Time efficiency and precision of splinted complete arch implant scanning

IF 4.8 2区 医学 Q1 DENTISTRY, ORAL SURGERY & MEDICINE Journal of Prosthetic Dentistry Pub Date : 2025-07-01 Epub Date: 2025-03-13 DOI:10.1016/j.prosdent.2025.02.017
Aly Abdelrehim BDS, MSc , Eshamsul bin Sulaiman BDS, MFD.RCS, MClinDent , Hazrina Sofian BCompSc, MCompSc, PhD , Nosizana Mohd Salleh DDS, PhD
{"title":"Time efficiency and precision of splinted complete arch implant scanning","authors":"Aly Abdelrehim BDS, MSc ,&nbsp;Eshamsul bin Sulaiman BDS, MFD.RCS, MClinDent ,&nbsp;Hazrina Sofian BCompSc, MCompSc, PhD ,&nbsp;Nosizana Mohd Salleh DDS, PhD","doi":"10.1016/j.prosdent.2025.02.017","DOIUrl":null,"url":null,"abstract":"<div><h3>Statement of problem</h3><div>The precision and accuracy of intraoral scanners for complete arch implant scanning is questionable as it is affected by the absence of anatomic features, the presence of large homogenous areas, and the lack of clear reference points, all of which create inherent errors in the image stitching process by the scanner software program.</div></div><div><h3>Purpose</h3><div>The purpose of this in vitro study was to test a newly designed scan-aid apparatus for evaluating the effect of geometric heterogeneity and additional reference points on the time efficiency and scanning precision of complete arch implant scanning.</div></div><div><h3>Material and methods</h3><div>A total of 60 scans (20 per group) were performed using an intraoral scanner in 3 different clinical simulations: 2 parallel implants, 4 parallel implants, and 4 implants; 2 anterior implants in parallel axial orientation; and 2 distal implants with a 30-degree posterior inclination. The scanning procedure was alternated between scanning with the scan-aid apparatus (test groups; 4IP+, 4IA+, 2IP+) and without (control groups; 4IP-, 4IA-, 2IP-). A 3-dimensional (3D) inspection software program was used to assess scan precision. Test group scans were compared with a reference scan for repeatability, with and without the scan-aid apparatus. One scan per group was the reference; the others were superimposed by using a best-fit algorithm. The 3D standard deviations were recorded for repeatability analysis. To assess time efficiency, the total scanning time, actual scanning time, and time lost were measured. A paired <em>t</em> test was used to compare means between the groups (α=.05).</div></div><div><h3>Results</h3><div>Precision was significantly higher in the test groups using the scan-aid apparatus than in the control groups (<em>P</em>&lt;.05). Furthermore, scanning with the scan-aid apparatus was significantly more time efficient (<em>P</em>&lt;.05).</div></div><div><h3>Conclusions</h3><div>Additional reference points and geometric heterogeneity using a scan-aid apparatus significantly improved the precision and time efficiency of complete arch implant scanning</div></div>","PeriodicalId":16866,"journal":{"name":"Journal of Prosthetic Dentistry","volume":"134 1","pages":"Pages 223.e1-223.e7"},"PeriodicalIF":4.8000,"publicationDate":"2025-07-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Prosthetic Dentistry","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0022391325001015","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2025/3/13 0:00:00","PubModel":"Epub","JCR":"Q1","JCRName":"DENTISTRY, ORAL SURGERY & MEDICINE","Score":null,"Total":0}
引用次数: 0

Abstract

Statement of problem

The precision and accuracy of intraoral scanners for complete arch implant scanning is questionable as it is affected by the absence of anatomic features, the presence of large homogenous areas, and the lack of clear reference points, all of which create inherent errors in the image stitching process by the scanner software program.

Purpose

The purpose of this in vitro study was to test a newly designed scan-aid apparatus for evaluating the effect of geometric heterogeneity and additional reference points on the time efficiency and scanning precision of complete arch implant scanning.

Material and methods

A total of 60 scans (20 per group) were performed using an intraoral scanner in 3 different clinical simulations: 2 parallel implants, 4 parallel implants, and 4 implants; 2 anterior implants in parallel axial orientation; and 2 distal implants with a 30-degree posterior inclination. The scanning procedure was alternated between scanning with the scan-aid apparatus (test groups; 4IP+, 4IA+, 2IP+) and without (control groups; 4IP-, 4IA-, 2IP-). A 3-dimensional (3D) inspection software program was used to assess scan precision. Test group scans were compared with a reference scan for repeatability, with and without the scan-aid apparatus. One scan per group was the reference; the others were superimposed by using a best-fit algorithm. The 3D standard deviations were recorded for repeatability analysis. To assess time efficiency, the total scanning time, actual scanning time, and time lost were measured. A paired t test was used to compare means between the groups (α=.05).

Results

Precision was significantly higher in the test groups using the scan-aid apparatus than in the control groups (P<.05). Furthermore, scanning with the scan-aid apparatus was significantly more time efficient (P<.05).

Conclusions

Additional reference points and geometric heterogeneity using a scan-aid apparatus significantly improved the precision and time efficiency of complete arch implant scanning
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
夹板全弓种植体扫描的时效性和精度。
问题陈述:口腔内扫描仪用于全弓种植体扫描的精度和准确性值得怀疑,因为它受到缺乏解剖特征,存在大面积均匀区域以及缺乏明确参考点的影响,所有这些都会在扫描仪软件程序的图像拼接过程中产生固有的错误。目的:本体外研究的目的是测试一种新设计的扫描辅助装置,以评估几何异质性和附加参考点对全弓种植体扫描的时间效率和扫描精度的影响。材料和方法:使用口腔内扫描仪在3种不同的临床模拟中共进行60次扫描(每组20次):2次平行种植,4次平行种植和4次种植;2个平行轴向前路种植体;2个远端植入物后倾角为30度。扫描过程在扫描辅助设备(试验组;4IP+、4IA+、2IP+)和无(对照组;4ip -, 4ia -, 2ip -)。使用三维(3D)检测软件程序评估扫描精度。测试组扫描与参考扫描进行重复性比较,有无扫描辅助装置。每组一次扫描作为参考;通过使用最佳拟合算法,将其他参数叠加在一起。记录三维标准偏差进行重复性分析。为了评估时间效率,测量了总扫描时间、实际扫描时间和损失时间。采用配对t检验比较各组间均数(α= 0.05)。结果:使用扫描辅助设备的试验组的精度明显高于对照组(结论:使用扫描辅助设备的额外参考点和几何异质性显著提高了全弓种植体扫描的精度和时间效率。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
Journal of Prosthetic Dentistry
Journal of Prosthetic Dentistry 医学-牙科与口腔外科
CiteScore
7.00
自引率
13.00%
发文量
599
审稿时长
69 days
期刊介绍: The Journal of Prosthetic Dentistry is the leading professional journal devoted exclusively to prosthetic and restorative dentistry. The Journal is the official publication for 24 leading U.S. international prosthodontic organizations. The monthly publication features timely, original peer-reviewed articles on the newest techniques, dental materials, and research findings. The Journal serves prosthodontists and dentists in advanced practice, and features color photos that illustrate many step-by-step procedures. The Journal of Prosthetic Dentistry is included in Index Medicus and CINAHL.
期刊最新文献
Letter to the Editor regarding, "Evaluation of the effects of the small intestine submucosa biological membrane on guided bone regeneration and early loading of dental implants in the maxillary anterior region: A randomized controlled trial". LEGO-modified mounting plates to mount 3D printed casts on dental articulators. Response to Letter to the Editor regarding, "Evaluation of the effects of the small intestine submucosa biological membrane on guided bone regeneration and early loading of dental implants in the maxillary anterior region: A randomized controlled trial". Robot-assisted immediate implant placement with the socket shield technique in the esthetic zone: A prospective 10-patient series. Internal fit and marginal adaptation of digitally fabricated interim fixed dental prostheses.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1