Clinical Application of MPRAGE Wave Controlled Aliasing in Parallel Imaging (Wave-CAIPI): A Comparative Study with MPRAGE GRAPPA.

IF 4.6 Q2 MATERIALS SCIENCE, BIOMATERIALS ACS Applied Bio Materials Pub Date : 2022-10-01 Epub Date: 2021-10-01 DOI:10.2463/mrms.mp.2021-0065
Azusa Sakurama, Yasutaka Fushimi, Satoshi Nakajima, Akihiko Sakata, Takuya Hinoda, Sonoko Oshima, Sayo Otani, Krishna Pandu Wicaksono, Wei Liu, Takakuni Maki, Tomohisa Okada, Ryosuke Takahashi, Yuji Nakamoto
{"title":"Clinical Application of MPRAGE Wave Controlled Aliasing in Parallel Imaging (Wave-CAIPI): A Comparative Study with MPRAGE GRAPPA.","authors":"Azusa Sakurama,&nbsp;Yasutaka Fushimi,&nbsp;Satoshi Nakajima,&nbsp;Akihiko Sakata,&nbsp;Takuya Hinoda,&nbsp;Sonoko Oshima,&nbsp;Sayo Otani,&nbsp;Krishna Pandu Wicaksono,&nbsp;Wei Liu,&nbsp;Takakuni Maki,&nbsp;Tomohisa Okada,&nbsp;Ryosuke Takahashi,&nbsp;Yuji Nakamoto","doi":"10.2463/mrms.mp.2021-0065","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To compare reliability and elucidate clinical application of magnetization-prepared rapid gradient-echo (MPRAGE) with 9-fold acceleration by using wave-controlled aliasing in parallel imaging (Wave-CAIPI 3 × 3) in comparison to conventional MPRAGE accelerated by using generalized autocalibrating partially parallel acquisition (GRAPPA) 2 × 1.</p><p><strong>Methods: </strong>A total of 26 healthy volunteers and 33 patients were included in this study. Subjects were scanned with two MPRAGEs, GRAPPA 2 × 1 and Wave-CAIPI 3 × 3 acquired in 5 min 21 s and 1 min 42 s, respectively, on a 3T MR scanner. Healthy volunteers underwent additional two MPRAGEs (CAIPI 3 × 3 and GRAPPA 3 × 3). The image quality of the four MPRAGEs was visually evaluated with a 5-point scale in healthy volunteers, and the SNR of four MPRAGEs was also calculated by measuring the phantom 10 times with each MPRAGE. Based on the results of the visual evaluation, voxel-based morphometry (VBM) analyses, including subfield analysis, were performed only for GRAPPA 2 × 1 and Wave-CAIPI 3 × 3. Correlation of segmentation results between GRAPPA 2 × 1 and Wave-CAIPI 3 × 3 was assessed.</p><p><strong>Results: </strong>In visual evaluations, scores for MPRAGE GRAPPA 2 × 1 (mean rank: 4.00) were significantly better than those for Wave-CAIPI 3 × 3 (mean rank: 3.00), CAIPI 3 × 3 (mean rank: 1.83), and GRAPPA 3 × 3 (mean rank: 1.17), and scores for Wave-CAIPI 3×3 were significantly better than those for CAIPI 3 × 3 and GRAPPA 3 × 3. Image noise was evident at the center for additional MPRAGE CAIPI 3 × 3 and GRAPPA 3 × 3. The correlation of segmentation results between GRAPPA 2 × 1 and Wave-CAIPI 3 × 3 was higher than 0.85 in all VOIs except globus pallidus. Subfield analysis of hippocampus also showed a high correlation between GRAPPA 2 × 1 and Wave-CAIPI 3 × 3.</p><p><strong>Conclusion: </strong>MPRAGE Wave-CAIPI 3 × 3 shows relatively better contrast, despite of its short scan time of 1 min 42 s. The volumes derived from automated segmentation of MPRAGE Wave-CAIPI are considered to be reliable measures.</p>","PeriodicalId":2,"journal":{"name":"ACS Applied Bio Materials","volume":"21 4","pages":"633-647"},"PeriodicalIF":4.6000,"publicationDate":"2022-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://ftp.ncbi.nlm.nih.gov/pub/pmc/oa_pdf/ff/c9/mrms-21-633.PMC9618934.pdf","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"ACS Applied Bio Materials","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.2463/mrms.mp.2021-0065","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2021/10/1 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"MATERIALS SCIENCE, BIOMATERIALS","Score":null,"Total":0}
引用次数: 1

Abstract

Purpose: To compare reliability and elucidate clinical application of magnetization-prepared rapid gradient-echo (MPRAGE) with 9-fold acceleration by using wave-controlled aliasing in parallel imaging (Wave-CAIPI 3 × 3) in comparison to conventional MPRAGE accelerated by using generalized autocalibrating partially parallel acquisition (GRAPPA) 2 × 1.

Methods: A total of 26 healthy volunteers and 33 patients were included in this study. Subjects were scanned with two MPRAGEs, GRAPPA 2 × 1 and Wave-CAIPI 3 × 3 acquired in 5 min 21 s and 1 min 42 s, respectively, on a 3T MR scanner. Healthy volunteers underwent additional two MPRAGEs (CAIPI 3 × 3 and GRAPPA 3 × 3). The image quality of the four MPRAGEs was visually evaluated with a 5-point scale in healthy volunteers, and the SNR of four MPRAGEs was also calculated by measuring the phantom 10 times with each MPRAGE. Based on the results of the visual evaluation, voxel-based morphometry (VBM) analyses, including subfield analysis, were performed only for GRAPPA 2 × 1 and Wave-CAIPI 3 × 3. Correlation of segmentation results between GRAPPA 2 × 1 and Wave-CAIPI 3 × 3 was assessed.

Results: In visual evaluations, scores for MPRAGE GRAPPA 2 × 1 (mean rank: 4.00) were significantly better than those for Wave-CAIPI 3 × 3 (mean rank: 3.00), CAIPI 3 × 3 (mean rank: 1.83), and GRAPPA 3 × 3 (mean rank: 1.17), and scores for Wave-CAIPI 3×3 were significantly better than those for CAIPI 3 × 3 and GRAPPA 3 × 3. Image noise was evident at the center for additional MPRAGE CAIPI 3 × 3 and GRAPPA 3 × 3. The correlation of segmentation results between GRAPPA 2 × 1 and Wave-CAIPI 3 × 3 was higher than 0.85 in all VOIs except globus pallidus. Subfield analysis of hippocampus also showed a high correlation between GRAPPA 2 × 1 and Wave-CAIPI 3 × 3.

Conclusion: MPRAGE Wave-CAIPI 3 × 3 shows relatively better contrast, despite of its short scan time of 1 min 42 s. The volumes derived from automated segmentation of MPRAGE Wave-CAIPI are considered to be reliable measures.

Abstract Image

Abstract Image

Abstract Image

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
MPRAGE波控混叠在平行成像(Wave- caipi)中的临床应用:与MPRAGE GRAPPA的比较研究
目的:比较波控混叠平行成像(Wave-CAIPI 3 × 3)加速9倍的磁化制备快速梯度回波(MPRAGE)与广义自校准部分平行采集(GRAPPA) 2 × 1加速的传统MPRAGE的可靠性,并阐明其临床应用。方法:选取健康志愿者26例,患者33例。受试者在3T MR扫描仪上分别于5分21秒和1分42秒获得两种mprage, GRAPPA 2 × 1和Wave-CAIPI 3 × 3。健康志愿者额外进行2次MPRAGE (CAIPI 3 × 3和GRAPPA 3 × 3),以5分制视觉评价4次MPRAGE的图像质量,并通过每次MPRAGE测量10次幻像计算4次MPRAGE的信噪比。基于视觉评价结果,仅对GRAPPA 2 × 1和Wave-CAIPI 3 × 3进行基于体素的形态测量(VBM)分析,包括子场分析。评估GRAPPA 2 × 1与Wave-CAIPI 3 × 3分割结果的相关性。结果:在视觉评价中,MPRAGE grappa2 × 1评分(平均排名4.00)显著优于Wave-CAIPI 3×3评分(平均排名3.00)、CAIPI 3×3评分(平均排名1.83)和grappa3 ×3评分(平均排名1.17),Wave-CAIPI 3×3评分显著优于CAIPI 3×3和grappa3 ×3评分。附加MPRAGE CAIPI 3 × 3和GRAPPA 3 × 3时,图像中心有明显的噪声。在除苍白球外的所有VOIs中,GRAPPA 2 × 1与Wave-CAIPI 3 × 3分割结果的相关性均大于0.85。海马的子场分析也显示了GRAPPA 2 × 1与Wave-CAIPI 3 × 3之间的高度相关性。结论:MPRAGE Wave-CAIPI 3 × 3扫描时间较短,仅为1 min 42 s,但成像效果较好。由MPRAGE Wave-CAIPI自动分割得到的体积被认为是可靠的测量方法。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
ACS Applied Bio Materials
ACS Applied Bio Materials Chemistry-Chemistry (all)
CiteScore
9.40
自引率
2.10%
发文量
464
期刊最新文献
A Systematic Review of Sleep Disturbance in Idiopathic Intracranial Hypertension. Advancing Patient Education in Idiopathic Intracranial Hypertension: The Promise of Large Language Models. Anti-Myelin-Associated Glycoprotein Neuropathy: Recent Developments. Approach to Managing the Initial Presentation of Multiple Sclerosis: A Worldwide Practice Survey. Association Between LACE+ Index Risk Category and 90-Day Mortality After Stroke.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1