Jiwon Lee, Wontae Kim, Jaeho Shim, Yujin Kim, Seonmi Kang, Young-Kyoung Yoo, Young-Ho Lee, Kangmoon Seo
{"title":"Validation of Recalibrated TONOVET Plus and TONOVET Pro in Canine Eyes Compared to Manometer and TONOVET.","authors":"Jiwon Lee, Wontae Kim, Jaeho Shim, Yujin Kim, Seonmi Kang, Young-Kyoung Yoo, Young-Ho Lee, Kangmoon Seo","doi":"10.1111/vop.70002","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>To evaluate the accuracy of recalibrated TONOVET Plus (rTVP) and the newly released TONOVET Pro (TVPR) in measuring intraocular pressure (IOP) in ex vivo canine eyes compared to manometer and the established TONOVET (TV).</p><p><strong>Procedure: </strong>Sixteen enucleated canine eyes were cannulated to establish manometric IOPs ranging from 5 to 80 mmHg. IOP measurements were obtained using TV, rTVP, and TVPR.</p><p><strong>Results: </strong>All tonometers showed a strong correlation with the manometer (R<sup>2</sup> > 0.98). Measurements from rTVP and TVPR exhibited almost perfect agreement with the manometer (CCC = 0.9974 for both), whereas TV demonstrated lower agreement (CCC = 0.9874) associated with greater systematic bias. Specifically, TV showed a lower proportional bias (95% confidence interval [CI] of the slope = 0.987-1.006) but a larger systematic bias (intercept = -3.33 mmHg) compared to rTVP and TVPR (95% CI of the slope = 1.045-1.066 and 1.045-1.063; intercept = -2.532 and -2.633 mmHg, respectively). In accuracy analyses by range, rTVP and TVPR showed no significant difference from the manometer in the 30-70 mmHg and 35-80 mmHg IOP ranges, respectively.</p><p><strong>Conclusions: </strong>Among the ICare rebound tonometers, rTVP and TVPR provided the most accurate IOP measurements in the ex vivo analysis and demonstrated particularly strong agreement in the high IOP range, suggesting their suitability for IOP measurement in glaucomatous patients.</p>","PeriodicalId":23836,"journal":{"name":"Veterinary ophthalmology","volume":" ","pages":""},"PeriodicalIF":1.7000,"publicationDate":"2025-02-11","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Veterinary ophthalmology","FirstCategoryId":"97","ListUrlMain":"https://doi.org/10.1111/vop.70002","RegionNum":4,"RegionCategory":"农林科学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"VETERINARY SCIENCES","Score":null,"Total":0}
引用次数: 0
Abstract
Objective: To evaluate the accuracy of recalibrated TONOVET Plus (rTVP) and the newly released TONOVET Pro (TVPR) in measuring intraocular pressure (IOP) in ex vivo canine eyes compared to manometer and the established TONOVET (TV).
Procedure: Sixteen enucleated canine eyes were cannulated to establish manometric IOPs ranging from 5 to 80 mmHg. IOP measurements were obtained using TV, rTVP, and TVPR.
Results: All tonometers showed a strong correlation with the manometer (R2 > 0.98). Measurements from rTVP and TVPR exhibited almost perfect agreement with the manometer (CCC = 0.9974 for both), whereas TV demonstrated lower agreement (CCC = 0.9874) associated with greater systematic bias. Specifically, TV showed a lower proportional bias (95% confidence interval [CI] of the slope = 0.987-1.006) but a larger systematic bias (intercept = -3.33 mmHg) compared to rTVP and TVPR (95% CI of the slope = 1.045-1.066 and 1.045-1.063; intercept = -2.532 and -2.633 mmHg, respectively). In accuracy analyses by range, rTVP and TVPR showed no significant difference from the manometer in the 30-70 mmHg and 35-80 mmHg IOP ranges, respectively.
Conclusions: Among the ICare rebound tonometers, rTVP and TVPR provided the most accurate IOP measurements in the ex vivo analysis and demonstrated particularly strong agreement in the high IOP range, suggesting their suitability for IOP measurement in glaucomatous patients.
期刊介绍:
Veterinary Ophthalmology is a peer-reviewed, international journal that welcomes submission of manuscripts directed towards academic researchers of veterinary ophthalmology, specialists and general practitioners with a strong ophthalmology interest. Articles include those relating to all aspects of:
Clinical and investigational veterinary and comparative ophthalmology;
Prospective and retrospective studies or reviews of naturally occurring ocular disease in veterinary species;
Experimental models of both animal and human ocular disease in veterinary species;
Anatomic studies of the animal eye;
Physiological studies of the animal eye;
Pharmacological studies of the animal eye.