Effects of autoclave sterilization on the dimensional stability of implant surgical guides: A 3D in vitro analysis.

IF 3.4 2区 医学 Q1 DENTISTRY, ORAL SURGERY & MEDICINE Journal of Prosthodontics-Implant Esthetic and Reconstructive Dentistry Pub Date : 2025-03-17 DOI:10.1111/jopr.14049
Lucio Lo Russo, Pierluigi Mariani, Carlo Ercoli, Giusy Caprio, Ji-Man Park, Luigi Laino, Fariba Esperouz
{"title":"Effects of autoclave sterilization on the dimensional stability of implant surgical guides: A 3D in vitro analysis.","authors":"Lucio Lo Russo, Pierluigi Mariani, Carlo Ercoli, Giusy Caprio, Ji-Man Park, Luigi Laino, Fariba Esperouz","doi":"10.1111/jopr.14049","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>Surgical guides are medical devices used in static implant surgery to replicate the prosthetic and surgical implant position planned with computer-assisted software before surgery. Their use reduces invasiveness and improves safety and outcomes of implant surgery, provided that their accuracy is ensured from design to manufacturing and then to clinical usage. One of the factors that may potentially interfere with their accuracy is the sterilization procedure. The present study aimed to analyze how autoclave sterilization affects 3D-printed surgical guide stability.</p><p><strong>Materials and methods: </strong>Ten 3D-printed surgical guides were scanned before (T0) and after (T1) autoclave sterilization. T0 and T1 scans were compared with the corresponding CAD model in a surface matching software program. Trueness of the intaglio surface, linear and axial deviations of sleeves' housings at T0 and T1 were measured and statistically analyzed by generalized estimated equation (GEE) models to evaluate the effects of the sterilization procedure.</p><p><strong>Results: </strong>Surface variations, linear and axial deviations of sleeves' housings were found to be significantly (p < 0.05) affected by the sterilization only in combination with the volume or the total surface area of the surgical guides. The magnitude of the T0-T1 differences was small (0.016 mm and 0.035 mm for mean absolute deviation (MAD) at the intaglio surface in contact with teeth or mucosa; 0.009 mm for linear deviation of the central point of sleeves' housings; 0.084° and 0.180° for their axial deviation on the sagittal and frontal planes).</p><p><strong>Conclusion: </strong>The effects of autoclave sterilization on the stability of 3D-printed surgical guides remain limited and clinically insignificant.</p>","PeriodicalId":49152,"journal":{"name":"Journal of Prosthodontics-Implant Esthetic and Reconstructive Dentistry","volume":" ","pages":""},"PeriodicalIF":3.4000,"publicationDate":"2025-03-17","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Prosthodontics-Implant Esthetic and Reconstructive Dentistry","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1111/jopr.14049","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"DENTISTRY, ORAL SURGERY & MEDICINE","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: Surgical guides are medical devices used in static implant surgery to replicate the prosthetic and surgical implant position planned with computer-assisted software before surgery. Their use reduces invasiveness and improves safety and outcomes of implant surgery, provided that their accuracy is ensured from design to manufacturing and then to clinical usage. One of the factors that may potentially interfere with their accuracy is the sterilization procedure. The present study aimed to analyze how autoclave sterilization affects 3D-printed surgical guide stability.

Materials and methods: Ten 3D-printed surgical guides were scanned before (T0) and after (T1) autoclave sterilization. T0 and T1 scans were compared with the corresponding CAD model in a surface matching software program. Trueness of the intaglio surface, linear and axial deviations of sleeves' housings at T0 and T1 were measured and statistically analyzed by generalized estimated equation (GEE) models to evaluate the effects of the sterilization procedure.

Results: Surface variations, linear and axial deviations of sleeves' housings were found to be significantly (p < 0.05) affected by the sterilization only in combination with the volume or the total surface area of the surgical guides. The magnitude of the T0-T1 differences was small (0.016 mm and 0.035 mm for mean absolute deviation (MAD) at the intaglio surface in contact with teeth or mucosa; 0.009 mm for linear deviation of the central point of sleeves' housings; 0.084° and 0.180° for their axial deviation on the sagittal and frontal planes).

Conclusion: The effects of autoclave sterilization on the stability of 3D-printed surgical guides remain limited and clinically insignificant.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
求助全文
约1分钟内获得全文 去求助
来源期刊
CiteScore
7.90
自引率
15.00%
发文量
171
审稿时长
6-12 weeks
期刊介绍: The Journal of Prosthodontics promotes the advanced study and practice of prosthodontics, implant, esthetic, and reconstructive dentistry. It is the official journal of the American College of Prosthodontists, the American Dental Association-recognized voice of the Specialty of Prosthodontics. The journal publishes evidence-based original scientific articles presenting information that is relevant and useful to prosthodontists. Additionally, it publishes reports of innovative techniques, new instructional methodologies, and instructive clinical reports with an interdisciplinary flair. The journal is particularly focused on promoting the study and use of cutting-edge technology and positioning prosthodontists as the early-adopters of new technology in the dental community.
期刊最新文献
Effects of autoclave sterilization on the dimensional stability of implant surgical guides: A 3D in vitro analysis. What occlusal scheme should be used for tooth-supported fixed partial dentures? Relationship between palatal coverage and accuracy of texture perception: A preliminary study using an experimental palatal plate. Obtaining full-arch implant scan with smartphone video and deep learning: An in vitro investigation on trueness and precision. Sleep medicine, sleep apnea, and prosthodontics.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1