Is simultaneous placement of orthodontic mini-implants and skeletally anchored appliances advisable? : CBCT-based cadaver study comparing the accuracy of the digital workflow with methods requiring two visits.

IF 1.3 4区 医学 Q3 DENTISTRY, ORAL SURGERY & MEDICINE Journal of Orofacial Orthopedics-Fortschritte Der Kieferorthopadie Pub Date : 2025-03-04 DOI:10.1007/s00056-025-00578-x
Till Augustinowitz, Lucia Schiavon, Katharina Mücke, Beryl Schwarz-Herzke, Dieter Drescher, Giulia Brunello, Kathrin Becker
{"title":"Is simultaneous placement of orthodontic mini-implants and skeletally anchored appliances advisable? : CBCT-based cadaver study comparing the accuracy of the digital workflow with methods requiring two visits.","authors":"Till Augustinowitz, Lucia Schiavon, Katharina Mücke, Beryl Schwarz-Herzke, Dieter Drescher, Giulia Brunello, Kathrin Becker","doi":"10.1007/s00056-025-00578-x","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>For the fabrication of an orthodontic mini-implant (OMI)-borne appliance, the position of the inserted OMI can be detected by a silicone impression or an intraoral scan (IOS). In case of digital planning, it can be taken over from the planning and the appliance can be produced in advance. This study aimed to evaluate the accuracy of these three techniques and whether there is an association with the insertion angle.</p><p><strong>Methods: </strong>Two OMIs were digitally planned and placed in the anterior palate of 11 human cadavers with different insertion angles. Subsequently, the position of each OMI was detected by an IOS, a silicone impression, and a cone-beam computed tomography (CBCT) scan, whereby the CBCT scan was set as \"real position\". The measurements of accuracy were performed between the CBCT data as a reference and the preoperative digital planning, the IOS and the plaster model manufactured from the silicone impression.</p><p><strong>Results: </strong>The IOS was the most accurate in detecting the Top (mean deviation 0.14 mm) and the Apex (mean deviation 0.36 mm) of the OMIs. Significant linear deviations between the three modalities were registered for both Top (p < 0.001) and Apex (p = 0.010). The digital planning procedure achieved the lowest mean angular deviation of 3.7° and was significantly more accurate in this respect than the IOS (p < 0.001).</p><p><strong>Conclusion: </strong>All methods were subject to small, but clinically irrelevant deviations. Within the limitations of a cadaver study, all methods appear to be suitable for clinical use. However, the digital workflow could be advantageous, requiring only a single visit for OMI placement and simultaneous appliance fitting.</p>","PeriodicalId":54776,"journal":{"name":"Journal of Orofacial Orthopedics-Fortschritte Der Kieferorthopadie","volume":" ","pages":""},"PeriodicalIF":1.3000,"publicationDate":"2025-03-04","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Orofacial Orthopedics-Fortschritte Der Kieferorthopadie","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1007/s00056-025-00578-x","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"DENTISTRY, ORAL SURGERY & MEDICINE","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: For the fabrication of an orthodontic mini-implant (OMI)-borne appliance, the position of the inserted OMI can be detected by a silicone impression or an intraoral scan (IOS). In case of digital planning, it can be taken over from the planning and the appliance can be produced in advance. This study aimed to evaluate the accuracy of these three techniques and whether there is an association with the insertion angle.

Methods: Two OMIs were digitally planned and placed in the anterior palate of 11 human cadavers with different insertion angles. Subsequently, the position of each OMI was detected by an IOS, a silicone impression, and a cone-beam computed tomography (CBCT) scan, whereby the CBCT scan was set as "real position". The measurements of accuracy were performed between the CBCT data as a reference and the preoperative digital planning, the IOS and the plaster model manufactured from the silicone impression.

Results: The IOS was the most accurate in detecting the Top (mean deviation 0.14 mm) and the Apex (mean deviation 0.36 mm) of the OMIs. Significant linear deviations between the three modalities were registered for both Top (p < 0.001) and Apex (p = 0.010). The digital planning procedure achieved the lowest mean angular deviation of 3.7° and was significantly more accurate in this respect than the IOS (p < 0.001).

Conclusion: All methods were subject to small, but clinically irrelevant deviations. Within the limitations of a cadaver study, all methods appear to be suitable for clinical use. However, the digital workflow could be advantageous, requiring only a single visit for OMI placement and simultaneous appliance fitting.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
求助全文
约1分钟内获得全文 去求助
来源期刊
CiteScore
3.90
自引率
0.00%
发文量
64
审稿时长
>12 weeks
期刊介绍: The Journal of Orofacial Orthopedics provides orthodontists and dentists who are also actively interested in orthodontics, whether in university clinics or private practice, with highly authoritative and up-to-date information based on experimental and clinical research. The journal is one of the leading publications for the promulgation of the results of original work both in the areas of scientific and clinical orthodontics and related areas. All articles undergo peer review before publication. The German Society of Orthodontics (DGKFO) also publishes in the journal important communications, statements and announcements.
期刊最新文献
Is simultaneous placement of orthodontic mini-implants and skeletally anchored appliances advisable? : CBCT-based cadaver study comparing the accuracy of the digital workflow with methods requiring two visits. Mitteilungen der DGKFO. Investigation of the cleaning performance of commercial orthodontic cleaning tablets regarding biofilm removal on PMMA test specimens : An ex vivo study. Comparison of six different CAD/CAM retainers vs. the stainless steel twistflex retainer: an in vitro investigation of survival rate and stability. Changes in the distribution of occlusal forces in the course of the orthodontic retention phase : A prospective cohort study.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1