Evaluation of an automated assay for eosinophil-derived neurotoxin in serum.

IF 2.5 3区 医学 Q2 MEDICAL LABORATORY TECHNOLOGY Clinical biochemistry Pub Date : 2025-01-24 DOI:10.1016/j.clinbiochem.2025.110890
Carley Karsten, Theodore Stier, Christina Wood-Wentz, Carin Smith, Yifei K Yang, Melissa Snyder
{"title":"Evaluation of an automated assay for eosinophil-derived neurotoxin in serum.","authors":"Carley Karsten, Theodore Stier, Christina Wood-Wentz, Carin Smith, Yifei K Yang, Melissa Snyder","doi":"10.1016/j.clinbiochem.2025.110890","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>Eosinophil-derived neurotoxin (EDN) is a promising biomarker for eosinophil activation during inflammatory responses. Here we evaluate the analytical performance of an automated fluorescence enzyme immunoassay for EDN in serum and explore its relationship with eosinophil counts in both healthy participants and those with eosinophilic conditions.</p><p><strong>Materials and methods: </strong>Paired serum samples were collected from individuals for whom a complete blood count with differential was ordered. EDN was measured using the ImmunoCAP EDN Assay Kit (research use only, Phadia AB / provided by Thermo Fisher Scientific) and 40 samples were also measured using an ELISA kit (research use only, ALPCO).</p><p><strong>Results: </strong>The analytical measurement range of the ImmunoCAP assay was 2.6-200 µg/L. The imprecision across different EDN concentrations was ≤ 7.0 %. Stability and preanalytical requirements were determined. To minimize ex vivo degranulation and false elevation of EDN levels, serum should be removed from the cell pellet immediately after centrifugation. There was strong correlation for EDN measurements between ImmunoCAP and the comparative ELISA (r = 0.974), although a significant bias was observed. A 95th percentile reference range in 180 presumed healthy adults was calculated at 101 µg/L. Overall EDN was significantly higher in serum from patients with elevated circulating eosinophil counts (median = 120.0; P < 0.0001). However, individual patients may present with discordant presentation of eosinophil counts and EDN concentration.</p><p><strong>Conclusions: </strong>Together these results demonstrate that the ImmunoCAP EDN Assay Kit can reliably measure EDN in serum and may be useful for the evaluation of patients with conditions associated with hypereosinophilia.</p>","PeriodicalId":10172,"journal":{"name":"Clinical biochemistry","volume":" ","pages":"110890"},"PeriodicalIF":2.5000,"publicationDate":"2025-01-24","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Clinical biochemistry","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1016/j.clinbiochem.2025.110890","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Introduction: Eosinophil-derived neurotoxin (EDN) is a promising biomarker for eosinophil activation during inflammatory responses. Here we evaluate the analytical performance of an automated fluorescence enzyme immunoassay for EDN in serum and explore its relationship with eosinophil counts in both healthy participants and those with eosinophilic conditions.

Materials and methods: Paired serum samples were collected from individuals for whom a complete blood count with differential was ordered. EDN was measured using the ImmunoCAP EDN Assay Kit (research use only, Phadia AB / provided by Thermo Fisher Scientific) and 40 samples were also measured using an ELISA kit (research use only, ALPCO).

Results: The analytical measurement range of the ImmunoCAP assay was 2.6-200 µg/L. The imprecision across different EDN concentrations was ≤ 7.0 %. Stability and preanalytical requirements were determined. To minimize ex vivo degranulation and false elevation of EDN levels, serum should be removed from the cell pellet immediately after centrifugation. There was strong correlation for EDN measurements between ImmunoCAP and the comparative ELISA (r = 0.974), although a significant bias was observed. A 95th percentile reference range in 180 presumed healthy adults was calculated at 101 µg/L. Overall EDN was significantly higher in serum from patients with elevated circulating eosinophil counts (median = 120.0; P < 0.0001). However, individual patients may present with discordant presentation of eosinophil counts and EDN concentration.

Conclusions: Together these results demonstrate that the ImmunoCAP EDN Assay Kit can reliably measure EDN in serum and may be useful for the evaluation of patients with conditions associated with hypereosinophilia.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
求助全文
约1分钟内获得全文 去求助
来源期刊
Clinical biochemistry
Clinical biochemistry 医学-医学实验技术
CiteScore
5.10
自引率
0.00%
发文量
151
审稿时长
25 days
期刊介绍: Clinical Biochemistry publishes articles relating to clinical chemistry, molecular biology and genetics, therapeutic drug monitoring and toxicology, laboratory immunology and laboratory medicine in general, with the focus on analytical and clinical investigation of laboratory tests in humans used for diagnosis, prognosis, treatment and therapy, and monitoring of disease.
期刊最新文献
Oxidative stress and obesity are associated with endothelial dysfunction and subclinical atherosclerosis in adolescents. Evaluation of an automated assay for eosinophil-derived neurotoxin in serum. Assessing the predictive value of elevated triglycerides, triglyceride-glucose index (TyG), and TG/HDL ratios for cardiovascular disease and mortality during 20 years of follow-up: Tehran lipid and glucose study. Utility of the combination of high fluorescence cells and tumor markers for the diagnosis of malignant pleural effusions. Clinical significance of fecal Syndecan-2 gene methylation combined with blood tumor abnormal protein detection in the diagnosis of colorectal cancer and precancerous lesions.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1