes can become soluble C5b-9 (sC5b-9) when they bind to Protein S. Elevated sC5b-9 levels are associated with increased risk of hematopoietic stem cell transplant-associated thrombotic microangiopathy (TA-TMA), a serious condition which can be improved with eculizumab therapy. Early detection of TA-TMA is essential for improving patient outcomes and optimizing the use of this costly treatment. We assessed the Quidel Microvue sC5b-9 Plus Enzyme Immunoassay on the Dynex-DS2 platform to screen for patients at risk of TA-TMA.
Methods: EDTA plasma samples were collected from bone marrow transplant (BMT) patients and others not at risk for TA-TMA. Assay validation included correlation with another laboratory, precision, linearity, and hemolysis interference. Additionally, clinical accuracy was assessed through retrospective patient data analyses.
Results: The assay showed acceptable intra- and interday precision, with less than 13 % variation. Linearity ranged from 80 to 1600 ng/mL, and there was no hemolysis interference up to 100 mg/dL. Clinical data revealed that monitoring sC5b-9 levels could detect significant increases indicative of TA-TMA, facilitating timely eculizumab intervention. Analytically, changes in sC5b-9 by 2- to 3-fold were significant for patient monitoring of TA-TMA. Lastly, a retrospective analysis on utility of the assay demonstrated effective utilization at our institution.
Conclusion: The Quidel Microvue sC5b-9 Plus Enzyme Immunoassay demonstrated good analytical and clinical performance in screening patients with increased risk of TA-TMA.
期刊介绍:
The Official Journal of the International Federation of Clinical Chemistry and Laboratory Medicine (IFCC)
Clinica Chimica Acta is a high-quality journal which publishes original Research Communications in the field of clinical chemistry and laboratory medicine, defined as the diagnostic application of chemistry, biochemistry, immunochemistry, biochemical aspects of hematology, toxicology, and molecular biology to the study of human disease in body fluids and cells.
The objective of the journal is to publish novel information leading to a better understanding of biological mechanisms of human diseases, their prevention, diagnosis, and patient management. Reports of an applied clinical character are also welcome. Papers concerned with normal metabolic processes or with constituents of normal cells or body fluids, such as reports of experimental or clinical studies in animals, are only considered when they are clearly and directly relevant to human disease. Evaluation of commercial products have a low priority for publication, unless they are novel or represent a technological breakthrough. Studies dealing with effects of drugs and natural products and studies dealing with the redox status in various diseases are not within the journal''s scope. Development and evaluation of novel analytical methodologies where applicable to diagnostic clinical chemistry and laboratory medicine, including point-of-care testing, and topics on laboratory management and informatics will also be considered. Studies focused on emerging diagnostic technologies and (big) data analysis procedures including digitalization, mobile Health, and artificial Intelligence applied to Laboratory Medicine are also of interest.