Suture debris from high-tensile sutures contributes significantly to particle-induced tissue response in shoulder arthroplasty.

IF 2.9 2区 医学 Q1 ORTHOPEDICS Journal of Shoulder and Elbow Surgery Pub Date : 2025-02-25 DOI:10.1016/j.jse.2025.02.005
John P Scanaliato, Deborah J Hall, Songyun Liu, Gregory P Nicholson, Grant E Garrigues, Robin Pourzal
{"title":"Suture debris from high-tensile sutures contributes significantly to particle-induced tissue response in shoulder arthroplasty.","authors":"John P Scanaliato, Deborah J Hall, Songyun Liu, Gregory P Nicholson, Grant E Garrigues, Robin Pourzal","doi":"10.1016/j.jse.2025.02.005","DOIUrl":null,"url":null,"abstract":"<p><strong>Background and hypothesis: </strong>Polyethylene wear debris has long been known to be important for the development of osteolysis and aseptic loosening in total joint replacements. Evaluation of shoulder hemiarthroplasty (HA) specimens provided a unique opportunity to study the histopathologic response when a polyethylene-bearing surface was absent. We hypothesized that HAs would exhibit no significant inflammatory periprosthetic tissue response due to the absence of significant numbers of wear particles from the articulation.</p><p><strong>Methods: </strong>We analyzed 13 shoulder HAs. The explants were examined for damage to the bearing surface and taper damage using a stereo-microscope. The periprosthetic tissues were examined histologically for wear debris and cellular biological response. Fourier Transform Infrared Spectroscopic imaging (FTIR-I) and scanning electron microscopy (SEM) with energy dispersive x-ray spectroscopy (EDS) were used to characterize debris, if present, within tissue samples.</p><p><strong>Results: </strong>The average patient age at the time of implantation was 57.2 ± 7.3 and the average time in situ was 34.1 ± 29.4 months. The metal-bearing surface damage was primarily mild, with an average score of 1.5±0.7. The same was true for taper damage on both humeral stem and head tapers, with average scores of 1.7±0.5 and 1.7±0.9, respectively. The histopathological analysis revealed the considerable presence of metal debris in 11 (91.7%), cement debris in 4 (33.3%), and suture debris in 11 (83.3%) cases. SEM/EDS revealed titanium alloy debris to be the most dominant type of metal particle present, while FTIR-I scans suggested polyester to be the most commonly occurring type of suture debris. Particles were mostly found within and around macrophages. The mean macrophage score was 3.1 ± 0.8 and the mean foreign-body giant cell (FBGC) score was 2.3±1.1. There was, on average, no significant lymphocyte or neutrophil presence, except for a single septic case.</p><p><strong>Discussion and conclusion: </strong>This surgically retrieved shoulder HA cohort exhibited a considerable macrophage and FBGC response within the periprosthetic environment. Interestingly, given the low amount of damage to metallic surfaces and the absence of a polyethylene bearing, the tissue response does not appear to be driven by wear particles from the metal-bearing surface or taper junctions but rather from suture and, to a much lesser degree, cement and metal debris. Overall, these findings highlight the potential impact that high-tensile suture material, frequently used in high quantities around shoulder arthroplasty, may have on the periprosthetic environment and identify these materials as a potential driver of osteolysis.</p>","PeriodicalId":50051,"journal":{"name":"Journal of Shoulder and Elbow Surgery","volume":" ","pages":""},"PeriodicalIF":2.9000,"publicationDate":"2025-02-25","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Shoulder and Elbow Surgery","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1016/j.jse.2025.02.005","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"ORTHOPEDICS","Score":null,"Total":0}
引用次数: 0

Abstract

Background and hypothesis: Polyethylene wear debris has long been known to be important for the development of osteolysis and aseptic loosening in total joint replacements. Evaluation of shoulder hemiarthroplasty (HA) specimens provided a unique opportunity to study the histopathologic response when a polyethylene-bearing surface was absent. We hypothesized that HAs would exhibit no significant inflammatory periprosthetic tissue response due to the absence of significant numbers of wear particles from the articulation.

Methods: We analyzed 13 shoulder HAs. The explants were examined for damage to the bearing surface and taper damage using a stereo-microscope. The periprosthetic tissues were examined histologically for wear debris and cellular biological response. Fourier Transform Infrared Spectroscopic imaging (FTIR-I) and scanning electron microscopy (SEM) with energy dispersive x-ray spectroscopy (EDS) were used to characterize debris, if present, within tissue samples.

Results: The average patient age at the time of implantation was 57.2 ± 7.3 and the average time in situ was 34.1 ± 29.4 months. The metal-bearing surface damage was primarily mild, with an average score of 1.5±0.7. The same was true for taper damage on both humeral stem and head tapers, with average scores of 1.7±0.5 and 1.7±0.9, respectively. The histopathological analysis revealed the considerable presence of metal debris in 11 (91.7%), cement debris in 4 (33.3%), and suture debris in 11 (83.3%) cases. SEM/EDS revealed titanium alloy debris to be the most dominant type of metal particle present, while FTIR-I scans suggested polyester to be the most commonly occurring type of suture debris. Particles were mostly found within and around macrophages. The mean macrophage score was 3.1 ± 0.8 and the mean foreign-body giant cell (FBGC) score was 2.3±1.1. There was, on average, no significant lymphocyte or neutrophil presence, except for a single septic case.

Discussion and conclusion: This surgically retrieved shoulder HA cohort exhibited a considerable macrophage and FBGC response within the periprosthetic environment. Interestingly, given the low amount of damage to metallic surfaces and the absence of a polyethylene bearing, the tissue response does not appear to be driven by wear particles from the metal-bearing surface or taper junctions but rather from suture and, to a much lesser degree, cement and metal debris. Overall, these findings highlight the potential impact that high-tensile suture material, frequently used in high quantities around shoulder arthroplasty, may have on the periprosthetic environment and identify these materials as a potential driver of osteolysis.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
求助全文
约1分钟内获得全文 去求助
来源期刊
CiteScore
6.50
自引率
23.30%
发文量
604
审稿时长
11.2 weeks
期刊介绍: The official publication for eight leading specialty organizations, this authoritative journal is the only publication to focus exclusively on medical, surgical, and physical techniques for treating injury/disease of the upper extremity, including the shoulder girdle, arm, and elbow. Clinically oriented and peer-reviewed, the Journal provides an international forum for the exchange of information on new techniques, instruments, and materials. Journal of Shoulder and Elbow Surgery features vivid photos, professional illustrations, and explicit diagrams that demonstrate surgical approaches and depict implant devices. Topics covered include fractures, dislocations, diseases and injuries of the rotator cuff, imaging techniques, arthritis, arthroscopy, arthroplasty, and rehabilitation.
期刊最新文献
Suture debris from high-tensile sutures contributes significantly to particle-induced tissue response in shoulder arthroplasty. The Effect of the Central Post and Screw Constructs on the Univers Revers Total Shoulder System. Does dynamically tensioning suture alter outcomes after rotator cuff repair? Comparative and predictive modelling of modern anatomic total shoulder arthroplasty revision rates in osteoarthritis with different polyethylene glenoid designs. Glenohumeral Arthrodesis in Brachial Plexus Palsies: Open Surgery or Arthroscopy? A Retrospective Comparative Study.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1