Kelly Byron, Fraser Robb, Pascal Stang, Shreyas Vasanawala, John Pauly, Greig Scott
{"title":"An RF-gated wireless power transfer system for wireless MRI receive arrays","authors":"Kelly Byron, Fraser Robb, Pascal Stang, Shreyas Vasanawala, John Pauly, Greig Scott","doi":"10.1002/cmr.b.21360","DOIUrl":null,"url":null,"abstract":"<p>In MRI systems, cable-free receive arrays would simplify setup while reducing the bulk and weight of coil arrays and improve patient comfort and throughput. Since battery power would limit scan time, wireless power transfer (WPT) is a viable option to continuously supply several watts of power to on-coil electronics. To minimize added noise and decouple the wireless power system from MRI coils, restrictions are placed on the coil geometry of the wireless power system, which are shown to limit its efficiency. Continuous power harvesting can also cause a large increase in the background noise of the image due to diode rectifier up-conversion of noise around the frequency of the transmitted power. However, by RF gating the transmitted power off during the MRI receive time while continuing to supply power from a storage capacitor, WPT is demonstrated to have minimal impact on image quality at received power levels up to 11 W. The integration of WPT with a 1.5T scanner is demonstrated.</p>","PeriodicalId":50623,"journal":{"name":"Concepts in Magnetic Resonance Part B-Magnetic Resonance Engineering","volume":"47B 4","pages":""},"PeriodicalIF":0.9000,"publicationDate":"2018-02-14","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1002/cmr.b.21360","citationCount":"13","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Concepts in Magnetic Resonance Part B-Magnetic Resonance Engineering","FirstCategoryId":"3","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1002/cmr.b.21360","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"CHEMISTRY, PHYSICAL","Score":null,"Total":0}
引用次数: 13
Abstract
In MRI systems, cable-free receive arrays would simplify setup while reducing the bulk and weight of coil arrays and improve patient comfort and throughput. Since battery power would limit scan time, wireless power transfer (WPT) is a viable option to continuously supply several watts of power to on-coil electronics. To minimize added noise and decouple the wireless power system from MRI coils, restrictions are placed on the coil geometry of the wireless power system, which are shown to limit its efficiency. Continuous power harvesting can also cause a large increase in the background noise of the image due to diode rectifier up-conversion of noise around the frequency of the transmitted power. However, by RF gating the transmitted power off during the MRI receive time while continuing to supply power from a storage capacitor, WPT is demonstrated to have minimal impact on image quality at received power levels up to 11 W. The integration of WPT with a 1.5T scanner is demonstrated.
期刊介绍:
Concepts in Magnetic Resonance Part B brings together engineers and physicists involved in the design and development of hardware and software employed in magnetic resonance techniques. The journal welcomes contributions predominantly from the fields of magnetic resonance imaging (MRI), nuclear magnetic resonance (NMR), and electron paramagnetic resonance (EPR), but also encourages submissions relating to less common magnetic resonance imaging and analytical methods.
Contributors come from both academia and industry, to report the latest advancements in the development of instrumentation and computer programming to underpin medical, non-medical, and analytical magnetic resonance techniques.